Common-Mode Driven Synchronous Filtering of the Powerline Interference in ECG
Source abstract
Abstract
Powerline interference (PLI) is a major disturbing factor in ground-free biopotential acquisition systems. PLI produces both common-mode and differential input voltages. The first is suppressed by a high common-mode rejection ratio of bioamplifiers. However, the differential PLI component evoked by the imbalance of electrode impedances is amplified together with the diagnostic differential biosignal. Therefore, PLI filtering is always demanded and commonly managed by analog or digital band-rejection filters. In electrocardiography (ECG), PLI filters are not ideal, inducing QRS and ST distortions as a transient reaction to steep slopes, or PLI remains when its amplitude varies and PLI frequency deviates from the notch. This study aims to minimize the filter errors in wide deviation ranges of PLI amplitudes and frequencies, introducing a novel biopotential readout circuit with a software PLI demodulator–remodulator concept for synchronous processing of both differential-mode and common-mode signals. A closed-loop digital synchronous filtering (SF) algorithm is designed to subtract a PLI estimation from the differential-mode input in real time. The PLI estimation branch connected to the SF output includes four stages: (i) prefilter and QRS limiter; (ii) quadrature demodulator of the output PLI using a common-mode driven reference; (iii) two servo loops for low-pass filtering and the integration of in-phase and quadrature errors; (iv) quadrature remodulator for synthesis of the estimated PLI using the common-mode signal as a carrier frequency. A simulation study of artificially generated PLI sinusoids with frequency deviations (48–52 Hz, slew rate 0.01–0.1 Hz/s) and amplitude deviations (root mean square (r.m.s.) 50–1000 μV, slew rate 10–200 μV/s) is conducted for the optimization of SF servo loop settings with artificial signals from the CTS-ECG calibration database (10 s, 1 lead) as well as for the SF algorithm test with 40 low-noise recordings from the Physionet PTB Diagnostic ECG database (10 s, 12 leads) and CTS-ECG analytical database (10 s, 8 leads). The statistical study for the PLI frequencies (48–52 Hz, slew rate ≤ 0.1 Hz/s) and amplitudes (≤1000 μV r.m.s., slew rate ≤ 40 μV/s) show that maximal SF errors do not exceed 15 μV for any record and any lead, which satisfies the standard requirements for a peak ringing noise of < 25 μV. The signal-to-noise ratio improvement reaches 57–60 dB. SF is shown to be robust against phase shifts between differential- and common-mode PLI. Although validated for ECG signals, the presented SF algorithm is generalizable to different biopotential acquisition settings via surface electrodes (electroencephalogram, electromyogram, electrooculogram, etc.) and can benefit many diagnostic and therapeutic medical devices.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: ECG Monitoring and Analysis · Analog and Mixed-Signal Circuit Design · Advancements in PLL and VCO Technologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Institute of Biophysics and Biomedical Engineering, Bulgarian Academy of Sciences, Acad. G. Bonchev Str. Bl. 105, 1113 Sofia, Bulgaria
Funding
Bulgarian National Science Fund
References (67)
Show source-provided references
1. Electrograms (ECG, EEG, EMG, EOG)10.3233/THC-2010-0604 ↗
2. Kaniusas, E. (2019). Biomedical Signals and Sensors III, Linking Electric Biosignals and Biomedical Sensors, Springer Nature.10.1007/978-3-319-74917-4 ↗
3. Webster, J., and Nimunkar, A. (2020). Biopotential amplifiers. Medical Instrumentation: Applications and Design, John Wiley & Sons. [5th ed.].
4. Simple, fast method for patient body capacitance and power-line electric interference measurement10.1007/BF02442332 ↗
5. Estimation of stray coupling capacitances in biopotential measurements10.1007/s11517-011-0811-6 ↗
6. Pallas-Areny, R., and Colominas, J. (1989, January 9–12). Differential mode interferences in biopotential amplifiers. Proceedings of the Images of the Twenty-First Century. Proceedings of the Annual International Engineering in Medicine and Biology Society, Seattle, WA, USA.
7. Filters: When, Why, and How (Not) to Use Them10.1016/j.neuron.2019.02.039 ↗
8. Suppression of power-line interference by analog notch filtering in the ECG signal for heart rate variability analysis: To do or not to do?
9. A hybrid approach to the simultaneous eliminating of power-line interference and associated ringing artifacts in electrocardiograms10.1186/1475-925X-12-42 ↗
10. Smith, S. (2003). Digital Signal Processing. A Practical Guide for Engineers and Scientists, Elsevier.
11. Fast integer coefficient FIR filters to remove the ac interference and the high-frequency noise10.1007/BF02441495 ↗
12. Tanji, A.K., de Brito, M., Alves, M., Garcia, R., Chen, L., and Ama, N. (2021). Improved noise cancelling algorithm for electrocardiogram based on moving average adaptive filter. Electronics, 10.10.3390/electronics10192366 ↗
13. High-Q comb filter for mains interference suppression
14. Suppressing harmonic powerline interference using multiple-notch filtering methods with improved transient behavior10.1016/j.measurement.2012.03.004 ↗
15. Power line interference reduction in ECG using combination of MA method and IIR notch filter
16. Online digital filter and QRS detector applicable in low resource ECG monitoring systems10.1007/s10439-008-9553-5 ↗
17. Adaptive tunable notch filter for ECG signal enhancement10.1016/j.procs.2015.07.347 ↗
18. Developments in ECG Acquisition, Preprocessing, Parameter Measurement and Recording10.1109/51.664031 ↗
19. Removal of power-line interference from the ECG: A review of the subtraction procedure10.1186/1475-925X-4-50 ↗
20. Power-line interference elimination from ECG in case of non-multiplicity between the sampling rate and the power-line frequency10.1016/j.bspc.2008.04.006 ↗
21. Power-line interference removal from high sampled ECG signals using modified version of the subtraction procedure10.7546/ijba.2020.24.4.000802 ↗
22. Tompkins, W. (2006). Biomedical Signal Processing. C-Language Examples and Laboratory Experiments for the IBM PC, Prentice-Hall.
23. Haykin, S. (2014). Adaptive Filter Theory, Pearson Education Limited.
24. Maniruzzaman, M., Billah, K., Biswas, U., and Gain, B. (2012, January 18–19). Least-mean-square algorithm based adaptive filters for removing power line interference from ECG signal. Proceedings of the International Conference on Informatics Electronics & Vision, Dhaka, Bangladesh.10.1109/ICIEV.2012.6317472 ↗
25. Makwana, G., and Gupta, L. (2015, January 4–6). De-noising of Electrocardiogram (ECG) with Adaptive Filter Using MATLAB. Proceedings of the Fifth International Conference on Communication Systems and Network Technologies, Gwalior, India.10.1109/CSNT.2015.126 ↗
26. Ren, A., Du, Z., Li, J., Hu, F., Yang, X., and Abbas, H. (2017). Adaptive interference cancellation of ECG signals. Sensors, 17.10.3390/s17050942 ↗
27. Denoising ECG Signals Using Adaptive Filter Algorithm
28. An efficient cardiac signal enhancement using time–frequency realization of leaky adaptive noise cancelers for remote health monitoring systems10.1016/j.measurement.2013.07.009 ↗
29. Removal of multiple artifacts from ECG signal using cascaded multistage adaptive noise cancellers10.1016/j.array.2022.100133 ↗
30. Digital lock-in techniques for adaptive power-line interference extraction10.1088/0967-3334/29/7/009 ↗
31. A novel PLI suppression method in ECG by notch filtering with a modulation-based detection and frequency estimation scheme10.1016/j.bspc.2020.102150 ↗
32. Grandmaison, M., Belzile, J., Thibeault, C., and Gagnon, F. (2004, January 23). Frequency domain filter using an accurate reconfigurable FFT/IFFT core. Proceedings of the IEEE Workshop on Circuits and Systems, Montreal, QC, Canada.
33. Frequency-domain digital filtering techniques for the removal of powerline noise with application to the electrocardiogram10.1016/0010-4809(90)90035-B ↗
34. Real-time FFT-based notch filter for single-frequency noise cancellation: Application to electrocardiogram signal denoising10.4103/jmss.JMSS_3_20 ↗
35. Zhang, D., Wang, S., Li, F., Wang, J., Sangaiah, A., Sheng, V., and Ding, X. (2019). An ECG signal de-noising approach based on wavelet energy and sub-band smoothing filter. Appl. Sci., 9.10.3390/app9224968 ↗
36. A wavelet-based method for power-line interference removal in ECG signals10.1590/2446-4740.01817 ↗
37. High-quality biopotential acquisition without a reference electrode: Power-line interference reduction by adaptive impedance balancing in a mixed analog–digital design10.1007/s11517-022-02586-0 ↗
38. Software PLL for power-line interference synchronization: Design, modeling and simulation
39. Software PLL for power-line interference synchronization: Implementation and results
40. Zywietz, C. (1999). CTS-ECG Test Atlas, Center for Computer Electrocardiography, Biosignal Processing, Medical School. Available online: https://www.medteq.net/ctscse-database-information.
41. (2011). Medical Electrical Equipment—Part 2-25: Particular Requirements for the Basic Safety and Essential Performance of Electrocardiographs (Standard No. IEC 60601-2-25:2011).
42. Nutzung der EKG-signaldatenbank CARDIODAT der PTB über das Internet
43. PhysioBank PhysioToolkit and PhysioNet: Components of a new research resource for complex physiologic signals10.1161/01.CIR.101.23.e215 ↗
44. Driven-right-leg circuit design10.1109/TBME.1983.325168 ↗
45. Dobrev, D., and Neycheva, T. (2020, January 16–18). Software automatic gain control. Proceedings of the XXIX International Scientific Conference Electronics (ET), Sozopol, Bulgaria.
46. Drentea, C. (2010). Modern Communications Receiver Design and Technology, Artech House Inc.
47. Synchronous filtering10.1109/TCSI.2006.879055 ↗
48. A Third Method of Generation and Detection of Single-Sideband Signals10.1109/JRPROC.1956.275061 ↗
49. Dobrev, D., and Neycheva, T. (2022, January 13–15). Open-loop Software Automatic Gain Control: Common-mode Power-line Interference Stabilization During ECG Recording. Proceedings of the XXXI International Scientific Conference Electronics (ET), Sozopol, Bulgaria.10.1109/ET55967.2022.9920322 ↗
50. Perez, J., Pueyo, S., and Lopez, B. (2011). Automatic Gain Control. Techniques and Architectures for RF Receivers, Springer Nature AG.
51. Bench study of the accuracy of a commercial AED arrhythmia analysis algorithm in the presence of electromagnetic interferences10.1088/0967-3334/30/7/012 ↗
52. The effect of 50/60 Hz notch filter application on human and rat ECG recordings10.1088/0967-3334/31/1/004 ↗
53. Electrocardiogram signal filtering using circulant singular spectrum analysis and cascaded Savitzky-Golay filter10.1016/j.bspc.2022.103583 ↗
54. An Improved Adaptive Power Line Interference Canceller for Electrocardiography10.1109/TBME.2006.883631 ↗
55. An Intelligent Adaptive Filter for Elimination of Power Line Interference from High Resolution Electrocardiogram10.1109/ACCESS.2016.2548362 ↗
56. Saxena, S., Jais, R., and Hota, M.K. (2019, January 4–6). Removal of Powerline Interference from ECG Signal using FIR, IIR, DWT and NLMS Adaptive Filter. Proceedings of the 2019 IEEE International Conference on Communication and Signal Processing, Chennai, India.10.1109/ICCSP.2019.8698112 ↗
57. Power Line Interference Removal for High-Quality Continuous Biosignal Monitoring With Low-Power Wearable Devices10.1109/JSEN.2016.2536363 ↗
58. Biswas, U., and Maniruzzaman, M. (2014, January 10–12). Removing power line interference from ECG signal using adaptive filter and notch filter. Proceedings of the IEEE International Conference on Electrical Engineering and Information & Communication Technology, Dhaka, Bangladesh.10.1109/ICEEICT.2014.6919072 ↗
59. Noise-aware dictionary-learning-based sparse representation framework for detection and removal of single and combined noises from ECG signal10.1049/htl.2016.0077 ↗
60. An adaptive denoising approach to powerline interference reduction in ECG recording
61. Removal of Power-Line Interference from ECG Using Decomposition Methodologies and Kalman Filter Framework: A Comparative Study10.18280/ts.380334 ↗
62. Robust nonlinear aggregation operator for ECG powerline interference reduction10.1016/j.bspc.2021.102675 ↗
63. Neural Network Based Canceller for Powerline Interference in ECG signals
64. Qiu, Y., Huang, K., Xiao, F., and Shen, H. (2018, January 14–16). Power-Line Interference Suppression in Electrocardiogram Using Recurrent Neural Networks. Proceedings of the 10th International Congress on Image and Signal Processing, BioMedical Engineering and Informatics, Shanghai, China.10.1109/CISP-BMEI.2017.8302231 ↗
65. An adaptive filtering approach for electrocardiogram (ECG) signal noise reduction using neural networks10.1016/j.neucom.2013.02.010 ↗
66. Sparsity-based modified wavelet de-noising autoencoder for ECG signals10.1016/j.sigpro.2022.108605 ↗
67. Neycheva, T., and Dobrev, D. (2022, January 13–15). Design of Fractional Filters for Power-line Interference Suppression in ECG Signals. Proceedings of the XXXI International Scientific Conference Electronics (ET), Sozopol, Bulgaria.10.1109/ET55967.2022.9920330 ↗