Modelling, Simulation and Performance Validation of the Pneumatic Actuation System of a Rehabilitation Device of the Human Hand Joints
Source abstract
Abstract
The passive mobilization of the hand joints by means of dedicated equipment accelerates patient recovery and decreases significantly the costs of therapy. For this reason, research and development of such equipment is essential. Important reductions in the development cycle duration of such equipment can be achieved by means of a specific technique known as Model-Based Design. Starting from these considerations, this paper puts forward a Model-Based Design approach to the study of a new concept of rehabilitation equipment of the hand joints actuated by a pneumatic muscle. The originality of the paper consists in the MATLAB-based rendering of the functional model of the rehabilitation equipment actuation system and in the presented simulation results. The purpose of this research was to obtain information concerning the behavior of the proposed system and to predict its performance prior to it being built physically. After simulation, the results are compared to the operational performance of the experimental model. The conclusion shows that the proposed operational model describes accurately the actual behavior of the system and can be used for future optimization of the rehabilitation equipment.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Prosthetics and Rehabilitation Robotics · Stroke Rehabilitation and Recovery · Muscle activation and electromyography studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Industrial Engineering and Management, Transilvania University of Brasov, 500036 Brașov, Romania
References (21)
Show source-provided references
1. Bulbul, M.K. (2015). Studiu Privind Incidenţa Sindromului de Tunel Carpian, Complicaţie Tardivă a Fracturilor Metafizo-Epifizare Distale Ale Antebraţului, în Corelaţie Cu Tratamentul Acestora. [Ph.D. Thesis, Oradea University].
2. Skirven, T.M., Osterman, A., Fedorczyk, J., Amadio, P., Felder, S., and Shin, E. (2020). Rehabilitation of the Hand and Upper Extremity, Elsevier. [7th ed.].
3. (2022, August 23). Kinetec Maestra Hand & Wrist Tabletop CPM. Available online: https://www.medcomgroup.com/kinetec-maestra-hand-wrist-tabletop-cpm/.
4. (2022, August 23). Kinetec—8091 Portable Hand CPM. Available online: https://www.medcomgroup.com/content/Kinetec%208091%20Portable%20Hand%20CPM%20Operations%20Manual.pdf.
5. Robot-Aided Neurorehabilitation: A Robot for Wrist Rehabilitation10.1109/TNSRE.2007.903899 ↗
6. (2022, August 23). QAL Medical W2 Wrist CPM. Available online: https://www.medcomgroup.com/qal-medical-w2-wrist-cpm/.
7. (2020, August 30). OrthoAgility 6000 Hand CPM Specification Brochure. Available online: https://pdf.medicalexpo.com/pdf/qal-medical/orthoagility-6000-hand-cpm-specification-brochure/93399-189906.html#open.
8. (2020, August 30). Waveflex CPM. Available online: https://www.remingtonmedical.com/product/waveflex-cpm/.
9. Pneumatic Muscle Actuated Wrist Rehabilitation Equipment Based on the Fin Ray Principle
10. Compliant Actuator Designs10.1109/MRA.2009.933629 ↗
11. (2020, August 31). The Hand Mentor Pro™. Available online: http://stage.motusnova.com/products/hand-mentor-pro/.
12. Andrikopoulos, G., Nikolakopoulos, G., and Manesis, S. (2015, January 8–11). Motion Control of a Novel Robotic Wrist Exoskeleton via Pneumatic Muscle Actuators. Proceedings of the 20th IEEE International Conference on Emerging Technologies and Factory Automation (ETFA), Luxembourg.10.1109/ETFA.2015.7301464 ↗
13. Meng, W., Sheng, B., Klinger, M., Liu, Q., Zhou, Z., and Xie, S.Q. (2015, January 7–11). Design and control of a robotic wrist orthosis for joint rehabilitation. Proceedings of the IEEE ASME International Conference on Advanced Intelligent Mechatronics, Busan, Republic of Korea.10.1109/AIM.2015.7222708 ↗
14. Filip, O., Deaconescu, A., and Deaconescu, T. (2021). Mechanical Design of a Bioinspired Compliant Robotic Wrist Rehabilitation Equipment. Appl. Sci., 11.10.3390/app11031246 ↗
15. (2022, August 28). Festo Fluidic Muscle DMSP/MAS. Available online: https://www.festo.com/rep/en_corp/assets/pdf/info_501_en.pdf.
16. Modeling and Simulation (2022, August 28). Wikipedia. Available online: https://en.wikipedia.org/wiki/Modeling_and_simulation.
17. Carson, J.S. (2005, January 4). Introduction to modeling and simulation. Proceedings of the 2005 Winter Simulation Conference, Orlando, FL, USA.
18. Marin, C., and Vasile, G. (2011). Modeling and Simulation Techniques in Mechanical Engineering (Tehnici de Modelare si Simulare in Ingineria Mecanică), Bibliotheca. (In Romanian).
19. Simulink (2022, August 29). MathWorks. Available online: https://www.mathworks.com/products/simulink.html.
20. (2022, August 29). Simulink Basics Tutorial. Available online: https://ctms.engin.umich.edu/CTMS/index.php?aux=Basics_Simulink.
21. Measurement and modeling of McKibben Pneumatic Artificial Muscles10.1109/70.481753 ↗