Improving Prediction of the Potential Distribution Induced by Cylindrical Electrodes within a Homogeneous Rectangular Grid during Irreversible Electroporation
Source abstract
Abstract
Background: Irreversible electroporation (IRE) is an ablation technique based on the application of short, high-voltage pulses between needle electrodes (diameter: ~1.0 × 10−3 m). A Finite Difference-based software simulating IRE treatment generally uses rectangular grids, yielding discretization issues when modeling cylindrical electrodes and potentially affecting the validity of treatment planning simulations. Aim: Develop an Electric-Potential Estimation (EPE) method for accurate prediction of the electric-potential distribution in the vicinity of cylindrical electrodes. Methods: The electric-potential values in the voxels neighboring the cylindrical electrode voxels were corrected based on analytical solutions derived for coaxial/cylindrical electrodes. Simulations at varying grid resolutions were validated using analytical models. Low-resolution heterogeneous simulations at 2.0 × 10−3 m excluding/including EPE were compared with high-resolution results at 0.25 × 10−3 m. Results: EPE significantly reduced maximal errors compared to analytical results for the electric-potential distributions (26.6–71.8%→0.4%) and for the electrical resistance (30%→1–6%) at 3.0 × 10−3 m voxel-size. EPE significantly improved the mean-deviation (43.1–52.8%→13.0–24.3%) and the calculation-time gain (>15,000×) of low-resolution compared to high-resolution heterogeneous simulations. Conclusions: EPE can accurately predict the potential distribution of neighboring cylindrical electrodes, regardless of size, position, and orientation in a rectangular grid. The simulation time of treatment planning can therefore be shortened by using large voxel-sized models without affecting accuracy of the electric-field distribution, enabling real-time clinical IRE treatment planning.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Microbial Inactivation Methods · Microfluidic and Bio-sensing Technologies · Ultrasound and Hyperthermia Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Radiation Oncology, Amsterdam UMC, University of Amsterdam, Cancer Center Amsterdam, 1105 AZ Amsterdam, The Netherlands
Department of Surgery, Amsterdam UMC, University of Amsterdam, Cancer Center Amsterdam, 1105 AZ Amsterdam, The Netherlands
Department of Radiology Amsterdam UMC, University of Amsterdam, Cancer Center Amsterdam, 1105 AZ Amsterdam, The Netherlands
Funding
Dutch Cancer Society
References (42)
Show source-provided references
1. High-Voltage Electrical Pulses in Oncology: Irreversible Electroporation, Electrochemotherapy, Gene Electrotransfer, Electrofusion, and Electroimmunotherapy10.1148/radiol.2020192190 ↗
2. Vogel, J.A., van Veldhuisen, E., Agnass, P., Crezee, J., Dijk, F., Verheij, J., van Gulik, T.M., Meijerink, M.R., Vroomen, L.G., and van Lienden, K.P. (2016). Time-Dependent Impact of Irreversible Electroporation on Pancreas, Liver, Blood Vessels and Nerves: A Systematic Review of Experimental Studies. PLoS ONE, 11.10.1371/journal.pone.0166987 ↗
3. Heating technology for malignant tumors: A review10.1080/02656736.2020.1779357 ↗
4. In vivo evaluation of bronchial injury of irreversible electroporation in a porcine lung ablation model by using laboratory, pathological, and CT findings
5. Thermodynamic profiling during irreversible electroporation in porcine liver and pancreas: A case study series
6. Membrane Electroporation and Electropermeabilization: Mechanisms and Models10.1146/annurev-biophys-052118-115451 ↗
7. Dynamics of Cell Death After Conventional IRE and H-FIRE Treatments10.1007/s10439-020-02462-8 ↗
8. Experimental Characterization and Numerical Modeling of Tissue Electrical Conductivity during Pulsed Electric Fields for Irreversible Electroporation Treatment Planning10.1109/TBME.2012.2182994 ↗
9. Irreversible Electroporation Ablation: Is All the Damage Nonthermal?10.1148/radiol.12120609 ↗
10. Irreversible Electroporation: Treatment Effect Is Susceptible to Local Environment and Tissue Properties10.1148/radiol.13122590 ↗
11. Agnass, P., Rodermond, H.M., Zweije, R., Sijbrands, J., Vogel, J., van Lienden, K.P., van Gulik, T.M., van Veldhuisen, E., Franken, N.A.P., and Oei, A.L. (2020). HyCHEED System for Maintaining Stable Temperature Control during Preclinical Irreversible Electroporation Experiments at Clinically Relevant Temperature and Pulse Settings. Sensors, 20.10.3390/s20216227 ↗
12. Mathematical modeling of the thermal effects of irreversible electroporation for in vitro, in vivo, and clinical use: A systematic review10.1080/02656736.2020.1753828 ↗
13. OpenEP: An open-source simulator for electroporation-based tumor treatments10.1038/s41598-020-79858-y ↗
14. EView: An electric field visualization web platform for electroporation-based therapies10.1016/j.cmpb.2020.105682 ↗
15. Numerical modelling challenges for clinical electroporation ablation technique of liver tumors10.1051/mmnp/2019037 ↗
16. Comsol, Inc. (2016, January 06). Finite Element Mesh Refinement. Available online: https://www.comsol.com/multiphysics/mesh-refinement.
17. First Human Trial of High-Frequency Irreversible Electroporation Therapy for Prostate Cancer10.1177/1533033818789692 ↗
18. A 2D Capacitance Problem
19. A three-dimensional iterative scheme for an electromagnetic capacitive applicator10.1109/10.102810 ↗
20. Intrinsic dependencies of CT radiomic features on voxel size and number of gray levels10.1002/mp.12123 ↗
21. Simulation of radiofrequency ablation in real human anatomy10.3109/02656736.2014.968639 ↗
22. Sequential finite element model of tissue electropermeabilization10.1109/TBME.2005.845212 ↗
23. Intracranial Nonthermal Irreversible Electroporation: In Vivo Analysis10.1007/s00232-010-9284-z ↗
24. Feasibility of on-line temperature-based hyperthermia treatment planning to improve tumour temperatures during locoregional hyperthermia10.1080/02656736.2017.1400120 ↗
25. Planning, optimisation and evaluation of hyperthermia treatments10.1080/02656736.2017.1295323 ↗
26. Predictive value of simulated SAR and temperature for changes in measured temperature after phase-amplitude steering during locoregional hyperthermia treatments10.1080/02656736.2018.1500720 ↗
27. A 3-D SAR model for current source interstitial hyperthermia10.1109/10.536905 ↗
28. Surface charges and electric field in a two-wire resistive transmission line
29. Incropera, F.P., Dewitt, D.P., Bergman, T.L., and Lavine, A.S. (2007). The Conduction Shape Factor and the Dimensionless Conduction Heat Rate. Fundamentals of Heat and Mass Transfer, John Wiley & Sons, Inc.. [6th ed.].
30. Time-Dependent Impact of Irreversible Electroporation on Pathology and Ablation Size in the Porcine Liver: A 24-Hour Experimental Study10.1177/1533033819876899 ↗
31. Hasgall, P.A., Di Gennaro, F., Baumgartner, C., Neufeld, E., Lloyd, B., Gosselin, M.C., Payne, D., Klingenböck, A., and Kuster, N. (2018, May 15). IT’IS Database for Thermal and Electromagnetic Parameters of Biological Tissues. Available online: https://itis.swiss/virtual-population/tissue-properties/database/database-summary/.
32. Irreversible electroporation for the ablation of pancreatic malignancies: A patient-specific methodology10.1002/jso.24566 ↗
33. Ablation of Locally Advanced Pancreatic Cancer with Percutaneous Irreversible Electroporation: Results of the Phase I/II PANFIRE Study10.1148/radiol.2016152835 ↗
34. A study of the sink effect by blood vessels in radiofrequency ablation10.1016/j.compbiomed.2014.12.014 ↗
35. Predictive therapeutic planning for irreversible electroporation treatment of spontaneous malignant glioma10.1002/mp.12401 ↗
36. Tissue heating during tumor ablation with irreversible electroporation
37. Effects of internal electrode cooling on irreversible electroporation using a perfused organ model
38. The Feasibility of Irreversible Electroporation for the Treatment of Breast Cancer and Other Heterogeneous Systems10.1007/s10439-009-9796-9 ↗
39. Retrospective study for validation and improvement of numerical treatment planning of irreversible electroporation ablation for treatment of liver tumors10.1109/TBME.2021.3075772 ↗
40. Numerical workflow of irreversible electroporation for deep-seated tumor10.1088/1361-6560/ab00c4 ↗
41. Tissue heterogeneity in structure and conductivity contribute to cell survival during irreversible electroporation ablation by “electric field sinks”10.1038/srep08485 ↗
42. Scheffer, H.J., Vogel, J.A., Bos, W.V.D., Neal, R.E., van Lienden, K.P., Besselink, M.G.H., van Gemert, M.J.C., van der Geld, C.W.M., Meijerink, M.R., and Klaessens, J.H. (2016). The Influence of a Metal Stent on the Distribution of Thermal Energy during Irreversible Electroporation. PLoS ONE, 11.10.1371/journal.pone.0148457 ↗