journal article·2019-06-27Open access

Characterisation of the Material and Mechanical Properties of Atomic Force Microscope Cantilevers with a Plan-View Trapezoidal Geometry

Ashley D. Slattery, Adam J. Blanch, Cameron J. Shearer, Andrew J. Stapleton, Renee V. Goreham, Sarah L. Harmer, Jamie S. Quinton, Christopher T. Gibson

Applied Sciences· Volume 9· Issue 13

Abstract

Cantilever devices have found applications in numerous scientific fields and instruments, including the atomic force microscope (AFM), and as sensors to detect a wide range of chemical and biological species. The mechanical properties, in particular, the spring constant of these devices is crucial when quantifying adhesive forces, material properties of surfaces, and in determining deposited mass for sensing applications. A key component in the spring constant of a cantilever is the plan-view shape. In recent years, the trapezoidal plan-view shape has become available since it offers certain advantages to fast-scanning AFM and can improve sensor performance in fluid environments. Euler beam equations relating cantilever stiffness to the cantilever dimensions and Young’s modulus have been proven useful and are used extensively to model cantilever mechanical behaviour and calibrate the spring constant. In this work, we derive a simple correction factor to the Euler beam equation for a beam-shaped cantilever that is applicable to any cantilever with a trapezoidal plan-view shape. This correction factor is based upon previous analytical work and simplifies the application of the previous researchers formula. A correction factor to the spring constant of an AFM cantilever is also required to calculate the torque produced by the tip when it contacts the sample surface, which is also dependent on the plan-view shape. In this work, we also derive a simple expression for the torque for triangular plan-view shaped cantilevers and show that for the current generation of trapezoidal plan-view shaped AFM cantilevers, this will be a good approximation. We shall apply both these correction factors to determine Young’s modulus for a range of trapezoidal-shaped AFM cantilevers, which are specially designed for fast-scanning. These types of AFM probes are much smaller in size when compared to standard AFM probes. In the process of analysing the mechanical properties of these cantilevers, important insights are also gained into their spring constant calibration and dimensional factors that contribute to the variability in their spring constant.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Force Microscopy Techniques and Applications · Mechanical and Optical Resonators · Advanced MEMS and NEMS Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Flinders Institute for NanoScale Science and Technology, College of Science and Engineering, Flinders University, Bedford Park 5042, Australia

Adelaide Microscopy, The University of Adelaide, Adelaide 5005, Australia

Department of Biochemistry and Molecular Biology, Bio21 Institute, University of Melbourne, Parkville 3052, Australia

Department of Chemistry, The University of Adelaide, Adelaide 5005, Australia

School of Mathematical and Physical Sciences, University of Newcastle, Callaghan 2308, Australia

Flinders Microscopy and Microanalysis, College of Science and Engineering, Flinders University, Bedford Park 5042, Australia

References (60)

Show source-provided references
1. Theoretical analysis of the static deflection of plates for atomic force microscope applications10.1063/1.354137 ↗
2. atomic force microscopy as a tool applied to nano/biosensors10.3390/s120608278 ↗
3. Detecting protein aggregates on untreated human tissue samples by atomic force microscopy recognition imaging10.1016/j.bpj.2010.06.044 ↗
4. Nanobiosensors based on chemically modified afm probes: a useful tool for metsulfuron-methyl detection10.3390/s130201477 ↗
5. Atomic force microscopy-based antibody recognition imaging of proteins in the pathological deposits in pseudoexfoliation syndrome10.1016/j.ultramic.2011.03.008 ↗
6. Determination of the elastic properties of tomato fruit cells with an atomic force microscope10.3390/s130912175 ↗
7. Atomic force microscopy analysis of wool fibre surfaces in air and under water10.1016/S0968-4328(99)00076-1 ↗
8. Interaction of silver nanoparticles with tethered bilayer lipid membranes10.1021/acs.langmuir.5b00586 ↗
9. Application of atomic force microscopy to the study of micromechanical properties of biological materials10.1023/A:1005604028444 ↗
10. Accurate thickness measurement of graphene10.1088/0957-4484/27/12/125704 ↗
11. Cantilever transducers as a platform for chemical and biological sensors10.1063/1.1763252 ↗
12. Micromechanical cantilever-based biosensors10.1016/S0925-4005(01)00856-5 ↗
13. A nondestructive method for determining the spring constant of cantilevers for scanning force microscopy10.1063/1.1144209 ↗
14. Parallel beam approximation for V-shaped atomic force microscope cantilevers10.1063/1.1145292 ↗
15. A method for calculating the spring constant of atomic force microscopy cantilevers with a nonrectangular cross section10.1021/ac048828h ↗
16. Calibration of rectangular atomic force microscope cantilevers10.1063/1.1150021 ↗
17. Spring constant calibration of atomic force microscope cantilevers of arbitrary shape10.1063/1.4757398 ↗
18. Calibration of atomic force microscope cantilevers using only their resonant frequency and quality factor10.1063/1.4901227 ↗
19. A virtual instrument to standardise the calibration of atomic force microscope cantilevers10.1063/1.4962866 ↗
20. Atomic force microscope cantilever calibration using a focused ion beam10.1088/0957-4484/23/28/285704 ↗
21. Nondestructive and noncontact method for determining the spring constant of rectangular cantilevers10.1063/1.2720727 ↗
22. Practical implementation of dynamic methods for measuring atomic force microscope cantilever spring constants10.1088/0957-4484/17/9/010 ↗
23. Measuring the spring constant of atomic force microscope cantilevers: Thermal fluctuations and other methods10.1088/0957-4484/13/1/307 ↗
24. Cantilever spring constant calibration using laser Doppler vibrometry10.1063/1.2743272 ↗
25. Accurate and precise calibration of AFM cantilever spring constants using laser Doppler vibrometry10.1088/0957-4484/23/37/375702 ↗
26. Accurate measurement of atomic force microscope cantilever deflection excluding tip-surface contact with application to force calibration10.1016/j.ultramic.2013.03.009 ↗
27. Comparison of calibration methods for atomic-force microscopy cantilevers10.1088/0957-4484/14/1/301 ↗
28. Calibration of atomic force microscope cantilevers using standard and inverted static methods assisted by FIB-milled spatial markers10.1088/0957-4484/24/1/015710 ↗
29. Improved methods and uncertainty analysis in the calibration of the spring constant of an atomic force microscope cantilever using static experimental methods10.1088/0957-0233/20/12/125501 ↗
30. Accurate determination of spring constant of atomic force microscope cantilevers and comparison with other methods10.1016/j.measurement.2009.12.020 ↗
31. Quantitative analytical atomic force microscopy: A cantilever reference device for easy and accurate AFM spring-constant calibration10.1088/0957-0233/15/7/016 ↗
32. Precise atomic force microscope cantilever spring constant calibration using a reference cantilever array10.1063/1.2764372 ↗
33. Experimental determination of spring constants in atomic force microscopy10.1021/la00016a600 ↗
34. Method to determine the spring constant of atomic force microscope cantilevers10.1063/1.1642750 ↗
35. A calibration method for the higher modes of a micro-mechanical cantilever10.1063/1.4984222 ↗
36. Determining the spring constant of arbitrarily shaped cantilevers in viscous environments10.1063/1.5009071 ↗
37. A non-destructive method to calibrate the torsional spring constant of atomic force microscope cantilevers in viscous environments10.1063/1.5046648 ↗
38. Static and dynamic calibration of torsional spring constants of cantilevers10.1063/1.5045679 ↗
39. Calibrating torsional eigenmodes of micro-cantilevers for dynamic measurement of frictional forces10.1063/1.5038967 ↗
40. Susceptibility of atomic force microscope cantilevers to lateral forces10.1063/1.1544421 ↗
41. Susceptibility of atomic force microscope cantilevers to lateral forces: Experimental verification10.1063/1.1616657 ↗
42. High-speed AFM and nano-visualization of biomolecular processes10.1007/s00424-007-0406-0 ↗
43. Spring constant calibration techniques for next-generation fast-scanning atomic force microscope cantilevers10.1088/0957-4484/25/33/335705 ↗
44. Tailoring beam mechanics towards enhancing detection of hazardous biological species10.1007/s11340-006-9015-7 ↗
45. A new sensitivity improving approach for mass sensors through integrated optimization of both cantilever profile and cross-section10.1016/j.snb.2014.09.033 ↗
46. Longqi, X., Guiming, Z., Libo, Z., Yulong, Z., Zhuangde, J., Rahman, H., Hongyan, W., and Zhigang, L. (2013). A fluid viscosity sensor with resonant trapezoidal micro cantilever. IEEE Int. Symp. Assem. Manuf. (ISAM), 131–134.
47. A trapezoidal cantilever density sensor based on MEMS technology10.1631/jzus.C12MNT06 ↗
48. A trapezoidal flexoelectric accelerometer10.1177/1045389X13491021 ↗
49. Calibration of the effective spring constant of ultra-short cantilevers for a high-speed atomic force microscope10.1088/0957-0233/26/6/065001 ↗
50. Characterisation of the mechanical properties of thin film cantilevers with the atomic force microscope10.4028/www.scientific.net/MSF.189-190.107 ↗
51. Some properties of vapor deposited silicon nitride films using the SiH4-NH3-H2 system10.1149/1.2426719 ↗
52. Tai, Y.C., and Muller, R.S. (1990). Measurement of Young’s modulus on microfabricated structures using a surface profiler. Proc. IEEE Micro. Electro. Mechanical Syst., 147.10.1109/MEMSYS.1990.110267 ↗
53. Method for the calibration of atomic force microscope cantilevers10.1063/1.1145439 ↗
54. Microfabrication of cantilever styli for the atomic force microscope10.1116/1.576520 ↗
55. Influence of atomic force microscope cantilever tilt and induced torque on force measurements10.1063/1.2885734 ↗
56. Accurate analytical measurements in the atomic force microscope: A microfabricated spring constant standard potentially traceable to the SI10.1088/0957-4484/14/12/009 ↗
57. Shigley, J. (2004). Machine elements that absorb and store energy. Standard Handbook of Machine Design, McGraw-Hill.
58. Effects of aqueous exposure on the mechanical properties of wool fibers—Analysis by atomic force microscopy10.1177/004051750107100702 ↗
59. Normal stiffness calibration of microfabricated tri-layer conducting polymer actuators10.1088/0964-1726/18/6/065013 ↗
60. Characterization of silicon nitride micromechanical cantilevers and bridges