journal article·2022-04Open access

Translation of a spinal bone cement product from bench to bedside

Fei Feng, Mengmeng Chen, Xuan Wang, Hongwei Zhang, Hongtao Nie, Hai Tang

Abstract

Spinal acrylic bone cements (ABCs) are used clinically for percutaneous vertebroplasty (PVP) and kyphoplasty (PKP) to treat osteoporotic vertebral compression fractures. Product translation of spinal ABC products followed the design control processes including design verification and validation. The bench to bedside translation of the first Chinese spinal ABC product (Alliment®, namely Alliment Cement) approved by National Medical Products Administration of China was investigated and another commercial product served as the control (Osteopal®V, namely Osteopal V Cement). Results of non-clinical bench performance verification tests of compression, bending and monomer release showed that the newly marketed Alliment Cement is similar to the Osteopal V Cement with properties of both meeting the criteria specified by standards. The Alliment Cement demonstrated good biocompatibility during the 26 weeks' bone implantation test. Porcine cadaver validation tests further revealed that the Alliment Cement satisfied the needs for both PVP and PKP procedures. A post-approval, retrospective clinical investigation further demonstrated the safety and efficacy of the Alliment Cement, with a significant reduction of pain and the improved stability of the fractured vertebral bodies. A successful translation of biomaterial medical products needs close collaborations among academia, industry, healthcare professionals and regulatory agencies.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Spinal Fractures and Fixation Techniques · Spine and Intervertebral Disc Pathology · Pelvic and Acetabular Injuries

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (66)

Show source-provided references
1. The genesis and evolution of acrylic bone cement10.1016/j.ocl.2004.06.012 ↗
2. Clinical development and current status: Europe10.1016/j.ocl.2004.07.002 ↗
3. Acrylic bone cement: clinical development and current status in North America10.1016/j.ocl.2004.06.005 ↗
4. Alternative acrylic bone cement formulations for cemented arthroplasties: present status, key issues, and future prospects10.1002/jbm.b.30873 ↗
5. Properties of acrylic bone cement: state of the art review10.1002/(SICI)1097-4636(199722)38:2<155::AID-JBM10>3.0.CO;2-C ↗
6. Viscoelastic properties of injectable bone cements for orthopaedic applications: state-of-the-art review10.1002/jbm.b.31835 ↗
7. Fatigue testing and performance of acrylic bone-cement materials: state-of-the-art review10.1002/jbm.b.10018 ↗
8. Toward standardization of methods of determination of fracture properties of acrylic bone cement and statistical analysis of test results10.1002/1097-4636(2000)53:6<748::AID-JBM18>3.0.CO;2-Z ↗
9. Acrylic bone cements: composition and properties10.1016/j.ocl.2004.06.010 ↗
10. Acrylic bone cements: mechanical and physical properties10.1016/j.ocl.2004.06.011 ↗
11. Bone cements and fillers: a review10.1023/A:1026394530192 ↗
12. Tensile characteristics of ten commercial acrylic bone cements10.1002/1097-4636(200009)53:5<605::AID-JBM22>3.0.CO;2-5 ↗
13. Properties of antibiotic-loaded acrylic bone cements for use in cemented arthroplasties: a state-of-the-art review10.1002/jbm.b.31220 ↗
14. Backgrounds of antibiotic-loaded bone cement and prosthesis-related infection10.1016/S0142-9612(03)00554-4 ↗
15. Influence of the method of blending an antibiotic powder with an acrylic bone cement powder on physical, mechanical, and thermal properties of the cured cement10.1016/j.biomaterials.2004.11.003 ↗
16. Preliminary focus on the mechanical and antibacterial activity of a PMMA-based bone cement loaded with gold nanoparticles
17. Studies on the cytocompatibility, mechanical and antimicrobial properties of 3D printed poly(methyl methacrylate) beads
18. Antibiotic-loaded bone cement is associated with a lower risk of revision following primary cemented total knee arthroplasty: an analysis of 731,214 cases using National Joint Registry data10.1302/0301-620X.101B11.BJJ-2019-0196.R1 ↗
19. Injectable bone cements for use in vertebroplasty and kyphoplasty: state-of-the-art review10.1002/jbm.b.30398 ↗
20. Bone cements: review of their physiochemical and biochemical properties in percutaneous vertebroplasty
21. Vertebroplasty and kyphoplasty: filler materials10.1016/j.spinee.2005.02.020 ↗
22. Computational modelling of bone augmentation in the spine10.1016/j.jot.2015.09.003 ↗
23. Future directions Augmentation of osteoporotic vertebral bodies10.1097/00007632-199712151-00007 ↗
24. Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty10.1097/00007632-200107150-00009 ↗
25. Reporting height restoration in vertebral compression fractures10.1097/01.BRS.0000092424.29886.C9 ↗
26. Chemical and physical properties of bone cement for vertebroplasty10.4103/2319-4170.112750 ↗
27. Bone cements for percutaneous vertebroplasty and balloon kyphoplasty: current status and future developments10.1016/j.jot.2014.11.002 ↗
28. A review of the design process for implantable orthopedic medical devices10.2174/1874120700903010021 ↗
29. Heraeus
30. ISO
31. ASTM
32. Kühn · 2000 · 247
33. ISO
34. Is Kiva implant advantageous to balloon kyphoplasty in treating osteolytic metastasis to the spine? Comparison of 2 percutaneous minimal invasive spine techniques: a prospective randomized controlled short-term study10.1097/BRS.0000000000000112 ↗
35. The Oswestry disability index10.1097/00007632-200011150-00017 ↗
36. Clinical applications of visual analogue scales: a critical review10.1017/S0033291700009934 ↗
37. Stentoplasty with resorbable calcium salt bone void fillers for the treatment of vertebral compression fracture: evaluation after 3 years10.2147/CIA.S308667 ↗
38. A novel and convenient method to evaluate bone cement distribution following percutaneous vertebral augmentation10.1038/s41598-020-73513-2 ↗
39. Heraeus
40. Kühn · 2000 · 95
41. [Preliminary note on the treatment of vertebral angioma by percutaneous acrylic vertebroplasty]
42. Vertebral compression fractures: pain reduction and improvement in functional mobility after percutaneous polymethylmethacrylate vertebroplasty retrospective report of 245 cases10.1148/radiol.2262010906 ↗
43. Translational status of biomedical Mg devices in China
44. Kyphoplasty and vertebroplasty in the management of osteoporosis with subsequent vertebral compression fractures
45. Determination of interfacial fracture toughness of bone-cement interface using sandwich Brazilian disks10.1016/j.engfracmech.2006.02.014 ↗
46. Mechanical properties of bone cement: a review10.1002/jbm.820180411 ↗
47. Improving mechanical and antibacterial properties of PMMA via polyblend electrospinning with silk fibroin and polyethyleneimine towards dental applications
48. Synthesis of tetracalcium phosphate from mechanochemically activated reactants and assessment as a component of bone cements10.1007/s10856-008-3403-8 ↗
49. Vivo temperature profile of intervertebral discs and vertebral endplates during vertebroplasty: an experimental study in sheep10.1097/01.brs.0000224193.52587.d8 ↗
50. The effects of microwave heating on osteoinduction of demineralized bone matrix in rabbits
51. The European Spine Society AcroMed Prize 1994. Acute thermal nerve root injury10.1007/BF02200140 ↗
52. Bone cement implantation syndrome10.1093/bja/aen328 ↗
53. Clinical features and echocardiography of embolism during cemented hip arthroplasty10.1007/BF03012997 ↗
54. Kyphoplasty and vertebroplasty: how good is the evidence?10.1007/s11926-007-0023-0 ↗
55. Percutaneous vertebroplasty with high- versus low-viscosity bone cement for osteoporotic vertebral compression fractures10.1186/s13018-020-01835-y ↗
56. High-viscosity bone cement for vertebral compression fractures: a prospective study on intravertebral diffusion and leakage of bone cement10.1186/s12891-020-03613-7 ↗
57. Influence of mixing method on the cement temperature-mixing time history and doughing time of three acrylic cements for vertebroplasty10.1002/jbm.b.20009 ↗
58. A comparative study of high-viscosity cement percutaneous vertebroplasty vs. low-viscosity cement percutaneous kyphoplasty for treatment of osteoporotic vertebral compression fractures10.1007/s11596-016-1597-4 ↗
59. Cement leakage in percutaneous vertebroplasty for osteoporotic vertebral compression fractures: identification of risk factors10.1016/j.spinee.2011.07.027 ↗
60. Combined extraforaminal and intradiscal cement leakage following percutaneous vertebroplasty10.1097/BRS.0b013e31805c0b25 ↗
61. A comparison of high viscosity bone cement and low viscosity bone cement vertebroplasty for severe osteoporotic vertebral compression fractures10.1016/j.clineuro.2014.11.018 ↗
62. A cadaver study to compare vertebral augmentation with a high-viscosity cement to augmentation with conventional lower-viscosity cement10.1097/BSD.0b013e3182337d69 ↗
63. An updated comparison of high- and low-viscosity cement vertebroplasty in the treatment of osteoporotic thoracolumbar vertebral compression fractures: a retrospective cohort study10.1016/j.ijsu.2017.05.067 ↗
64. High-viscosity cement significantly enhances uniformity of cement filling in vertebroplasty: an experimental model and study on cement leakage10.1097/01.brs.0000240695.58651.62 ↗
65. Influence of cement viscosity on cement interdigitation and venous fat content under in vivo conditions: a bilateral study of 13 sheep10.1080/00016470216320 ↗
66. Vertebroplasty: cement leakage into the disc increases the risk of new fracture of adjacent vertebral body