journal article·2022-04-18Open access

Can AI Automatically Assess Scan Quality of Hip Ultrasound?

Abhilash Rakkunedeth Hareendrananthan, Myles Mabee, Baljot S. Chahal, Sukhdeep K. Dulai, Jacob L. Jaremko

Applied Sciences· Volume 12· Issue 8

Abstract

Ultrasound images can reliably detect Developmental Dysplasia of the Hip (DDH) during early infancy. Accuracy of diagnosis depends on the scan quality, which is subjectively assessed by the sonographer during ultrasound examination. Such assessment is prone to errors and often results in poor-quality scans not being reported, risking misdiagnosis. In this paper, we propose an Artificial Intelligence (AI) technique for automatically determining scan quality. We trained a Convolutional Neural Network (CNN) to categorize 3D Ultrasound (3DUS) hip scans as ‘adequate’ or ‘inadequate’ for diagnosis. We evaluated the performance of this AI technique on two datasets—Dataset 1 (DS1) consisting of 2187 3DUS images in which each image was assessed by one reader for scan quality on a scale of 1 (lowest quality) to 5 (optimal quality) and Dataset 2 (DS2) consisting of 107 3DUS images evaluated semi-quantitatively by four readers using a 10-point scoring system. As a binary classifier (adequate/inadequate), the AI technique gave highly accurate predictions on both datasets (DS1 accuracy = 96% and DS2 accuracy = 91%) and showed high agreement with expert readings in terms of Intraclass Correlation Coefficient (ICC) and Cohen’s kappa coefficient (K). Using our AI-based approach as a screening tool during ultrasound scanning or postprocessing would ensure high scan quality and lead to more reliable ultrasound hip examination in infants.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Hip disorders and treatments · Cardiac Valve Diseases and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Radiology and Diagnostic Imaging, University of Alberta, Edmonton, AB T6G 2B7, Canada

College of Medicine, University of Saskatchewan, 107 Wiggins Rd, Saskatoon, SK S7N 5E5, Canada

References (35)

Show source-provided references
1. Hip Disease and the Prognosis of Total Hip Replacements10.1302/0301-620X.83B4.0830579 ↗
2. The Epidemiology and Demographics of Hip Dysplasia10.5402/2011/238607 ↗
3. Hip Dysplasia: A Significant Risk Factor for the Development of Hip Osteoarthritis. A Cross-Sectional Survey10.1093/rheumatology/keh436 ↗
4. The Economic Burden of Osteoarthritis
5. Artrocad Study Group Economic Burden of Knee and Hip Osteoarthritis in Spain10.1002/art.24214 ↗
6. Prevention of Hip Dysplasia in Children and Adults10.1016/j.ocl.2012.05.001 ↗
7. Indicators of Successful Use of the Pavlik Harness in Infants with Developmental Dysplasia of the Hip10.1007/s00264-006-0097-8 ↗
8. Developmental Dysplasia of the Hip: Real World Data from a Retrospective Analysis to Evaluate the Effectiveness of Universal Screening10.1007/s40477-020-00463-w ↗
9. Indirect Effects of COVID-19 on Child Health Care: Delayed Diagnosis of Developmental Dysplasia of the Hip10.1007/s40477-020-00502-6 ↗
10. Cochrane Review: Screening Programmes for Developmental Dysplasia of the Hip in Newborn Infants10.1002/ebch.1891 ↗
11. Developmental Dysplasia of the Hip10.1016/S0140-6736(07)60710-7 ↗
12. Risk Factors for Developmental Dysplasia of the Hip: Ultrasonographic Findings in the Neonatal Period
13. Ultrasound Screening of Hips at Risk for CDH. Failure to Reduce the Incidence of Late Cases10.1302/0301-620X.71B1.2644290 ↗
14. Fundamentals of Sonographic Diagnosis of Infant Hip Dysplasia10.1097/01241398-198411000-00015 ↗
15. Potential for Change in US Diagnosis of Hip Dysplasia Solely Caused by Changes in Probe Orientation: Patterns of Alpha-Angle Variation Revealed by Using Three-Dimensional US10.1148/radiol.14140451 ↗
16. Reliability of 2D and 3D Ultrasound for Infant Hip Dysplasia in the Hands of Novice Users10.1007/s00330-018-5699-1 ↗
17. Impact of Scan Quality on AI Assessment of Hip Dysplasia Ultrasound
18. Localizing Target Structures in Ultrasound Video—A Phantom Study10.1016/j.media.2013.05.003 ↗
19. Standard Plane Localization in Ultrasound by Radial Component Model and Selective Search10.1016/j.ultrasmedbio.2014.06.006 ↗
20. Integration of Local and Global Features for Anatomical Object Detection in Ultrasound
21. Maraci, M.A., Napolitano, R., Papageorghiou, A., and Noble, J.A. (2014). Searching for Structures of Interest in an Ultrasound Video Sequence. Machine Learning in Medical Imaging, Springer International Publishing.10.1007/978-3-319-10581-9_17 ↗
22. Hareendranathan, A.R., Zonoobi, D., Mabee, M., Cobzas, D., Punithakumar, K., Noga, M., and Jaremko, J.L. (2017, January 18–21). Toward Automatic Diagnosis of Hip Dysplasia from 2D Ultrasound. Proceedings of the 2017 IEEE 14th International Symposium on Biomedical Imaging (ISBI 2017), Melbourne, VIC, Australia.10.1109/ISBI.2017.7950680 ↗
23. Tran, D., Bourdev, L., Fergus, R., Torresani, L., and Paluri, M. (2015, January 7–13). Learning Spatiotemporal Features with 3d Convolutional Networks. Proceedings of the IEEE International Conference on Computer Vision, Santiago, Chile.10.1109/ICCV.2015.510 ↗
24. Carreira, J., and Zisserman, A. (2017, January 21–26). Quo Vadis, Action Recognition? A New Model and the Kinetics Dataset. Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition, Los Alamitos, CA, USA.10.1109/CVPR.2017.502 ↗
25. Çiçek, Ö., Abdulkadir, A., Lienkamp, S.S., Brox, T., and Ronneberger, O. (2016, January 17–21). 3D U-Net: Learning Dense Volumetric Segmentation from Sparse Annotation. Proceedings of the Medical Image Computing and Computer-Assisted Intervention—MICCAI 2016, Athens, Greece.10.1007/978-3-319-46723-8_49 ↗
26. Milletari, F., Navab, N., and Ahmadi, S.-A. (2016, January 25–28). V-Net: Fully Convolutional Neural Networks for Volumetric Medical Image Segmentation. Proceedings of the 2016 Fourth International Conference on 3D Vision (3DV), Stanford, CA, USA.10.1109/3DV.2016.79 ↗
27. 3D Deeply Supervised Network for Automated Segmentation of Volumetric Medical Images10.1016/j.media.2017.05.001 ↗
28. Paserin, O., Mulpuri, K., Cooper, A., Hodgson, A.J., and Abugharbieh, R. (2017, January 14). Automatic Near Real-Time Evaluation of 3D Ultrasound Scan Adequacy for Developmental Dysplasia of the Hip. Proceedings of the Computer Assisted and Robotic Endoscopy and Clinical Image-Based Procedures, Québec City, QC, Canada.10.1007/978-3-319-67543-5_12 ↗
29. Ultrasound Standard Plane Detection Using a Composite Neural Network Framework10.1109/TCYB.2017.2685080 ↗
30. Paserin, O., Mulpuri, K., Cooper, A., Hodgson, A.J., and Garbi, R. (2018, January 16–20). Real Time RNN Based 3D Ultrasound Scan Adequacy for Developmental Dysplasia of the Hip. Proceedings of the Medical Image Computing and Computer Assisted Intervention—MICCAI 2018, Granada, Spain.10.1007/978-3-030-00928-1_42 ↗
31. AIUM Practice Guideline for the Performance of an Ultrasound Examination for Detection and Assessment of Developmental Dysplasia of the Hip10.7863/jum.2009.28.1.114 ↗
32. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research10.1016/j.jcm.2016.02.012 ↗
33. A Technique for Semiautomatic Segmentation of Echogenic Structures in 3D Ultrasound, Applied to Infant Hip Dysplasia10.1007/s11548-015-1239-5 ↗
34. Toward Automated Classification of Acetabular Shape in Ultrasound for Diagnosis of DDH: Contour Alpha Angle and the Rounding Index10.1016/j.cmpb.2016.03.013 ↗
35. Quader, N., Hodgson, A., and Abugharbieh, R. (2014). Confidence Weighted Local Phase Features for Robust Bone Surface Segmentation in Ultrasound. Workshop on Clinical Image-Based Procedures, Springer.10.1007/978-3-319-13909-8_10 ↗