Optimization of Image Quality in Digital Mammography with the Response of a Selenium Detector by Monte Carlo Simulation

Marek Szewczuk, Adam Konefał

Applied Sciences· Volume 13· Issue 1

Abstract

Mammography machines must meet high requirements to ensure the quality of the generated images. On the other hand, due to the use of ionizing radiation, there is a need to minimize the dose received by patients. To optimize both of these parameters (dose and image quality), the response characteristics of image detectors and, depending on the composition of the breasts, the physical contrast of the examined structures should be considered. This study aimed to determine the optimal voltage values for a given breast thickness during imaging with the use of a selenium image detector. Analysis was carried out using the Monte Carlo simulation method with the GEANT4 code. Our results reveal that the combination of Mo anode together with Mo filtration (the system recommended in analog mammography) was the least favorable combination among those used in digital mammography machines with a selenium detector. Moreover, the use of Rh filtration instead of Mo was advantageous regardless of the thickness of the breast and resulted in a significant improvement in image quality with the same dose absorbed in the breast. The most advantageous solution was found to be an X-ray tube with a W anode. The highest values of the image quality-to-dose ratio were observed for breasts with dimensions ranging from 53 mm to 60 mm in thickness. Lower image quality was observed for breasts with smaller dimensions due to high breast glandularity, resulting in the deterioration of the physical contrast.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Digital Radiography and Breast Imaging · AI in cancer detection · Medical Imaging Techniques and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Katowice Cancer Center, 40-074 Katowice, Poland

Institute of Physics, University of Silesia in Katowice, 40-007 Katowice, Poland

References (36)

Show source-provided references
1. WHO (2022, November 20). Available online: https://www.who.int/news-room/fact-sheets/detail/breast-cancer.
2. Cancer Genome Landscapes10.1126/science.1235122 ↗
3. Impact of mammography screening programmes on breast cancer mortality in Switzerland, a country with different regional screening policies10.1136/bmjopen-2017-017806 ↗
4. Iwamoto, Y., Kaucher, S., Lorenz, E., Barnighausen, T., and Winkler, V. (2019). Development of breast cancer mortalityconsidering the implementation of mammography screening programs-acomparison of western European countries. BMC Public Health, 19.10.1186/s12889-019-7166-6 ↗
5. Effect of screening and adjuvant therapy on mortality from breast cancer10.1056/NEJMoa050518 ↗
6. The Swedish Organised Service Screening Evaluation Group (2006). Reduction in breast cancer mortality from organized service screening with mammography: 1. Further confirmation with extended data. Cancer Epidemiol. Biomark. Prev., 15, 45–51.10.1158/1055-9965.EPI-05-0349 ↗
7. Radiation effects on breast cancer risk: A pooled analysis of eight cohorts10.1667/0033-7587(2002)158[0220:REOBCR]2.0.CO;2 ↗
8. Bourguignon MH. Breast cancer induced by X-ray mammography screening? A review based on recent understanding of low-dose radiobiology10.1159/000442442 ↗
9. Organ doses and cancer risk assessment in patients exposed to high doses from recurrent CT exams10.1016/j.ejrad.2022.110224 ↗
10. International Atomic Energy Agency, IAEA (2022, November 01). Human Health Series 17. Quality Assurance Programme for Digital Mammography, Vienna. Available online: https://www.iaea.org/publications/8560/quality-assurance-programme-for-digital-mammography.
11. Radiation dose optimization in routine computed tomography: A study of feasibility in a University Hospital10.1590/S0100-39842010000400008 ↗
12. Oduko, J.M., Young, K.C., Burch, A., Castellano, E., Kulama, E., Lawinski, C., and Marshall, N. (2009). Review of Measurements on Full Field Digital Mammography Systems—NHSBSP Equipment Report 0901, NHS Cancer Screening Programmes.
13. Young, K.C., Oduko, J.M., and Woolley, L. (2007). Technical Evaluation of the Hologic Selenia Full Field Digital Mammography System—NHSBSP Equipment Report 0701, NHS Cancer Screening Programmes.
14. X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study10.1118/1.2351951 ↗
15. Organ doses from CT localizer radiographs: Development, validation, and application of a Monte Carlo estimation technique10.1002/mp.13781 ↗
16. Infuence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: A Monte Carlo study10.1259/bjr.73.874.11271898 ↗
17. Optimization of the exposure parameters in digital mammography for diverse glandularities using the contrast-detail metric10.1016/j.ejmp.2022.08.009 ↗
18. The myth of the 50-50 breast10.1118/1.3250863 ↗
19. Cullen, D.E. (2015). EPICS2014: Electron Photon Interaction Cross Sections, The Nuclear Energy Agency.
20. Validation of recent Geant4 physics models for application in carbon ion therapy10.1016/j.nimb.2010.04.008 ↗
21. Validation of the Geant4 electromagnetic photon cross sections for elements and compounds. Nucl. Instrum
22. GEANT4 simulation of electron energy deposition in extended media. Nucl. Instrum
23. Molybdenium, rhodium andtungsten anode spectral models using interpolationg polynomials with application to mammography10.1118/1.598100 ↗
24. Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose10.1088/0031-9155/35/9/002 ↗
25. Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol10.1088/0031-9155/45/11/308 ↗
26. Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols10.1088/0031-9155/54/14/002 ↗
27. Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography10.1088/0031-9155/50/24/013 ↗
28. Glandular breast dose for monoenergetic and high-energy X-raybeams: Monte Carlo assessment10.1148/radiology.213.1.r99oc3923 ↗
29. Homogeneous vs. patient specific breast models for Monte Carlo evaluation of mean glandular dose in mammography10.1016/j.ejmp.2018.04.392 ↗
30. Absorbed radiation dose in mammography10.1148/130.2.485 ↗
31. X-ray characterization of normal and neoplastic breast tissues10.1088/0031-9155/32/6/002 ↗
32. Ullman, G., Sandborg, M., Hunt, R., Dance, D.R., and Carlsson, G.A. (2003). Implementation of Pathologies in the Monte Carlo Model in Chest and Breast Imaging, Institutionen för Radiologi, Universitetet in Linköping. Report 94.
33. Perry, N., Broeders, M., de Wolf, C., Törnberg, S., Holland, R., and von Karsa, L. (2006). European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, Office for Official Publications of the European Communities. [4th ed.].
34. Comparison of depth-dose distributions of proton therapeutic beams calculated by means of logical detectors and ionization chamber modeled in Monte Carlo codes10.1016/j.nima.2016.04.090 ↗
35. X-ray detectors for digital radiography10.1088/0031-9155/42/1/001 ↗
36. Pruszyński, B. (2007). Diagnostic Imaging—Theoretical Basis and Research Methodology, Wydawnictwo Lekarskie PZWL. (In Polish).