Application of Stockwell Transform and Shannon Energy for Pace Pulses Detection in a Single-Lead ECG Corrupted by EMG Artifacts
Source abstract
Abstract
Electrocardiogram (ECG) analysis is important for the detection of pace pulse artifacts, since their existence indicates the presence of a pacemaker. ECG gives information on the proper functionality of the device and could help to evaluate the reaction of the heart. Beyond the challenges related to the diversity of ECG arrhythmias and pace pulses, the existence of electromyogram (EMG) noise could cause serious problems for the correct detection of pace pulses. This study reveals the potential of a methodology based on Stockwell transformation (S-transform), subsequent Shannon energy calculation and a threshold-based rule for pace artifact detection in a single-lead ECG corrupted with EMG noise. The design, validation and test are performed on a large, publicly available artificial database acquired with high amplitude and time resolution. It includes various combinations of ECG arrhythmias and pace pulses with different amplitudes, rising edges and total pulse durations, as well as timing that corresponds to different pacemaker modes. The training was done over 312 (ECG + EMG) signals. The method was validated on 390 clean ECGs and independently tested on 312 (ECG + EMG) and 390 clean ECGs. The achieved accuracy over the test dataset was Se = 100%, PPV = 98.0% for ECG corrupted by EMG artifacts and Se = 99.9%, PPV = 98.3% for clean ECG signals. This shows that, despite EMG artifacts, the S-transform could distinctly localize the pace pulse positions and, together with the applied ShE, could provide precise pace pulses detection in the time domain.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: ECG Monitoring and Analysis · Muscle activation and electromyography studies · EEG and Brain-Computer Interfaces
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Institute of Biophysics and Biomedical Engineering, Bulgarian Academy of Sciences, Acad. G. Bonchev Str. Bl 105, 1113 Sofia, Bulgaria
Department of Electronics, Faculty of Electronic Engineering and Technologies, Technical University of Sofia, 8 Kliment Ohridski Blvd., 1000 Sofia, Bulgaria
Funding
Bulgarian National Science Fund
References (32)
Show source-provided references
1. Andreozzi, E., Gargiulo, G., Fratini, A., Esposito, D., and Bifulco, P. (2018). A Contactless Sensor for Pacemaker Pulse Detection: Design Hints and Performance Assessment. Sensors, 18.10.3390/s18082715 ↗
2. The presence of pacing artifacts may impede diagnosis of ventricular fibrillation during cardiac arrest10.1016/j.resuscitation.2014.04.034 ↗
3. Detecting and distinguishing cardiac pacing artifacts
4. A review on pacemakers: Device types, operating modes and pacing pulses. Problems related to the pacing pulses detection
5. Rose, W. (2020, September 02). Electromyogram analysis. Available online: http://www1.udel.edu/biology/rosewc/kaap686/notes/EMG%20analysis.pdf.
6. Calabria, T. (2020, September 02). Hardware Pace Using Slope Detection: Texas Instruments Precision Designs: Verified Design. Available online: https://www.ti.com/lit/ug/slau511/slau511.pdf?ts%20=%201603612787115&ref_url%20=%20https%253A%252F%252F%20www.google.com%252F.
7. High resolution electrocardiography optimized for recording pulses from electronic pacemakers: Evaluation of a new pacemaker sensing system
8. Performance study of digital pacer spike detection as sampling rate changes
9. A simple, wide bandwidth, biopotential amplifier to record pacemaker pulse waveform10.2147/MDER.S97902 ↗
10. The relationship between programmed pacemaker pulse amplitude and he surface electrocardiogram recorded amplitude: Application of a new high-bandwidth electrocardiogram system10.1016/j.jelectrocard.2008.06.023 ↗
11. Observations of pacemaker pulses in high-bandwidth electrocardiograms and Dower-estimated vectorcardiograms10.1016/j.jelectrocard.2010.09.014 ↗
12. Herleikson, E. (1997). ECG Pace Pulse Detection and Processing. (5682902), U.S. Patent.
13. Polpetta, A., and Banelli, P. (2008, January 20–25). Fully digital pacemaker detection in ECG signals using a non-linear filtering approach. Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine & Biology Society, EMBS, Vancouver, BC, Canada.10.1109/IEMBS.2008.4650437 ↗
14. Real-time detection of pace pulses in a single lead ECG
15. Kalinichenko, A., Chaykovskaya, M., and Danilova, D. (2016). Paced ECG Analysis in Mobile Cardiac Monitor. Comput. Cardiol., 43.10.22489/CinC.2016.190-202 ↗
16. Precise Pacing Artefact Detection
17. Vector-based pacemaker pulse detection algorithm for the surface ECG
18. Nallathambi, G., Selvaraj, N., and Rajbhandary, P. (2020, January 20–24). An innovative hybrid approach for detection of pacemaker pulses at low sampling frequency. Proceedings of the 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Montreal, QC, Canada.10.1109/EMBC44109.2020.9176390 ↗
19. American National Standard (2007). Diagnostic Electrocardiographic Devices, American National Standards Institute. ANSI/AAMI EC11.
20. American National Standard (2011). Medical Electrical Equipment—Part 2–27: Particular Requirements for the Basic Safety and Essential Performance of Electrocardiographic Monitoring Equipment, American National Standards Institute. ANSI/AAMI/IEC 60601-2-27.
21. Performance Challenges in ECG Pacemaker Pulse Detection Systems
22. Jekova, I., Iliev, I., Tsibulko, V., and Tabakov, S. (2018, January 13–15). Algorithm for pace pulses detection in a single lead ECG: Performance in case of EMG artifacts. Proceedings of the IEEE XXVII International Scientific Conference Electronics–ET2018, Sozopol, Bulgaria.10.1109/ET.2018.8549603 ↗
23. Locally-adaptive Myriad filtration of one-dimensional complex signal10.7546/ijba.2018.22.3.275-296 ↗
24. Pseudo real-time low-pass filter in ECG, self-adjustable to the frequency spectra of the waves10.1007/s11517-017-1625-y ↗
25. Separation of the electromyographic from the electrocardiographic signals and vice versa. A topical review of the dynamic procedure10.7546/ijba.2020.24.3.000744 ↗
26. A novel method based on Adaptive Periodic Segment Matrix and Singular Value Decomposition for removing EMG artifact in ECG signal10.1016/j.bspc.2020.102060 ↗
27. Surface electromyography low-frequency content: Assessment in isometric conditions after electrocardiogram cancellation by the Segmented-Beat Modulation Method10.1016/j.imu.2018.10.006 ↗
28. Localization of the complex spectrum: The S transform10.1109/78.492555 ↗
29. ECG signal analysis using modified S-transform
30. Shannon’s Energy Based Algorithm in ECG Signal Processing10.1155/2017/8081361 ↗
31. QRS detection using S-Transform and Shannon energy10.1016/j.cmpb.2014.04.008 ↗
32. ECG Database Applicable for Development and Testing of Pace Detection Algorithms