Abstract

Abstract Authors have observed an unusual phenomenon that the collector current in an Extractor vacuum gauge, which several times experienced a bakeout of 570 K, about 70 h, becomes negative at pressures in or near the XHV region. To investigate the causes, the temperature dependence of the collector current, the electromotive force (e.m.f.), and the insulation resistance of the collector electrode for both an “unusual” and a new gauge were measured. As a result, their dependence for the unusual gauge showed strikingly different behaviors from those for the new gauge. These results suggest that some new structure is formed in the gauge head by several bakeouts. Also it was found that the new gauge shows a positive larger e.m.f., but the current due to this e.m.f. was very small because the collector electrode has a very high insulation resistance.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Electrical Measurement Techniques · Radioactive Decay and Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (3)

Show source-provided references
1. Bayard · Rev. Sci. Instrum. · 1950 · 21 · 57110.1063/1.1745653 ↗
2. J.W. Ackley, C.F. Lothrop, W.R. Wheeler, Trans. 9th AVS Vac. Symp. (1962) 452.
3. Kokubun · J. Vac. Soc. Jpn · 1997 · 40 · 13710.3131/jvsj.40.137 ↗