What size of cluster is most appropriate for SIMS?
Source abstract
Abstract
Abstract Secondary ion emission with large gas cluster ion is reviewed from the point of view of secondary ion mass spectroscopy (SIMS). We have proposed to use large cluster ions to realize fragment-free ionization for SIMS analysis for various organic materials, such as amino acids and peptides. When large cluster ions with optimized size and energy were incident on biomolecular samples, the relative yields of the fragment ions decreased drastically with the velocity of incident cluster ions. Molecular depth profiling capability is also demonstrated by using large gas cluster ions as the primary ion for SIMS.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Ion-surface interactions and analysis · Mass Spectrometry Techniques and Applications · Analytical chemistry methods development
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (32)
Show source-provided references
1. Yamada · Mater. Sci. Eng. · 2001 · R34 · 23110.1016/S0927-796X(01)00034-1 ↗
2. Seki · Nucl. Instrum. Methods · 2003 · B59/60 · 21
3. Akizuki · Nucl. Instrum. Methods · 1995 · B99 · 225
4. Qin · Mater. Chem. Phys. · 1998 · 54 · 25810.1016/S0254-0584(98)00109-6 ↗
5. Toyoda · Nucl. Instrum. Methods · 1999 · B148 · 63910.1016/S0168-583X(98)00769-1 ↗
6. Fenner · Mater. Res. Soc. Symp. Proc. · 2000 · 585 · 2710.1557/PROC-585-27 ↗
7. Goto · IEDM Technol. Dig. · 1997 · 471
8. Gillen · J. Vac. Sci. Technol. · 1999 · A17 · 84510.1116/1.581657 ↗
9. Gillen · J. Vac. Sci. Technol. · 2000 · B18 · 50310.1116/1.591221 ↗
10. Gillen · J. Vac. Sci. Technol. · 2001 · A19 · 56810.1116/1.1340651 ↗
11. Bouneau · Phys. Rev. · 2002 · B65 · 14410610.1103/PhysRevB.65.144106 ↗
12. Weibel · Anal. Chem. · 2003 · 75 · 175410.1021/ac026338o ↗
13. Mahoney · Rapid Commun. Mass Spectrom. · 1991 · 5 · 44110.1002/rcm.1290051004 ↗
14. Takats · Science · 2004 · 471 · 306
15. Aksyonov · Rapid Commun. Mass Spectrom. · 2001 · 15 · 200110.1002/rcm.470 ↗
16. Cody · Anal. Chem. · 2005 · 306 · 229710.1021/ac050162j ↗
17. Hiraoka · J. Mass Spectrom. · 2006 · 41 · 89410.1002/jms.1048 ↗
18. Winograd · Anal. Chem. · 2005 · 143A · 77
19. Matsuo · Nucl. Instrum. Methods · 2004 · B219 · 46310.1016/j.nimb.2004.01.103 ↗
20. Wittmaack · Surf. Sci. · 1996 · 557 · 90
21. Aoki · Nucl. Instrum. Methods · 2003 · B202 · 27810.1016/S0168-583X(02)01869-4 ↗
22. Toyoda · Nucl. Instrum. Methods · 2005 · B241 · 60910.1016/j.nimb.2005.07.079 ↗
23. Toyoda · Nucl. Instrum. Methods · 2005 · B242 · 466
24. Houzumi · Jpn. J. Appl. Phys. · 2005 · 44 · 625210.1143/JJAP.44.6252 ↗
25. Ninomiya · Nucl. Instrum. Methods · 2007 · B256 · 49310.1016/j.nimb.2006.12.047 ↗
26. Tanaka · Proc. Second Jpn.-China Joint Symp. Mass Spectrom. · 1987 · 185
27. Fenn · Science · 1989 · 246 · 6410.1126/science.2675315 ↗
28. Benninghoven · Phys. Status Solidi · 1969 · 34 · K16910.1002/pssb.19690340267 ↗
29. Touboul · Anal. Chem. · 2004 · 76 · 155010.1021/ac035243z ↗
30. Novikov · Rapid Commun. Mass Spectrom. · 2005 · 19 · 185110.1002/rcm.1995 ↗
31. Molecular dynamics simulations of sputtering of organic overlayers by slow, large clusters
32. Shard · Surf. Interface Anal. · 2006 · 39 · 29410.1002/sia.2525 ↗