Abstract

(1) Background: Microwave ablation (MWA) has an efficient killing effect on primary and metastatic lung cancer. However, the treatment effect will be affected by the air in the lung, which makes it very difficult to accurately predict and control the ablation area; (2) Methods: In this paper, in vitro experiments combined with simulations are used to study the microwave ablation area of inflated porcine lung. The in vitro experiment is divided into inflated group and deflated group, combined with different ablation power (40 W, 50 W, 60 W) and ablation time (100 s, 200 s, 300 s) for experiment, each power and time combination are repeated five times. A total of 90 ablation experiments were performed. The simulation experiment uses COMSOL Multiphysics software to simulate the microwave ablation area of the inflated lung; (3) Results and Conclusions: When the ablation power is 40 W, 50 W, and 60 W, the average long diameter of the deflated group are 20.8–30.9%, 7.6–22.6%, 10.4–19.8% larger than those of the inflated group, respectively; the average short diameter of the deflated group is 24.5–41.4%, 31.6–45.7%, 27.3–42.9% larger than that of the inflated group. The results show that the ablation area of inflated lung is smaller than deflated lung, which is mainly due to the smaller ablation short diameter.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Ultrasound and Hyperthermia Applications · Ultrasound and Cavitation Phenomena · Ultrasound in Clinical Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Faculty of Environment and Life, Beijing International Science and Technology Cooperation Base for Intelligent Physiological Measurement and Clinical Transformation, Beijing University of Technology, Beijing 100124, China

Funding

National Natural Science Foundation of China · Beijing Municipal Natural Science Foundation · Jiangsu Science and Technology Department (Jiangsu Science and Technology Committee)

References (42)

Show source-provided references
1. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries10.3322/caac.21492 ↗
2. Lung Cancer Ablation: Technologies and Techniques10.1055/s-0033-1342955 ↗
3. Temperature field of microwave ablation of rabit lung tumors based on CT
4. Research status of influence factors in therapeutic effect of microwave ablation for lung tumors
5. Experimental and numerical study of microwave ablation on ex-vivo porcine lung10.1080/15368378.2019.1669635 ↗
6. Ablation therapy of non-colorectal cancer lung metastases: Retrospective analysis of tumour response post-laser-induced interstitial thermotherapy (LITT), radiofrequency ablation (RFA) and microwave ablation (MWA)
7. Application and prospect of thermal ablation in treatment of lung cancer
8. Expert consensus workshop report: Guidelines for thermal ablation of primary and metastatic lung tumors (2018 edition)10.4103/jcrt.JCRT_221_18 ↗
9. Local incompressible registration for liver ablation surgery assessment10.1002/mp.12535 ↗
10. Percutaneous radiofrequency ablation of malignancies in the lung10.2214/ajr.174.1.1740057 ↗
11. Percutaneous thermal ablation of primary and secondary lung tumors: Comparison between microwave and radiofrequency ablation10.1016/j.diii.2019.07.008 ↗
12. Yamazaki, N., Watanabe, H., Seki, M., Hoshi, T., Kobayashi, Y., Miyashita, T., and Fujie, M.G. (September, January 30). Modeling the internal pressure dependence of thermal conductivity and in vitro temperature measurement for lung RFA. Proceedings of the 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Boston, MA, USA.10.1109/IEMBS.2011.6091424 ↗
13. Status and advancement of microwave ablation in China10.1080/02656736.2016.1243261 ↗
14. Clinical experiences with microwave thermal ablation of lung malignancies10.1080/02656736.2016.1204630 ↗
15. Image-guided ablative therapies for lung cancer10.1016/j.rcl.2012.06.004 ↗
16. Update in Percutaneous Lung Ablation10.1007/s40134-019-0340-x ↗
17. Application of Microwave Ablation for Lung Cancer
18. Thermal ablation of lung tumors: Focus on microwave ablation10.1055/s-0043-109010 ↗
19. Microwave ablation of the lung in a porcine model vessel diameter predicts pulmonary artery occlusion10.1007/s00270-017-1689-y ↗
20. Short term treatment results of local ablation with water-cooled microwave antenna for liver cancer: Comparison with radiofrequency ablation
21. Microwave ablation at 915 MHz vs. 2.45 GHz: A theoretical and experimental investigation10.1118/1.4931959 ↗
22. Yamazaki, N., Watanabe, H., Lu, X.W., Isobe, Y., Kobayashi, Y., Miyashita, T., Fujie, M.G., Watanabe, H., Isobe, Y., and Kobayashi, Y. (2013, January 3–7). The relation between temperature distribution for lung RFA and electromagnetic wave frequency dependence of electrical conductivity with changing a lung’s internal air volumes. Proceedings of the 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan.10.1109/EMBC.2013.6609518 ↗
23. Analysis of thermal damage parameters in microwave ablation of liver cancer
24. Computer Modeling for Microwave Ablation in Breast Cancer Using a Coaxial Slot Antenna10.1007/s10765-015-1931-2 ↗
25. Estimaiton of SAR Distribution of a Tip-Split Array Applicator for Microwave Coagulation Therapy Using the Finite Element Method
26. Experimental validation of a combined electromagnetic and thermal FDTD model of a microwave heating process
27. Analysis of temperature distributions generated by square array applicator composed of coaxial-slot antennas for hyperthermia
28. Analysis of tissue and arterial blood temperatures in the resting human forearm10.1152/jappl.1998.85.1.5 ↗
29. Microwave Ablation (MWA): Basics, Technique and Results in Primary and Metastatic Liver Neoplasms—Review Article10.1055/s-0043-117410 ↗
30. Microwave Ablation of Lung Tissue: Impact of Single-Lung Ventilation on Ablation Size10.1016/j.athoracsur.2010.05.068 ↗
31. Comparison of Conventional and Cone-Beam CT for Monitoring and Assessing Pulmonary Microwave Ablation in a Porcine Model10.1016/j.jvir.2018.04.035 ↗
32. Effect of changes in lung physical properties on microwave ablation zone during respiration10.1007/s13534-019-00145-5 ↗
33. Effect of the Blood Vessel on the Microwave Ablation Temperature Field of Liver
34. Physical modeling of microwave ablation zone clinical margin variance10.1118/1.4942980 ↗
35. Katsuyama, Y., Yamazaki, N., Kobayashi, Y., Hoshi, T., and Miyashita, T. (September, January 28). A study on estimation of the deformation behavior in the collapse process of lung. Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, San Diego, CA, USA.
36. Yang, D. (2020). Research on Microwave Hyperthermia Planning System, Changchun University of Science and Technology.
37. Thermal ablation of biological tissues in disease treatment: A review of computational models and future directions10.1080/15368378.2020.1741383 ↗
38. Radmilović, R.M., Sabo, M., Prnova, M., Šoltes, L., and Radjenović, B. (2021). Finite Element Analysis of the Microwave Ablation Method for Enhanced Lung Cancer Treatment. Cancers, 13.10.3390/cancers13143500 ↗
39. Numerical Analysis of Human Cancer Therapy Using Microwave Ablation10.3390/app10010211 ↗
40. Hepatic microwave ablation with multiple antennae results in synergistically larger zones of coagulation necrosis10.1245/ASO.2003.03.045 ↗
41. In vivo evaluation of lung microwave ablation in a porcine tumor mimic model10.1007/s00270-012-0399-8 ↗
42. Large nearly spherical ablation zones are achieved with simultaneous multi-antenna microwave ablation applied to treat liver tumours10.1007/s00330-019-06431-1 ↗