Vibration Reduction in a Pulse Tube-Cooled High Purity Germanium Detector
Source abstract
Abstract
High-purity germanium (HPGe) is a vibration-sensitive, high-resolution detector. It has been found that microphonic noise induced by vibration and electromagnetic interference from EMC causes energy resolution degradation in HPGe detectors. In this paper, to enhance the energy resolution of a pulse tube-cooled high purity germanium detector, a passive vibration isolation system for the portable pulse tube cooler (PTC) was employed and investigated. In order to evaluate the performance of the vibration isolation system, micro-vibration tests of the compressor with a vibration isolation kit and detector mounting interface were performed. Meanwhile, detector energy resolution and detector signal noise were tested, respectively, under the PTC, with the isolation system switched to either an on or off condition. The test results indicate that the effectiveness of passive vibration isolation techniques is confirmed, and no degradation was induced by PTC vibration on detector energy resolution.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Thermodynamic Systems and Engines · Superconducting and THz Device Technology · Acoustic Wave Resonator Technologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Key Laboratory of Technology on Space Energy Conversion, Technical Institute of Physics and Chemistry, CAS, Beijing 100190, China
University of Chinese Academy of Sciences, Beijing 100049, China
Funding
Supported by National Key R&D Program of China
References (15)
Show source-provided references
1. Recent developments in the fabrication and operation of germanium detectors10.1146/annurev.nucl.56.080805.140525 ↗
2. New cooling methods for HPGE detectors and associated electronics10.1007/s10967-005-0684-y ↗
3. Pulse tube refrigeration for detectors10.1007/s10967-012-2002-9 ↗
4. A passive launch and on-orbit vibration isolation system for the spaceborne cryocooler10.1016/j.ast.2012.11.013 ↗
5. Vibration-free stirling cryocooler for high-definition microscopy10.1016/j.cryogenics.2009.08.002 ↗
6. Control of dynamic disturbances produced by a pulse tube refrigerator in a vibration-sensitive instrumentation10.1016/j.cryogenics.2008.08.010 ↗
7. Micro-jitter attenuation of spaceborne cooler by using a blade-type hyperelastic shape memory alloy passive isolator10.1016/j.cryogenics.2017.08.011 ↗
8. Viscoelastic multilayered blade-type passive vibration isolation system for a spaceborne cryogenic cooler10.1016/j.cryogenics.2019.102982 ↗
9. Ultra-low-vibration pulse-tube cryocooler system–cooling capacity and vibration10.1016/j.cryogenics.2008.04.001 ↗
10. Vibration suppression of advanced space cryocoolers: An overview
11. Yakovlev, O., Malgin, V., Gostilo, V., and Viba, Y. (2017, January 17–22). Constructive and Technological Aspects of the Development of Cryostats for HPGe Detectors with Electric Cooling. Proceedings of the European Planetary Science Congress, Riga, Latvia.
12. Kasahara, K., Yamamoto, Y., Uchiyama, T., Miyoki, S., Ohashi, M., Kuroda, K., Tomartu, T., Suzuki, T., and Shintomi, T. (August, January 31). Study of heat links for a cryogenic laser interferometric gravitational wave detector. Proceedings of the International Cosmic Ray Conference, Tsukuba, Japan.
13. A miniature compact HPGe gamma-spectrometer for space applications10.1088/1748-0221/12/05/P05017 ↗
14. Performance evaluation of spaceborne cryocooler micro-vibration isolation system employing pseudoelastic SMA mesh washer10.1016/j.cryogenics.2015.01.002 ↗
15. Low vibration cooling using a pulse tube cooler and cryostat for the GRAVITY beam combiner instrument at the VLTI