Model-Guided Manufacturing of Transducer Arrays Based on Single-Fibre Piezocomposites
Source abstract
Abstract
For breast cancer imaging, ultrasound computer tomography (USCT) is an emerging technology. To improve the image quality of our full 3-D system, a new transducer array system (TAS) design was previously proposed. This work presents a manufacturing approach which realises this new design. To monitor the transducer quality during production, the electro-mechanical impedance (EMI) was measured initially and after each assembly step. To evaluate the measured responses, an extended Krimholtz–Leedom–Matthaei (KLM) transducer model was used. The model aids in interpreting the measured responses and presents a useful tool for evaluating parasitic electric effects and attenuation at resonance. For quality control, the phase angle at thickness resonance φ t was found to be the most specific EMI property. It can be used to verify the functionality of the piezocomposites and allows reliable detection of faults in the acoustic backing. Evaluating the final response of 68 transducers showed 5% variance of the series resonance frequency. This indicates good consistency of derived ultrasound performance parameters.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ultrasound Imaging and Elastography · Ultrasonics and Acoustic Wave Propagation · Photoacoustic and Ultrasonic Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Karlsruhe Institute of Technology, Institute for Data Processing and Electronics, 76344 Eggenstein-Leopoldshafen, Germany
References (30)
Show source-provided references
1. Breast Cancer Screening for High-Risk Patients of Different Ages and Risk—Which Modality Is Most Effective?
2. Mammographic Breast Density: Impact on Breast Cancer Risk and Implications for Screening10.1148/rg.352140106 ↗
3. Differences in the attenuation of ultrasound by normal, benign, and malignant breast tissue10.1002/jcu.1870040404 ↗
4. Johnson, S.A., Abbott, T., Bell, R., Berggren, M., Borup, D., Robinson, D., Wiskin, J., Olsen, S., and Hanover, B. (2007). Non-Invasive Breast Tissue Characterization Using Ultrasound Speed and Attenuation. Acoustical Imaging, Springer.10.1007/1-4020-5721-0_17 ↗
5. Three-dimensional Imaging in Ultrasound10.1007/BF00995508 ↗
6. Ruiter, N.V., Göbel, G., Berger, L., Zapf, M., and Gemmeke, H. (2011). Realization of an Optimized 3D USCT. Proc. SPIE Med. Imaging, 7961–7968.10.1117/12.877520 ↗
7. Ruiter, N.V., Schwarzenberg, G.F., Zapf, M., and Gemmeke, H. (2008). Conclusions from an experimental 3D Ultrasound Computer Tomograph. IEEE Nucl. Sci. Symp. Conf. Rec., 4502–4509.10.1109/NSSMIC.2008.4774292 ↗
8. 3D Ultrasound Computer Tomography for Medical Imaging10.1016/j.nima.2007.06.116 ↗
9. Ruiter, N., Zapf, M., Dapp, R., Hopp, T., Kaiser, W.A., and Gemmeke, H. (2013). First Results of a Clinical Study with 3D Ultrasound Computer Tomography. IEEE Int. Ultrason. Symp., 651–654.10.1109/ULTSYM.2013.0168 ↗
10. Development of Single-Fiber Piezocomposite Transducers for 3-D Ultrasound Computer Tomography10.1002/adem.201800423 ↗
11. Development of 40-MHz Ultrasonic Transducers via Soft Mold Process10.1109/TUFFC.2019.2923554 ↗
12. Development of fine-scale piezoelectric composites for transducers10.1002/aic.690431334 ↗
13. Gebhardt, S.E., Hohlfeld, K., Günther, P., and Neubert, H. (2017, January 1–3). Manufacturing Technologies for Ultrasonic Transducers in a Broad Frequency Range. Proceedings of the International Workshop on Medical Ultrasound Tomography, Speyer, Germany.
14. Structural Health Monitoring using Piezoelectric Impedance Measurements
15. Cochran, S., Demore, C., and Courtney, C. (2012). Modelling Ultrasonic-Transducer Performance: One-Dimensional Models, Woodhead Publishing.10.1533/9780857096302.2.187 ↗
16. New Equivalent Circuits for Elementary Piezoelectric Transducers
17. Zapf, M., Pfistner, P., Liberman, C.I., van Dongen, K., Jong, N.d., Leyrer, B., Gemmeke, H., and Ruiter, N.V. (2017, January 1–3). Dice-and-fill single element octagon transducers for next generation 3D USCT. Proceedings of the International Workshop on Medical Ultrasound Tomography, Speyer, Germany.
18. Kohout, B. (2014). Simulation, Analyse und Entwurf eines 3D Ultraschall-Computertomographen für Diagnose und Therapie. (German) [Simulation, Analysis and design of a 3D Ultrasound-Tomograph for Diagnosis and Therapy]. [Ph.D. Thesis, Karlsruhe Institute of Technology].
19. Gemmeke, H., Berger, L., Hopp, T., Zapf, M., Tan, W., Blanco, R., Leys, R., Peric, I., and Ruiter, N.V. (2017, January 1–3). The New Generation of the KIT 3D USCT. Proceedings of the International Workshop on Medical Ultrasound Tomography, Speyer, Germany.
20. PZT components derived from polysulphone spinning process10.1179/1743676114Y.0000000229 ↗
21. Birk, L. (2019). Aufbau und Charakterisierung einer Serie von Ultraschallwandlern fuer einen Ultraschall Computertomographen. (German) [Manufacturing and Characterization of a series of Ultrasound Transducers for Ultrasound Computer Tomography]. [Master’s Thesis, Karlsruhe Institute of Technology].
22. The design of efficient broad-band piezoelectric transducers10.1109/T-SU.1978.31001 ↗
23. KLM model for lossy piezoelectric transducers10.1016/S0041-624X(03)00101-X ↗
24. A calculation scheme for the optimum design of ultrasonic transducers10.1016/0041-624X(83)90033-1 ↗
25. Approximate Material Properties in Isotropic Materials10.1109/T-SU.1985.31608 ↗
26. Passive Materials for High-Frequency Ultrasound Transducers10.1117/12.350684 ↗
27. A benchmark for free vibration and effective coupling of thick piezoelectric smart structures10.1088/0964-1726/17/6/065007 ↗
28. ROC Graphs: Notes and Practical Considerations for Researchers
29. Jain, A.K., Murty, M.N., and Flynn, P.J. (1999). Data clustering. ACM Comput. Surv., 31.10.1145/331499.331504 ↗
30. Lenk, A., Ballas, R., Werthschützky, R., and Pfeifer, G. (2011). Electromechanical Systems in Microtechnology and Mechatronics, Springer.10.1007/978-3-642-10806-8 ↗