High-resolution positron lifetime measurement using ultra fast digitizers Acqiris DC211
Source abstract
Abstract
This work demonstrates that a significant improvement of the timing resolution, simultaneously keeping coincidence count rate high enough, can be achieved by use of the state-of-the-art ultra fast 8-bit digitizers which replace the traditional analog timing modules. Performance of the digital spectrometer in conditions of a routine measurement is compared with that of the same detectors connected to the analog setup. It was found that employment of the digital setup leads to an improvement of timing resolution from 169 ps for the traditional analog setup down to 146 ps for the digital setup.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Muon and positron interactions and applications · Particle Detector Development and Performance · Radiation Detection and Scintillator Technologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (6)
Show source-provided references
1. Saito · Nucl. Instrum. Methods A · 2002 · 487 · 61210.1016/S0168-9002(01)02172-6 ↗
2. Bečvář · Nucl. Instrum. Methods A · 2000 · 443 · 55710.1016/S0168-9002(99)01156-0 ↗
3. Nissilä · Nucl. Instrum. Methods A · 2005 · 538 · 77810.1016/j.nima.2004.08.102 ↗
4. Bečvář · Nucl. Instrum. Methods A · 2005 · 539 · 37210.1016/j.nima.2004.09.031 ↗
5. Bečvář · Nucl. Instrum. Methods B · 2007 · 261 · 87110.1016/j.nimb.2007.03.042 ↗
6. Seeger · Phys. Status Solidi (a) · 1987 · 102 · 17110.1002/pssa.2211020117 ↗