journal article·2022-03-05Open access

Optimizing the Minimum Detectable Difference of Gamma Camera SPECT Images via the Taguchi Analysis: A Feasibility Study with a V-Shaped Slit Gauge

Ching-Hsiu Ke, Wan-Ju Liu, Bing-Ru Peng, Lung-Fa Pan, Lung-Kwang Pan

Applied Sciences· Volume 12· Issue 5

Abstract

This study tried to propose an innovated idea of solidifying the resolution of gamma camera in routine quality control and recommended a quantified index as minimum detectable difference (MDD) of gamma camera SPECT images using the Taguchi analysis and an indigenous V-shaped slit gauge. The gauge was customized to fulfill the quantitative requirement of the Taguchi analysis. The MDD among slit gauge of derived SPECT image was calculated from two overlapped peak profiles collected from a tangent slice of the V-shaped slit gauge with two nearby peaks. In particular, MDD was evaluated as minimum distance between two peak centers through the Student’s t-test with a constant, 1.96, which indicates that two peak centers separated distant enough to create a 95% confidence level of separation. Eighteen combinations of six gamma camera scanned factors were organized according to Taguchi analysis. Accordingly, (A) collimator, (B) detector to targe distance, (C) total counts, (D) acquired energy width, (E) Matrix size, and (F) zoom of collected ROI with each of two or three levels were organized into 18 groups to collect the slit gauge images according to Taguchi L18 orthogonal array. Then, three well-trained radiologists were ranked the scanned gauge images to derive the fish-bone-plot of signal-to-noise ratio (S/N, dB) and correlated ANOVA. Furthermore, the quantified MDD was proposed to verify the optimal suggestion of gamma camera scanned protocol, and obtained the MDD as 8.4, 7.9, and 7.1 mm for the second group of original L18 preset, conventional, and the optimal preset, respectively. Thus, the optimal preset of gamma camera was achieved in this study. The MDD proved to be a successful index in quantifying the imaging resolution of a gamma camera.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Medical Imaging Techniques and Applications · Advanced X-ray and CT Imaging · Radiomics and Machine Learning in Medical Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Medical Imaging and Radiological Science, Central Taiwan University of Science and Technology, Takun, Taichung 406, Taiwan, China

Department of Optometry, Central Taiwan University of Science and Technology, Takun, Taichung 406, Taiwan, China

Department of Nuclear Medicine, Taichung Tzu Chi Hospital, Buddhist Tzu Chi Medical Foundation, Taichung 427, Taiwan, China

Department of Radiology, Taichung Armed Forces General Hospital, Taichung 411, Taiwan, China

Department of Cardiology, Taichung Armed Forces General Hospital, Taichung 411, Taiwan, China

Funding

Ministry of Science and Technology · Taichung armed forces general hospital

References (25)

Show source-provided references
1. Comparison of irregular flux viewer system with BONENAVI version for identification of Tc-99m MDP whole body bone scan metastasis images
2. Role of Tc-99m MDP bone scan in evaluation of osteoid osteoma at varied locations10.18203/2320-6012.ijrms20183256 ↗
3. Vegetation Ecology of Coastal belt of Khejuri area of Purba Medinipur District with special reference to Hijli coast, West Bengal, India
4. Rare splenic metastasis of renal cell carcinoma detected on 99m Tc-MDP bone scan10.4103/0972-3919.125784 ↗
5. The role of 18F–NaF PET/CT in metastatic bone disease10.1016/j.jbo.2015.08.002 ↗
6. Prospective Evaluation of 99mTc MDP Scintigraphy, 18F NaF PET/CT, and 18F FDG PET/CT for Detection of Skeletal Metastases10.1007/s11307-011-0486-2 ↗
7. Zhang, L., He, Q., Zhou, T., Zhang, B., Li, W., Peng, H., Zhong, X., Ma, L., and Zhang, R. (2019). Accurate characterization of 99mTc-MDP uptake in extraosseous neoplasm mimicking bone metastasis on whole-body bone scan: Contribution of SPECT/CT. BMC Med. Imaging, 19.10.1186/s12880-019-0345-1 ↗
8. Assessment of an in-house phantom for the quality control of a clinical gamma camera
9. Simulation-based evaluation and optimization of a new CdZnTe gamma-camera architecture (HiSens)10.1088/0031-9155/55/9/019 ↗
10. Proof of concept for low-dose molecular breast imaging with a dual-head CZT gamma camera. Part I. Evaluation in phantoms10.1118/1.4718665 ↗
11. Contrast-noise-ratio (CNR) analysis and optimisation of breast-specific gamma imaging (BSGI) acquisition protocols10.1186/2191-219X-3-21 ↗
12. The optimum Ga-67-citrate gamma camera imaging quality factors as first calculated and shown by the Taguchi’s analysis
13. The Optimization of the Single Photon Emission Computed Tomography Image Quality via Taguchi Analysis: A Feasibility Study of a V-Shaped Phantom10.1166/jmihi.2017.1997 ↗
14. Optimizing the TLD-100 readout system for various radiotherapy beam doses using the Taguchi methodology10.1016/j.apradiso.2009.12.001 ↗
15. Optimization for solidification of low-level-radioactive resin using Taguchi analysis10.1016/S0956-053X(01)00012-5 ↗
16. Roy, R.K. A Primer on the Taguchi Method, Society of Manufacturing Engineering. [2nd ed.].
17. The Fisher–Snedecor F distribution: A simple and accurate composite fading model10.1109/LCOMM.2017.2687438 ↗
18. Performance comparison of finely channeled LYSO- and GAGG-based Si-PM gamma cameras for high-resolution SPECT10.1016/j.nima.2017.08.013 ↗
19. Evaluation of three scatter correction methods based on estimation of photopeak scatter spectrum in SPECT imaging: A simulation study10.1016/j.ejmp.2014.05.008 ↗
20. Randomized double blind placebo control studies, the “Gold Standard” in intervention based studies10.4103/0253-7184.102130 ↗
21. (2004, October 04). MATLAB. Matrix Laboratory Developed by MathWorks. V7.0.1.24704 (R14). Available online: http://www.mathworks.com/products/matlab/whatsnew.html.
22. The statistical analysis of data from small groups10.1037/0022-3514.83.1.126 ↗
23. Taguchi method-based optimization of the minimum detectable difference of a cardiac X-ray imaging system using a precise line pair gauge10.1142/S021951941940030X ↗
24. Integration of Taguchi analysis with phantom and innovative gauges: Optimization of the CT scan protocol for peripheral arterial occlusive disease (paod) syndrome10.1142/S0219519420400059 ↗
25. Minimum detectable difference of CT angiography scans at various cardiac beats: Evaluation via a customized oblique V-shaped line gauge and pmma phantom10.1142/S0219519421400662 ↗