journal article·2022-04-06Open access

Clinical Mass Spectrometry in Immunosuppressant Analysis: Toward a Full Automation?

Chiara Fania, Marco Bagnati, Marina Albertario, Carlotta Ferraris, Marta Lamonaca, Umberto Dianzani

Applied Sciences· Volume 12· Issue 7

Abstract

The analysis of immunosuppressive drugs allows the physician to monitor, and eventually correct, immunosuppressive therapy. The panel of molecules under evaluation includes cyclosporine A (CsA), tacrolimus, sirolimus, and everolimus. Initially, assays were performed by immunometric methods, but in the past few years this methodology has been largely superseded by a more accurate and specific technique, liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS), which is now considered the “gold standard” for immunosuppressant analysis. Both LC-MS/MS and often also immunoassays require a preanalytical manual sample preparation, which involves time-consuming sequential operations whose traceability is often hampered and adds up to the probability of gross errors. The aim of this work was to compare an “open” LC-MS/MS with a fully automated system, consisting of LC instrumentation combined with a triple quadrupole MS, named Thermo ScientificTM CascadionTM SM Clinical Analyzer (Cascadion). Such automated systems suit the requirements of the reference method and are designed to completely eliminate all of the manual procedures. More than 2000 immunosuppressant samples were analyzed both with the open LC-MS/MS and with Cascadion. Statistics allowed the evaluation of linearity, intra- and inter-assay CV%, bias %, limit of detection and of quantitation, and Passing–Bablok and Bland–Altman plots. Results indicated a good correlation between the two methods. In both cases, methods confirmed their suitability for diagnostic settings. Cascadion could provide support when the presence of specialized personnel is lacking, and/or when great productivity and continuous workflow are required.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Renal Transplantation Outcomes and Treatments · Antibiotics Pharmacokinetics and Efficacy · Pneumocystis jirovecii pneumonia detection and treatment

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Clinical Chemistry Laboratory, Maggiore della Carità University Hospital, 28100 Novara, Italy

IRCAD & Department of Health Sciences, Amedeo Avogadro University of Eastern Piedmont, 28100 Novara, Italy

References (17)

Show source-provided references
1. (2022, April 01). NIH National Cancer Institute, Available online: www.cancer.gov.
2. Modern immunosuppression
3. Clinically unexpected cyclosporine levels using the ACMIA method on the RXL dimension analyser10.1093/ndt/gfr002 ↗
4. Increased cyclosporine concentrations in the absence of cyclosporine administration10.1373/clinchem.2010.148718 ↗
5. Therapeutic drug monitoring of immunosuppressants by liquid chromatography-mass spectrometry10.1016/j.cca.2015.12.027 ↗
6. False tacrolimus concentrations measured by antibody conjugated magnetic immunoassay in liver transplant patient: 2 case reports and literature review
7. Mass spectrometry and total laboratory automation: Opportunities and drawbacks10.1515/cclm-2019-0723 ↗
8. Impact of automation on mass spectrometry10.1016/j.cca.2015.08.027 ↗
9. Distribution and protein binding of FK506, a potent immunosuppressive macrolide lactone, in human blood and its uptake by erythrocytes10.1111/j.2042-7158.1994.tb03752.x ↗
10. Binding of cyclosporine by erythrocytes: Influence on cell shape and deformability10.1111/j.1365-2362.1993.tb00758.x ↗
11. After another decade: LC-MS/MS became routine in clinical diagnostics10.1016/j.clinbiochem.2020.03.004 ↗
12. Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyser suggests a new era in laboratory medicine
13. Thermo Scientific (2020). Maximizing Accuracy with LC-MS/MS Immunosuppressant Drugs Assay, Thermo Fisher Scientific Inc.. Available online: www.thermofisher.com/diagnostic-education/dam/clinical-mass-spectrometry/global/resources/library/assets/D20230-Cascadion-SM-ISD-TechSpecs-EN.pdf.
14. Thermo Scientific (2020). The Accuracy of LC-MS/MS Technology with the Convenience of Automation, Thermo Fisher Scientific Inc.. Available online: www.thermofisher.com/diagnostic-education/dam/clinical-mass-spectrometry/global/resources/library/assets/D19899-Cascadion-SM-Clinical-Analyzer-Brochure-US.pdf.
15. Thermo Scientific (2020). Fully Automated, Random Access Analyzer for In Vitro Diagnostic Use, Thermo Fisher Scientific Inc.. Available online: https://www.thermofisher.com/diagnostic-education/dam/clinical-mass-spectrometry/global/resources/library/assets/D15952-Cascadion-SM-Clinical-Analyzer-Technical-Specifications-EN.pdf.
16. Immunosuppressive regimens for adult liver transplant recipients in real-life practice: Consensus recommendations from an Italian Working Group10.1007/s12072-020-10091-5 ↗
17. Informed consent and ethics committee approval in laboratory medicine10.11613/BM.2018.030201 ↗