Design of a Metal 3D Printing Patient-Specific Repairing Thin Implant for Zygomaticomaxillary Complex Bone Fracture Based on Buttress Theory Using Finite Element Analysis
Source abstract
Abstract
This study developed a zygomaticomaxillary complex (ZMC) patient-specific repairing thin (PSRT) implant based on the buttress theory by integrating topology optimization and finite element (FE) analysis. An intact facial skeletal (IFS) model was constructed to perform topology optimization to obtain a hollow skeleton (HS) model with the structure and volume optimized. The PSRT implant was designed based on the HS contour which represented similar trends as vertical buttress pillars. A biomechanical analysis was performed on a ZMC fracture fixation with the PSRT implant and two traditional mini-plates under uniform axial loads applied on posterior teeth with 250 N. Results indicated that the variation in maximum bone stress and model volume between the IFS and HS models was 15.4% and 75.1%, respectively. Small stress variations between the IFS model and repairing with a PSRT implant (2.75–26.78%) were found for compressive stress at frontal process and tensile stress at the zygomatic process. Comparatively, large stress variations (30.67–96.26%) with different distributions between the IFS model and mini-plate models were found at the corresponding areas. This study concluded that the main structure/contour design of the ZMC repair implant according to the buttress position and orientation can obtain a favorable mechanical behavior.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Facial Trauma and Fracture Management · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Biomedical Engineering, National Yang-Ming University, Taipei 11221, Taiwan
Craniofacial Research Center, Department of Plastic and Reconstructive Surgery, Chang Gung Memorial Hospital, Linkou 33305, Taiwan
References (21)
Show source-provided references
1. Management of tripod fractures (zygomaticomaxillary complex) 1 point and 2 point fixations: A 5-year review10.4103/0975-7406.155937 ↗
2. Occlusal load distribution through the cortical and trabecular bone of the human mid-facial skeleton in natural dentition: A three-dimensional finite element study10.1016/j.aanat.2014.09.002 ↗
3. Structural biomechanics of the craniomaxillofacial skeleton under maximal masticatory loading: Inferences and critical analysis based on a validated computational model10.1016/j.bjps.2017.01.021 ↗
4. Experimental research of accurate reduction of zygomatic-orbitomaxillary complex fractures with individual templates10.1016/j.joms.2010.09.006 ↗
5. Surgical treatment of unilateral zygomaticomaxillary complex fractures: A 7-year observational study assessing treatment outcome in 153 cases10.1016/j.jcms.2016.09.002 ↗
6. Application of a computer-assisted navigation system (CANS) in the delayed treatment of zygomatic fractures: A randomized controlled trial10.1016/j.joms.2016.10.001 ↗
7. Review of in vivo bone strain studies and finite element models of the zygomatic complex in humans and nonhuman primates: Implications for clinical research and practice10.1002/ar.23486 ↗
8. Biomechanical analysis to verify the buttress theory when using the anatomical thin titanium mesh plate for zygomaticomaxillary complex bone fracture10.1142/S0219519419400256 ↗
9. Application of skeletal buttress analogy in composite facial reconstruction10.1055/s-0028-1098966 ↗
10. Design of a maxillofacial prosthesis based on topology optimization10.1142/S0219519418500240 ↗
11. Comparison of 2 temporomandibular joint total joint prosthesis systems10.1053/joms.2003.50112 ↗
12. Engineering alloplastic temporomandibular joint replacements
13. Influence of masticatory fatigue on the fracture resistance of the pulpless teeth restored with quartz-fiber post-core and crown10.1038/ijos.2012.78 ↗
14. The feasibility of a custom-made endoprosthesis in mandibular reconstruction: Implant design and finite element analysis10.1016/j.jcms.2015.10.004 ↗
15. Interbody fusion cage design using integrated global layout and local microstructure topology optimization10.1097/01.BRS.0000134573.14150.1A ↗
16. Stress-based topology optimization of frame structures under geometric uncertainty10.1016/j.cma.2016.10.039 ↗
17. A customized fixation plate with novel structure designed by topological optimization for mandibular angle fracture based on finite element analysis10.1186/s12938-017-0422-z ↗
18. Seebach, M., Theurer, F., Foehr, P., Deimling, C., Burgkart, R., and Zaeh, M.F. (2018). Advances in Structural and Multidisciplinary Optimization. Design of Bone Plates for Mandibular Reconstruction Using Topology and Shape Optimization, Springer. [1st ed.].10.1007/978-3-319-67988-4_154 ↗
19. Development of a novel anatomical thin titanium mesh plate with reduction guidance and fixation function for Asian zygomatic-orbitomaxillary complex fracture10.1016/j.jcms.2017.11.009 ↗
20. Buttress fixation with plates10.1016/S1043-1810(05)80018-1 ↗
21. The dynamic role of buttress reconstruction after maxillectomy10.1097/01.PRS.0000156978.16358.1E ↗