Signal Acquisition and Algorithm Design for Bioimpedance-Based Heart Rate Estimation from the Wrist
Source abstract
Abstract
Background: Heart rate (HR) is a critical biomarker that provides insights into overall health, stress levels, and the autonomic nervous system. Pulse wave signals contain valuable information about the cardiovascular system and heart status. However, signal acquisition in wearables poses challenges, particularly when using electrical sensors, due to factors like the distance from the heart, body movement, and suboptimal electrode placement. Methods: Electrical bioimpedance (EBI) measurements using bipolar and tetrapolar electrode systems were employed for pulse wave signal acquisition from the wrist in both perpendicular and distal configurations. Signal preprocessing techniques, including baseline removal via Hankel matrix methods, normalization, cross-correlation, and peak detection, were applied to improve signal quality. This study describes the combination of sensor-level signal acquisition and processing for accurate wearable HR estimation. Results: The bipolar system was shown to produce larger ΔZ(t), while the tetrapolar system demonstrated higher sensitivity. Distal placement of the electrodes yielded greater ΔZ(t) (up to 0.231 Ω) when targeting both wrist arteries. Bandpass filtering resulted in a better signal-to-noise ratio (SNR), achieving 3.6 dB for the best bipolar setup and 4.8 dB for the tetrapolar setup, compared to 2.6 and 3.3 dB SNR, respectively, with the Savitzky–Golay filter. The custom HR estimation algorithm presented in this paper demonstrated improved accuracy over a reference method, achieving an average error of 1.8 beats per minute for the best bipolar setup, with a mean absolute percentage error (MAPE) of 8%. Conclusions: The analysis supports the feasibility of using bipolar electrode setups on the wrist and highlights the importance of electrode positioning relative to the arteries. The proposed signal processing method, featuring a preprocessing pipeline and HR estimation algorithm, provides a proof-of-concept demonstration for HR estimation from EBI signals acquired at the wrist.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Non-Invasive Vital Sign Monitoring · Heart Rate Variability and Autonomic Control · ECG Monitoring and Analysis
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Institute of Electronics and Computer Science, Dzerbenes Street 14, LV-1006 Riga, Latvia
Faculty of Computing, University of Latvia, Raiņa Bulvāris 19, LV-1586 Riga, Latvia
Thomas Johann Seebeck Department of Electronics, Tallinn University of Technology, Ehitajate tee 5, 19086 Tallinn, Estonia
Funding
Estonian Research Council · European Union
References (42)
Show source-provided references
1. Li, K., Cardoso, C., Moctezuma-Ramirez, A., Elgalad, A., and Perin, E. (2023). Heart Rate Variability Measurement through a Smart Wearable Device: Another Breakthrough for Personal Health Monitoring?. Int. J. Environ. Res. Public Health, 20.10.20944/preprints202308.0732.v1 ↗
2. The physiological basis and measurement of heart rate variability in humans10.1186/s40101-016-0113-7 ↗
3. Arrhythmias and Heart Rate: Mechanisms and Significance of a Relationship
4. A Review of Methods for Non-Invasive Heart Rate Measurement on Wrist10.1016/j.irbm.2020.04.001 ↗
5. Fundamentals, Recent Advances, and Future Challenges in Bioimpedance Devices for Healthcare Applications
6. Martinsen, O.G., and Grimnes, S. (2014). Bioimpedance and Bioelectricity Basics, Academic Press. [3rd ed.].
7. Fundamentals of impedance cardiography10.1109/51.32403 ↗
8. Escalona, O., Cullen, N., Weli, I., McCallan, N., Ng, K.Y., and Finlay, D. (2023). Robust Arm Impedocardiography Signal Quality Enhancement Using Recursive Signal Averaging and Multi-Stage Wavelet Denoising Methods for Long-Term Cardiac Contractility Monitoring Armbands. Sensors, 23.10.3390/s23135892 ↗
9. Metshein, M., Gautier, A., Larras, B., Frappe, A., John, D., Cardiff, B., Annus, P., Land, R., and Martens, O. (2021, January 1–5). Study of Electrode Locations for Joint Acquisition of Impedance- and Electro-cardiography Signals. Proceedings of the 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Guadalajara, Mexico.10.1109/EMBC46164.2021.9629504 ↗
10. Pulse rate variability in cardiovascular health: A review on its applications and relationship with heart rate variability10.1088/1361-6579/ab998c ↗
11. Landaeta, R.G., Casas, O., and Pallas-Areny, R. (September, January 30). Heart Rate Detection from Plantar Bioimpedance Measurements. Proceedings of the 2006 International Conference of the IEEE Engineering in Medicine and Biology Society, New York, NY, USA.10.1109/IEMBS.2006.4398604 ↗
12. Review on heart-rate estimation from photoplethysmography and accelerometer signals during physical exercise10.1007/s41745-017-0037-1 ↗
13. Continuous cuffless blood pressure monitoring with a wearable ring bioimpedance device10.1038/s41746-023-00796-w ↗
14. Bioimpedance Sensors: A Tutorial10.1109/JSEN.2021.3110283 ↗
15. Metshein, M., Pesti, K., Lapsa, D., Annus, P., Janeliukstis, R., Elsts, A., and Martens, O. (2024, January 20–23). Evaluation of Two-Electrode System Configurations for Forearm Arteries Bioimpedance Measurement. Proceedings of the 2024 IEEE International Instrumentation and Measurement Technology Conference (I2MTC), Glasgow, UK.10.1109/I2MTC60896.2024.10561178 ↗
16. Yılmaz, K., Adeyemi, A., Antink, C.H., and Vehkaoja, A. (November, January 30). Comparison of Electrode Configurations for Impedance Plethysmography Based Heart Rate Estimation at the Forearm. Proceedings of the 2022 IEEE Sensors, Dallas, TX, USA.10.1109/SENSORS52175.2022.9967149 ↗
17. Lapsa, D., Janeliukštis, R., and Elsts, A. (2022, January 5–7). Electrode Comparison for Heart Rate Detection via Bioimpedance Measurements. Proceedings of the 2022 Workshop on Microwave Theory and Techniques in Wireless Communications (MTTW), Riga, Latvia.10.1109/MTTW56973.2022.9942569 ↗
18. Priidel, E., Annus, P., Metshein, M., Land, R., Märtens, O., and Min, M. (2018, January 8–10). Lock-in integration for detection of tiny bioimpedance variations. Proceedings of the 2018 16th Biennial Baltic Electronics Conference (BEC), Tallinn, Estonia.10.1109/BEC.2018.8600965 ↗
19. Jung, M.H., Namkoong, K., Lee, Y., Jun Koh, Y., Eom, K., Jang, H., Jung, W., Bae, J., and Park, J. (2021). Wrist-wearable bioelectrical impedance analyzer with miniature electrodes for daily obesity management. Sci. Rep., 11.10.1038/s41598-020-79667-3 ↗
20. Kekonen, A., Bergelin, M., Johansson, M., Kumar Joon, N., Bobacka, J., and Viik, J. (2019). Bioimpedance sensor array for long-term monitoring of wound healing from beneath the primary dressings and controlled formation of H2O2 using low-intensity direct current. Sensors, 19.10.3390/s19112505 ↗
21. Alsanie, S., Lim, S., and Wootton, S. (2022). Detecting low-intake dehydration using bioelectrical impedance analysis in older adults in acute care settings: A systematic review. BMC Geriatr., 22.10.1186/s12877-022-03589-0 ↗
22. Hoog Antink, C., Schulz, F., Leonhardt, S., and Walter, M. (2018). Motion Artifact Quantification and Sensor Fusion for Unobtrusive Health Monitoring. Sensors, 18.10.3390/s18010038 ↗
23. Lapsa, D., Janeliukstis, R., and Elsts, A. (2023). Adaptive Signal-to-Noise Ratio Indicator for Wearable Bioimpedance Monitoring. Sensors, 23.10.3390/s23208532 ↗
24. Investigating the possible effect of electrode support structure on motion artifact in wearable bioelectric signal monitoring10.1186/s12938-015-0044-2 ↗
25. Combining Bioimpedance and EMG Measurements for Reliable Muscle Contraction Detection10.1109/JSEN.2019.2936171 ↗
26. Effect of pressure and padding on motion artifact of textile electrodes10.1186/1475-925X-12-26 ↗
27. Dang, T.H., Jang, G.Y., Lee, K., and Oh, T.I. (2023). Motion Artifacts Reduction for Noninvasive Hemodynamic Monitoring of Conscious Patients Using Electrical Impedance Tomography: A Preliminary Study. Sensors, 23.10.20944/preprints202304.1076.v1 ↗
28. Metshein, M., Abdullayev, A., Gautier, A., Larras, B., Frappe, A., Cardiff, B., Annus, P., Land, R., and Märtens, O. (2023). Sensor-Location-Specific Joint Acquisition of Peripheral Artery Bioimpedance and Photoplethysmogram for Wearable Applications. Sensors, 23.10.3390/s23167111 ↗
29. Non-Invasive Cardiac and Respiratory Activity Assessment From Various Human Body Locations Using Bioimpedance10.1109/OJEMB.2021.3085482 ↗
30. Wang, T.W., Chu, H.W., Chou, L., Sung, Y.L., Shih, Y.T., Hsu, P.C., Cheng, H.M., and Lin, S.F. (2021). Bio-Impedance Measurement Optimization for High-Resolution Carotid Pulse Sensing. Sensors, 21.10.3390/s21051600 ↗
31. Ren, B., Wang, Z., Ma, K., Zhou, Y., and Liu, M. (2023). An Improved Method of Heart Rate Extraction Algorithm Based on Photoplethysmography for Sports Bracelet. Information, 14.10.3390/info14050297 ↗
32. Radial artery pulse wave estimation by compressed sensing measurements of wrist bio-impedance10.1016/j.measurement.2023.113174 ↗
33. Krivošei, A., Min, M., Annus, P., Märtens, O., Metshein, M., Lotamõis, K., and Rist, M. (2022, January 22–24). Hankel Matrix Based Algorithm for Cardiac Pulse Wave Base and Peak Lines Correction. Proceedings of the 2022 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Messina, Italy.10.1109/MeMeA54994.2022.9856564 ↗
34. Wójcikowski, M., and Pankiewicz, B. (2020). Photoplethysmographic Time-Domain Heart Rate Measurement Algorithm for Resource-Constrained Wearable Devices and its Implementation. Sensors, 20.10.3390/s20061783 ↗
35. Han, J., Ou, W., Xiong, J., and Feng, S. (2022). Remote Heart Rate Estimation by Pulse Signal Reconstruction Based on Structural Sparse Representation. Electronics, 11.10.3390/electronics11223738 ↗
36. He, Z. (2020). The control mechanisms of heart rate dynamics in a new heart rate nonlinear time series model. Sci. Rep., 10.10.1038/s41598-020-61562-6 ↗
37. Alharbi, Y., Alshrouf, A., and Mansouri, S. (2021, January 25–27). Heart Rate Monitoring Using Electrical Impedance. Proceedings of the 2021 Seventh International conference on Bio Signals, Images, and Instrumentation (ICBSII), Chennai, India.10.1109/ICBSII51839.2021.9445147 ↗
38. Zurich Instruments (2023). MFIA User Manual, Zurich Instruments.
39. A Novel Method to Estimate Body Capacitance to Ground at Mid Frequencies10.1109/TIM.2013.2258240 ↗
40. (2024, August 11). Matlab Filter Design. Available online: https://se.mathworks.com/help/signal/ref/filtfilt.html.
41. Lee, J., Kim, S., and Lee, D. (1998, January 1). Fast cross-correlation method for real time detection of fetal heart rate. Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol. 20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286), Hong Kong, China.
42. Detrended cross-correlation analysis for non-stationary time series with periodic trends10.1209/0295-5075/94/18007 ↗