journal article·2022-04Open access

Two-stage degradation and novel functional endothelium characteristics of a 3-D printed bioresorbable scaffold

Tieying Yin, Ruolin Du, Yang Wang, Junyang Huang, Shuang Ge, Yuhua Huang, Youhua Tan, Qing Liu, Zhong Chen, Hanqing Feng, Jie Du, Yazhou Wang, Guixue Wang

Abstract

Bioresorbable scaffolds have emerged as a new generation of vascular implants for the treatment of atherosclerosis, and designed to provide a temporary scaffold that is subsequently absorbed by blood vessels over time. Presently, there is insufficient data on the biological and mechanical responses of blood vessels accompanied by bioresorbable scaffolds (BRS) degradation. Therefore, it is necessary to investigate the inflexion point of degradation, the response of blood vessels, and the pathophysiological process of vascular, as results of such studies will be of great value for the design of next generation of BRS. In this study, abdominal aortas of SD rats were received 3-D printed poly-l-actide vascular scaffolds (PLS) for various durations up to 12 months. The response of PLS implanted aorta went through two distinct processes: (1) the neointima with desirable barrier function was obtained in 1 month, accompanied with slow degradation, inflammation, and intimal hyperplasia; (2) significant degradation occurred from 6 months, accompanied with decreasing inflammation and intimal hyperplasia, while the extracellular matrix recovered to normal vessels which indicate the positive remodeling. These in vivo results indicate that 6 months is a key turning point. This “two-stage degradation and vascular characteristics” is proposed to elucidate the long-term effects of PLS on vascular repair and demonstrated the potential of PLS in promoting endothelium function and positive remodeling, which highlights the benefits of PLS and shed some light in the future researches, such as drug combination coatings design.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Electrospun Nanofibers in Biomedical Applications · Aortic aneurysm repair treatments · Aortic Disease and Treatment Approaches

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (69)

Show source-provided references
1. Fully bioresorbable vascular scaffolds: lessons learned and future directions10.1038/s41569-018-0124-7 ↗
2. Bioresorbable scaffold: the advent of a new era in percutaneous coronary and peripheral revascularization?10.1161/CIRCULATIONAHA.110.971606 ↗
3. First-in-human evaluation of a novel poly-l-lactide based sirolimus-eluting bioresorbable vascular scaffold for the treatment of de novo native coronary artery lesions: meres-1 trial10.4244/EIJ-D-17-00306 ↗
4. A multicenter post-marketing evaluation of the Elixir DESolve ® Novolimus-eluting bioresorbable coronary scaffold system: first results from the DESolve PMCF study, Catheter10.1002/ccd.27550 ↗
5. Neovas randomized controlled trial investigators, a randomized trial comparing the neovas sirolimus-eluting bioresorbable scaffold and metallic everolimus-eluting stents10.1016/j.jcin.2017.09.037 ↗
6. Serruys · Lancet · 2015 · 385 · 4310.1016/S0140-6736(14)61455-0 ↗
7. Absorb bioresorbable vascular scaffold: what have we learned after 5 years of clinical experience?10.1016/j.ijcard.2015.07.101 ↗
8. Randomised comparison of a bioresorbable everolimus-eluting scaffold with a metallic everolimus-eluting stent for ischaemic heart disease caused by de novo native coronary artery lesions: the 2-year clinical outcomes of the absorb II trial10.4244/EIJY16M08_01 ↗
9. Absorb III investigators, clinical outcomes before and after complete everolimus-eluting bioresorbable scaffold resorption: five-year follow-up from the absorb III trial10.1161/CIRCULATIONAHA.119.042584 ↗
10. Long-term vasomotion after absorb: fact or fiction?10.1016/j.jcin.2018.05.028 ↗
11. Mechanical strength, biodegradation, and in vitro and in vivo biocompatibility of Zn biomaterials10.1021/acsami.8b20634 ↗
12. Evolution of the degradation mechanism of pure zinc stent in the one-year study of rabbit abdominal aorta model10.1016/j.biomaterials.2017.08.022 ↗
13. XINSORB randomized clinical trial, Twelve-month angiographic and clinical outcomes of the xinsorb bioresorbable sirolimus-eluting scaffold and a metallic stent in patients with coronary artery disease10.1016/j.ijcard.2019.06.053 ↗
14. Clinical and multimodality imaging results at 6 months of a bioresorbable sirolimus-eluting scaffold for patients with single de novo coronary artery lesions: the neovas first-in-man trial10.4244/EIJV12I10A209 ↗
15. Three-year clinical outcomes with the ABSORB bioresorbable vascular scaffold in real life: insights from the France ABSORB registry
16. Three-year clinical outcomes of a sirolimus-eluting bioresorbable scaffold (XINSORB) and a metallic stent to treat coronary artery stenosis10.21037/atm-20-6739 ↗
17. Final report of the 5-year clinical outcomes of the XINSORB bioresorbable sirolimus-eluting scaffold in the treatment of single de novo coronary lesions in a first-in-human study10.21037/atm-20-5668 ↗
18. A novel three-dimensional printed sirolimus-eluting bioresorbable vascular scaffold evaluated by optical coherence tomography in a porcine model
19. Drug loaded nanoparticle coating on totally bioresorbable PLLA stents to prevent in-stent restenosis10.1002/jbm.b.33794 ↗
20. A practical method to rapidly dissolve metallic stents10.1016/j.carpath.2009.01.001 ↗
21. Comparison of acute recoil between bioabsorbable poly-L-lactic acid XINSORB stent and metallic stent in porcine model10.1155/2012/413956 ↗
22. An independent bench comparison of two bioresorbable drug-eluting coronary scaffolds (Absorb and DESolve) with a durable metallic drug-eluting stent (ML8/Xpedition)10.4244/EIJY15M02_03 ↗
23. Biomechanical and histological evaluation of the application of biodegradable poly-L-lactic cushion to the plate internal fixation for bone fracture healing10.1016/j.clinbiomech.2008.01.005 ↗
24. Poly(lactic- co-glycolic acid)-based composite bone-substitute materials
25. Short poly(ethylene glycol) block initiation of poly(l-lactide) di-block copolymers: a strategy for tuning the degradation of resorbable devices10.1002/pi.5563 ↗
26. M2 macrophage-derived exosomes promote the c-KIT phenotype of vascular smooth muscle cells during vascular tissue repair after intravascular stent implantation10.7150/thno.46143 ↗
27. Vascular inflammation and repair: implications for re-endothelialization, restenosis, and stent thrombosis10.1016/j.jcin.2011.05.025 ↗
28. Current status of bioresorbable scaffolds in the treatment of coronary artery disease10.1016/j.jacc.2014.09.041 ↗
29. On the importance of modeling stent procedure for predicting arterial mechanics10.1115/1.4023094 ↗
30. Structural optimization and finite element analysis of poly-l-lactide acid coronary stent with improved radial strength and acute recoil rate10.1002/jbm.b.34605 ↗
31. Multi-objective optimisation of material properties and strut geometry for poly (L-lactic acid) coronary stents using response surface methodology10.1371/journal.pone.0218768 ↗
32. Finite element investigation of GO reinforced PLLA stent deployment
33. Hemodynamically driven stent strut design10.1007/s10439-009-9719-9 ↗
34. Endothelial barrier and its abnormalities in cardiovascular disease10.3389/fphys.2015.00365 ↗
35. Endothelial cell recovery between comparator polymer-based drug-eluting stents10.1016/j.jacc.2008.04.030 ↗
36. Direct targeting of the mTOR (mammalian target of rapamycin) kinase improves endothelial permeability in drug-eluting stents-brief report10.1161/ATVBAHA.118.311321 ↗
37. Overview of the latest developments in the field of drug-eluting stent technology10.1039/C9BM00468H ↗
38. Mussel adhesive protein fused with VE-cadherin extracellular domain promotes endothelial-cell tight junctions and in vivo endothelization recovery of vascular stent10.1002/jbm.b.34369 ↗
39. Arsenic trioxide-coated stent is an endothelium-friendly drug eluting stent
40. The control of vascular integrity by endothelial cell junctions: molecular basis and pathological implications10.1016/j.devcel.2009.01.004 ↗
41. Tgf‐β1 induces rearrangement of flk‐1–ve‐cadherin–β‐catenin complex at the adherens junction through vegf‐mediated signaling10.1002/jcb.21935 ↗
42. Neoatherosclerosis 5 years after bioresorbable vascular scaffold implantation10.1016/j.jacc.2018.02.051 ↗
43. Pcl-plla semi-ipn shape memory polymers (smps): degradation and mechanical properties10.1002/marc.201600414 ↗
44. Onuma · JACC Cardiovasc. Interv. · 2014 · 7 · 140010.1016/j.jcin.2014.06.016 ↗
45. Microvascular endothelial cells engulf myelin debris and promote macrophage recruitment and fibrosis after neural injury10.1038/s41593-018-0324-9 ↗
46. Tissue response to poly (l-lactic acid)-based blend with phospholipid polymer for biodegradable cardiovascular stents10.1016/j.biomaterials.2010.11.067 ↗
47. Reducing in-stent restenosis: therapeutic manipulation of mirna in vascular remodeling and inflammation10.1016/j.jacc.2015.03.549 ↗
48. Vascular macrophages in atherosclerosis
49. Myeloid dendritic cells: development, functions, and role in atherosclerotic inflammation10.1016/j.imbio.2014.12.010 ↗
50. Thrombomodulin and its role in inflammation, Semin10.1007/s00281-011-0282-8 ↗
51. Inflammatory and immune responses in the arterial media10.1161/CIRCRESAHA.116.301312 ↗
52. Klf4-dependent phenotypic modulation of smooth muscle cells has a key role in atherosclerotic plaque pathogenesis10.1038/nm.3866 ↗
53. Insulin-like growth factor-1 isoform mrna expression in women with endometriosis: eutopic endometrium versus endometriotic cyst10.1196/annals.1365.042 ↗
54. Igf-iec expression is associated with advanced clinical and pathological stage of prostate cancer
55. The role of bmps in endothelial cell function and dysfunction10.1016/j.tem.2014.05.003 ↗
56. Improving the osteogenic efficacy of bmp2 with mechano growth factor by regulating the signaling events in bmp pathway10.1007/s00441-015-2154-3 ↗
57. Biomimetic nanotherapies: red blood cell based core-shell structured nanocomplexes for atherosclerosis management10.1002/advs.201900172 ↗
58. Polydopamine modified tio2 nanotube arrays for long-term controlled elution of bivalirudin and improved hemocompatibility10.1021/acsami.7b06108 ↗
59. Modified magnesium mydroxide nanoparticles inhibit the inflammatory response to biodegradable poly(lactide-co-glycolide) implants10.1021/acsnano.8b02365 ↗
60. A novel mechanism of inhibiting in-stent restenosis with arsenic trioxide drug-eluting stent: enhancing contractile phenotype of vascular smooth muscle cells via YAP pathway
61. Design and testing of hydrophobic core/hydrophilic shell nano/micro particles for drug-eluting stent coating10.1038/s41427-018-0064-z ↗
62. Lactic acid-mediated endothelial to mesenchymal transition through TGF-β1 contributes to in-stent stenosis in poly-L-lactic acid stent10.1016/j.ijbiomac.2019.11.136 ↗
63. Role of animal models in coronary stenting10.1007/s10439-015-1414-4 ↗
64. Drug-eluting coronary stents--promise and uncertainty, N10.1056/NEJMe068306 ↗
65. Local delivery of anti-monocyte chemoattractant protein-1 by gene-eluting stents attenuates in-stent stenosis in rabbits and monkeys10.1161/ATVBAHA.107.154609 ↗
66. Local microRNA modulation using a novel anti-miR-21-eluting stent effectively prevents experimental in-stent restenosis10.1161/ATVBAHA.115.305597 ↗
67. Patency of perforating arteries after stent placement? A study using an in vivo experimental atherosclerosis-induced model
68. Two-dimensional intravascular near-infrared fluorescence molecular imaging of inflammation in atherosclerosis and stent-induced vascular injury10.1016/j.jacc.2011.02.036 ↗
69. Balloon-expandable intravascular stent10.2214/ajr.150.6.1263 ↗