journal article·2022-09-28Open access

Optimizing the Ultrasound Image Quality of Carotid Artery Stenosis Patients via Taguchi’s Dynamic Analysis and an Indigenous Water Phantom

Kai-Yu Hsiao, Chih-Sheng Lin, Wan-Ming Li, Shih-Hsun Huang, Yu-Ting Cho, Bing-Ru Peng, Lung-Kwang Pan, Lung-Fa Pan

Applied Sciences· Volume 12· Issue 19

Abstract

This study optimized the ultrasound image of carotid artery stenosis using Taguchi dynamic analysis and an indigenous water phantom. Eighteen combinations of seven essential factors of the ultrasound scan facility were organized according to Taguchi’s L18 orthogonal array. The seven factors were assigned as follows: (1) angle of probe; (2) signal gain; (3) resolution vs. speed; (4) dynamic range; (5) XRES; (6) zoom; (7) time gain compensation. An indigenous water phantom was customized to satisfy the quantified need in Taguchi’s analysis. Unlike the conventional dynamic Taguchi analysis, an innovative quantified index, the figure of merit (FOM), was proposed to integrate four specific quality characteristics, namely (i) average difference between the practical scan and theoretically preset area (78.5, 50.2 and 12.6 mm2) of stenosis, (ii) standard deviation of the average, (iii) practical scan’s sensitivity β to various stenosis diameters (10, 8, and 4 mm), and (iv) correlation coefficient r2 of the linear regressed sensitivity curve. The highest value (FOM = 0.413) was furnished by the optimal combination of factors on 18 groups under study, yielding high r2 and low β or standard deviation values and the best quality of ultrasound images for the further clinical diagnosis. The comparison between FOM and the conventional signal-to-noise (S/N) ratio in Taguchi’s analysis revealed that FOM compiled more quality characteristics that were superior by nature to fulfill the practical need in clinical diagnosis. The alternative choice in ultrasound scan optimization can be based on stenosis diameter variation from a different perspective to be explored in the follow-up study.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Photoacoustic and Ultrasonic Imaging · Cerebrovascular and Carotid Artery Diseases · Ultrasound Imaging and Elastography

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Surgery, Taichung Armed-Forces General Hospital, Taichung 411, Taiwan

Division of Thoracic Surgery, Department of Surgery, Tri-Service General Hospital, National Defense Medical Center, Taipei 114, Taiwan

Graduate Institute of Medical Science, National Defense Medical Center, Taipei 114, Taiwan

Department of Radiology, BenQ Medical Center, The Affiliated BenQ Hospital of the Nanjing Medical University, Nanjing 211166, China

Department of Biomedical Imaging and Radiological Science, China Medical University, Taichung 404, Taiwan

Department of Medical Imaging and Radiological Science, Central Taiwan University of Science and Technology, Taichung 40601, Taiwan

Department of Nursing, Taichung Armed Forces General Hospital, Taichung 411, Taiwan

Department of Cardiology, Taichung Armed Forces General Hospital, Taichung 411, Taiwan

References (24)

Show source-provided references
1. Ultrasound carotid plaque features, cardiovascular disease risk factors and events: The Multi-Ethnic Study of Atherosclerosis10.1016/j.atherosclerosis.2018.06.005 ↗
2. General principles of carotid Doppler ultrasonography10.14366/usg.13018 ↗
3. A comparative study of color Doppler imaging and contrast-enhanced ultrasound for the detection of ulceration in patients with carotid atherosclerotic disease10.1007/s00330-018-5773-8 ↗
4. Role of ultrasound and color Doppler imaging in the detection of carotid paragangliomas10.1016/j.jus.2012.05.001 ↗
5. Pixel distribution analysis of B-mode ultrasound scan images predicts histologic features of atherosclerotic carotid plaques10.1067/mva.2002.122888 ↗
6. A novel automatic algorithm for the segmentation of the lumen of the carotid artery in ultrasound B-mode images10.1016/j.eswa.2013.06.003 ↗
7. A new method for IVUS-based coronary artery disease risk stratification: A link between coronary & carotid ultrasound plaque burdens10.1016/j.cmpb.2015.10.022 ↗
8. Development of a Carotid Vulnerable Plaque Phantom Model Evaluated by Pixel Distribution Analysis10.1016/j.ultrasmedbio.2018.07.005 ↗
9. Multi element diverging beam from a linear array transducer for transverse cross sectional imaging of carotid artery: Simulations and phantom vessel validation10.1143/JJAP.50.07HF05 ↗
10. Noninvasive Vascular Modulography Method for Imaging the Local Elasticity of Atherosclerotic Plaques: Simulation and In Vitro Vessel Phantom Study10.1109/TUFFC.2017.2757763 ↗
11. Carotid Artery Disease Imaging: A Home-Produced, Easily Made Phantom for Two- and Three-Dimensional Ultrasound Simulation10.1177/154431671303700204 ↗
12. Wu, K.Y., Liang, C.C., Chuang, C.H., Pan, L.F., and Pan, L.K. (2022). Optimization of common iliac artery sonography image via an ingenious water phantom and Taguchi’s analysis: A feasibility study. Appl. Sci., 12.10.3390/app12168197 ↗
13. Taguchi method-based optimization of the minimum detectable difference of a cardiac X-ray imaging system using a precise line pair gauge10.1142/S021951941940030X ↗
14. Optimizing cardiac CT angiography minimum detectable difference via Taguchi’s dynamic algorithm, a V-shaped line gauge, and three PMMA phantoms10.3233/THC-228009 ↗
15. Optimizing left anterior oblique (LAO) caudal imaging in coronary angiography using the Taguchi method: A phantom study with clinical verification
16. Taguchi dynamic analysis application to computer tomography number-mass density linear dependence optimization10.1080/24699322.2017.1378791 ↗
17. Taguchi, G. (1992). Taguchi on Robust Technology Development: Bringing Quality Engineering Upstream, ASME Press.10.1115/1.800288 ↗
18. Application of adaptive image processing technique to real-time spatial compound ultrasound imaging improves image quality10.1097/01.RLI.0000057032.41715.14 ↗
19. Optimizing the TLD-100 readout system for various radiotherapy beam doses using the Taguchi methodology10.1016/j.apradiso.2009.12.001 ↗
20. Optimizing the TLD-100H readout system under various radioactive I-131 doses via the revised Taguchi dynamic quality loss function
21. Evaluating the quality characteristics of TLD-100T and TLD-100H exposed to diagnostic X-rays and 64 multislice CT using Taguchi’s quality loss function10.1016/j.radmeas.2015.06.010 ↗
22. Optimizing multiple qualities of Nd: YAG laser welding onto magnesium alloy via grey relational analysis10.1179/174329305X46727 ↗
23. Optimizing multiple quality characteristics via Taguchi method-based Grey analysis10.1016/j.jmatprotec.2006.07.015 ↗
24. Optimizing the minimum detectable difference of computed tomography scanned images via the Taguchi analysis: A feasibility study with an indigenous hepatic phantom and a line group gauge