Characterization of drug-eluting stent (DES) materials with cluster secondary ion mass spectrometry (SIMS)
Source abstract
Abstract
Secondary ion mass spectrometry (SIMS) employing an SF 5 + polyatomic primary ion source was utilized to analyze several materials commonly used in drug-eluting stents (DES). Poly(ethylene-co-vinyl acetate) (PEVA), poly(lactic-co-glycolic acid) (PLGA) and various poly(urethanes) were successfully depth profiled using SF 5 + bombardment. The resultant molecular depth profiles obtained from these polymeric films showed very little degradation in molecular signal as a function of increasing SF 5 + primary ion dose when experiments were performed at low temperatures (signal was maintained for doses up to ∼5 x 10 15 ions/cm 2 ). Temperature was determined to be an important parameter in both the success of the depth profiles and the mass spectral analysis of the polymers. In addition to the pristine polymer films, paclitaxel (drug released in Taxus™ stent) containing PLGA films were also characterized, where it was confirmed that both drug and polymer signals could be monitored as a function of depth at lower paclitaxel concentrations (10 wt%).
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Ion-surface interactions and analysis · Integrated Circuits and Semiconductor Failure Analysis · Analytical chemistry methods development
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (10)
Show source-provided references
1. Van der Hoeven · Int. J. Cardiol. · 2005 · 99 · 910.1016/j.ijcard.2004.01.021 ↗
2. Dogrell · Expert Opin. Pharmacother. · 2004 · 5 · 220910.1517/14656566.5.11.2209 ↗
3. Ellis · Crit. Rev. Ther. Drug Carr. Sys. · 2004 · 22 · 110.1615/CritRevTherDrugCarrierSyst.v22.i1.10 ↗
4. See product information available at: http://www.cordislabeling.com/pdf/5462872_4.pdf.
5. Mahoney · Anal. Chem. · 2004 · 76 · 319910.1021/ac035532n ↗
6. Mahoney · Anal. Chem. · 2005 · 77 · 357010.1021/ac048274i ↗
7. Wagner · Surf. Interface Anal. · 2005 · 37 · 4210.1002/sia.2006 ↗
8. G. Gillen, C. Mahoney, S. Wight, R. Lareau, Analysis and imaging of high explosive particles using cluster secondary ion mass spectrometry (SIMS), Rapid Commun. Mass Spectrom., in press.
9. Brox · 2000 · 777
10. C.M. Mahoney, A.J. Fahey, G. Gillen, C. Xu, J.D. Batteas, Temperature-controlled depth profiling of poly(methylmethacrylate) (PMMA) using cluster secondary ion mass spectrometry (SIMS), Anal. Chem., submitted for publication.