Evaluation of FeMnN alloy bioresorbable flow diverting stents in the rabbit abdominal aorta
Source abstract
Abstract
Flow diverting stents are braided, metallic endoluminal devices widely used to treat intracranial aneurysms. Bioresorbable flow diverters (BRFDs) are gaining traction as the next generation of flow diverter technology. BRFDs aim to occlude and heal the aneurysm before safely dissolving into the body, mitigating or eliminating complications associated with the permanent presence of conventional flow diverters such as thromboembolism and stenosis. Additional putative advantages of a BRFD include a reduction in metal induced medical imaging artifacts, a restoration of physiological vasoreactivity, and allowing physicians to re-access the aneurysm if an additional procedure is required. In this current study, iron-manganese-nitrogen (FeMnN) alloy BRFDs and permanent control FDs composed of an industry standard Cobalt-Nickel-Chromium alloy were deployed in the rabbit aorta. MicroCT and SEM corrosion analysis determined the FeMnN wire volumes and cross-sectional areas had reduced approximately 85 % and 95 % after 3- and 6-months implantation duration, respectively. Histological analysis demonstrated the BRFDs exhibited suitable biocompatibility, with no cases of in-stent thrombosis, clinically significant stenosis, or adverse tissue responses observed. Immunohistochemistry revealed the neointimas surrounding the BRFDs featured a confluent endothelium covering several layers of smooth muscle cells, with macrophages adjacent to the device wires. The macrophages were able to penetrate the corrosion product and were observed transporting corrosion products away from the implant site. This current work provides primary in vivo corrosion and biocompatibility data to the field for FeMn alloys, which we feel will stimulate and inform the design of next-generation bioresorbable endovascular devices. • Bioresorbable FeMnN alloy investigated in vivo in the arteries for the first time. • Their volume reduced ∼85 % and ∼95 % after 3- and 6-months in vivo , respectively. • Corrosion of the FeMnN alloy was heterogeneous and non-uniform. • The bioresorbable flow diverting stents exhibited suitable biocompatibility. • Inflammatory cells could infiltrate and transport corrosion products.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Aortic aneurysm repair treatments · Intracranial Aneurysms: Treatment and Complications · Aortic Disease and Treatment Approaches
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (52)
Show source-provided references
1. A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal—results 6–18 months after implantation into New Zealand white rabbits10.1136/heart.86.5.563 ↗
2. Bioresorbable vascular scaffolds—basic concepts and clinical outcome10.1038/nrcardio.2016.151 ↗
3. Recent advances and directions in the development of bioresorbable metallic cardiovascular stents: insights from recent human and in vivo studies10.1016/j.actbio.2021.03.058 ↗
4. Magnetic resonance (MR) safety and compatibility of a novel iron bioresorbable scaffold
5. In vivo serial invasive imaging of the second-generation drug-eluting absorbable metal scaffold (Magmaris—DREAMS 2G) in de novo coronary lesions: insights from the BIOSOLVE-II first-in-man trial10.1016/j.ijcard.2017.12.053 ↗
6. Absorbable metal stents for vascular use in pediatric cardiology: progress and outlook10.3389/fcvm.2024.1410305 ↗
7. Initial status and 3-month results relating to the use of biodegradable nitride iron stents in children and the evaluation of right ventricular function
8. Acute treatment of critical vascular stenoses with a bioabsorbable magnesium scaffold in infants with CHDs10.1017/S1047951120000384 ↗
9. BIOSOLVE‐IV‐registry: safety and performance of the Magmaris scaffold: 12‐month outcomes of the first cohort of 1,075 patients
10. First-in-human evaluation of a novel ultrathin sirolimus-eluting iron bioresorbable scaffold: 3-year outcomes of the IBS-FIM trial
11. Bioresorbable flow diverters for the treatment of intracranial aneurysms: review of current literature and future directions10.1136/neurintsurg-2022-018941 ↗
12. Intraluminal flow diverter design primer for neurointerventionalists10.3174/ajnr.A8076 ↗
13. Cellular mechanisms of aneurysm occlusion after treatment with a flow diverter10.1148/radiol.13130796 ↗
14. Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis10.1161/STROKEAHA.112.678151 ↗
15. Long-term clinical and angiographic outcomes following pipeline embolization device treatment of complex internal carotid artery aneurysms: five-year results of the pipeline for uncoilable or failed aneurysms trial10.1093/neuros/nyw014 ↗
16. Review of current intracranial aneurysm flow diversion technology and clinical use10.1136/neurintsurg-2020-015877 ↗
17. International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study10.3174/ajnr.A4111 ↗
18. Very late ischemic complications in flow-diverter stents: a retrospective analysis of a single-center series10.3171/2015.10.JNS15703 ↗
19. Phosphorylcholine surface modified flow diverter associated with reduced intimal hyperplasia10.1136/neurintsurg-2018-013776 ↗
20. Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia10.1136/neurintsurg-2013-010950 ↗
21. The implications of magnetic resonance angiography artifacts caused by different types of intracranial flow diverters10.1186/s12968-021-00753-0 ↗
22. Neurovascular stent artifacts in 3D‐TOF and 3D‐PCMRI: influence of stent design on flow measurement10.1002/mrm.27352 ↗
23. Benchtop proof of concept and comparison of iron-and magnesium-based bioresorbable flow diverters
24. Iron–manganese: new class of metallic degradable biomaterials prepared by powder metallurgy10.1179/174329008X284868 ↗
25. Cold-drawn bioabsorbable ferrous and ferrous composite wires: an evaluation of mechanical strength and fatigue durability10.1007/s11663-012-9661-3 ↗
26. Medical imaging compatibility of magnesium-and iron-based bioresorbable flow diverters10.3174/ajnr.A7873 ↗
27. Cold drawn bioabsorbable ferrous and ferrous composite wires: an evaluation of in vitro vascular cytocompatibility10.1016/j.actbio.2012.07.043 ↗
28. A novel biodegradable high nitrogen iron alloy with simultaneous enhancement of corrosion rate and local corrosion resistance10.1016/j.jmst.2023.01.009 ↗
29. Histopathologic and immunohistochemical comparison of human, rabbit, and swine aneurysms embolized with platinum coils
30. A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms10.3174/ajnr.A1530 ↗
31. Evaluation of flow diverters for cerebral aneurysm therapy: recommendations for imaging analyses in clinical studies, endorsed by ESMINT, ESNR, OCIN, SILAN, SNIS, and WFITN
32. Long-term biocompatibility of a corrodible peripheral iron stent in the porcine descending aorta10.1016/j.biomaterials.2006.05.029 ↗
33. Long-term in vivo corrosion behavior, biocompatibility and bioresorption mechanism of a bioresorbable nitrided iron scaffold10.1016/j.actbio.2017.03.020 ↗
34. Inflammation, neointimal hyperplasia, and restenosis: as the leukocytes roll, the arteries thicken
35. Biodegradable iron-based materials—what was done and what more can be done?10.3390/ma14123381 ↗
36. Radiopaque FeMnN-Mo composite drawn filled tubing wires for braided absorbable neurovascular devices
37. Semiconducting and passive film properties of nitrogen-containing type 316LN stainless steels10.1016/j.corsci.2006.05.041 ↗
38. A simplified in vivo approach for evaluating the bioabsorbable behavior of candidate stent materials10.1002/jbm.b.31922 ↗
39. Current status and outlook on the clinical translation of biodegradable metals10.1016/j.mattod.2018.05.018 ↗
40. Schaffer · 2012
41. A comparative study of pulsatile arterial hemodynamics in rabbits and Guinea pigs10.1152/ajplegacy.1976.230.4.868 ↗
42. Biodegradable magnesium materials regulate ROS-RNS balance in pro-inflammatory macrophage environment
43. Accelerating corrosion of pure magnesium co-implanted with titanium in vivo
44. Oxidant production from corrosion of nano-and microparticulate zero-valent iron in the presence of oxygen: a comparative study10.1016/j.jhazmat.2013.11.066 ↗
45. Oxygen free radicals and systemic autoimmunity10.1046/j.1365-2249.2003.02104.x ↗
46. Hydroxyl radicals and oxidative stress: the dark side of Fe corrosion10.1016/j.colsurfb.2019.110542 ↗
47. Endothelial dysfunction induced by hydroxyl radicals–the hidden face of biodegradable Fe-based materials for coronary stents10.1016/j.msec.2020.110938 ↗
48. Bioresorbable vascular metallic scaffolds: current status and research trends
49. Long-term safety and absorption assessment of a novel bioresorbable nitrided iron scaffold in porcine coronary artery
50. Restenosis and the proportional neointimal response to coronary artery injury: results in a porcine model10.1016/0735-1097(92)90476-4 ↗
51. Creation of saccular aneurysms in the rabbit: a model suitable for testing endovascular devices10.2214/ajr.174.2.1740349 ↗
52. Creating elastase aneurysms in rabbits: a video primer