Hydrodynamic-derived centrifugal blood pump design for stable-low-flow-rate rate performance: from surface to structure
Source abstract
Abstract
Long-lasting stability of the anticoagulant coating on centrifugal blood pumps (CBPs) is of vital importance to ensure the hemocompatibility of the blood circulation system therein. Heparin coatings are often prepared using static wet chemical technique, but these face risks of delamination or deactivation induced by blood flow. Inspired by the flow-shear-stress mediated conformation changes of von Willebrand factor, a novel fluid-driven deposition technique was employed to apply polydopamine-heparin coatings within CBPs. Moreover, most FDA-approved CBPs are designed for high-flow-rate CBPs of major organs like the heart and lungs (1000∼8000 ml/min). Few are tailored for low-flow-rate perfusion of other organs such as the liver, kidney and brain (<50–300 mL/min). Our approach addresses this gap by developing low-flow-rate CBPs through anti-thrombogenic coatings and anti-hemolytic structural optimizations. In this study, we introduced an axial magnetic direct drive motor with our optimized low-flow-rate CBPs, achieving a stable-low-flow-rate rate ranging from 16.3 mL/min (300 rpm) to 121.0 mL/min (2000 rpm). The resulting CBPs system exhibited enhanced flow stability and hemocompatibility in rabbit model experiments, demonstrating significantly lower hemolysis rates and lower thrombus formation risks. These results indicate that the polydopamine-assisted heparin coating provides short-term stability under dynamic flow, offering a promising strategy for low-flow-rate CBPs, though its long-term durability and clinical translation potential require further validation. • Shear flow enables stable, dense polydopamine-heparin deposition on pump surfaces. • CFD-guided impeller/gap design minimizes hemolysis risk (HI = 0.78 %) and thrombosis. • Enables stable low-flow perfusion (16–121 mL/min) for small organs. • Coated pump significantly reduced thrombus formation and hemolysis in rabbit models. • Addresses critical gap for brain/kidney/liver support circuits.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Mechanical Circulatory Support Devices · Intravenous Infusion Technology and Safety · Cardiac and Coronary Surgery Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (51)
Show source-provided references
1. CFD assisted evaluation of In vitro experiments on bearingless blood pumps10.1109/TBME.2020.3030316 ↗
2. Multi-indicator analysis of mechanical blood damage with five clinical ventricular assist devices10.1016/j.compbiomed.2022.106271 ↗
3. Feasibility and safety of extracorporeal CO2 removal to enhance protective ventilation in acute respiratory distress syndrome: the SUPERNOVA study10.1007/s00134-019-05567-4 ↗
4. Hemolysis at low blood flow rates: in-vitro and in-silico evaluation of a centrifugal blood pump10.1186/s12967-020-02599-z ↗
5. An experimental and numerical study of the flow and mass transfer in a model of the wearable artificial kidney dialyzer10.1186/1475-925X-9-21 ↗
6. Extracorporeal kidney-replacement therapy for Acute kidney injury10.1056/NEJMra2104090 ↗
7. Hypothermic machine perfusion in liver transplantation - a randomized trial10.1056/NEJMoa2031532 ↗
8. Safety, feasibility, and potential efficacy of intraarterial selective cooling infusion for stroke patients treated with mechanical thrombectomy10.1177/0271678X18790139 ↗
9. Evaluating the effect of shear stress on graft-to zwitterionic polycarboxybetaine coating stability using a flow cell10.1021/acs.langmuir.8b03078 ↗
10. The effect of electrostatic heparin/collagen layer-by-layer coating degradation on the biocompatibility10.1016/j.apsusc.2015.11.227 ↗
11. Comparative study on corrosion behaviour of pure Mg and WE43 alloy in static, stirring and flowing Hank's solution10.1179/1743278212Y.0000000006 ↗
12. Dynamic degradation behavior of MgZn alloy in circulating m-SBF10.1016/j.matlet.2010.06.011 ↗
13. Flow-induced corrosion behavior of absorbable magnesium-based stents10.1016/j.actbio.2014.08.034 ↗
14. The blood compatibility challenge. Part 1: Blood-contacting medical devices: the scope of the problem10.1016/j.actbio.2019.06.021 ↗
15. Evaluating the effect of shear stress on Graft-To zwitterionic polycarboxybetaine coating stability using a flow cell10.1021/acs.langmuir.8b03078 ↗
16. Bioinspired shear-flow-driven layer-by-layer in situ self-assembly10.1021/acsnano.9b02010 ↗
17. Flow-induced elongation of von Willebrand factor precedes tension-dependent activation10.1038/s41467-017-00230-2 ↗
18. Activation of von Willebrand factor via mechanical unfolding of its discontinuous autoinhibitory module10.1038/s41467-021-22634-x ↗
19. Layered nanocomposites by shear-flow-induced alignment of nanosheets10.1038/s41586-020-2161-8 ↗
20. Biofunctionalization via flow shear stress resistant adhesive polysaccharide, hyaluronic acid-catechol, for enhanced in vitro endothelialization10.1016/j.jiec.2015.11.015 ↗
21. Rapidly-Deposited polydopamine coating via high temperature and vigorous stirring: formation, characterization and biofunctional evaluation10.1371/journal.pone.0113087 ↗
22. Hemolysis and cardiopulmonary bypass: meta-analysis and systematic review of contributing factors10.1186/s13019-023-02406-y ↗
23. Dynamics of collapsed polymers under the simultaneous influence of elongational and shear flows10.1063/1.3606392 ↗
24. Dynamics of a tethered polymer in shear flow10.1103/PhysRevLett.84.4769 ↗
25. Schneider · 2007 · 104 · 7899
26. Chu · 2023 · 38 · 771
27. Protection of multiple ischemic organs by controlled reperfusion10.4103/bc.bc_59_21 ↗
28. Flow characteristics and coherent structures in a centrifugal blood pump10.1007/s10494-018-9994-3 ↗
29. Bioinspired shear-flow-driven layer-by-layer in situ self-assembly10.1021/acsnano.9b02010 ↗
30. Layered nanocomposites by shear-flow-induced alignment of nanosheets10.1038/s41586-020-2161-8 ↗
31. The intra‐arterial selective cooling infusion system: a mathematical temperature analysis and in vitro experiments for acute ischemic stroke therapy10.1111/cns.13883 ↗
32. Improvement of hemolysis performance in a hydrodynamically levitated centrifugal blood pump by optimizing a shroud size10.1007/s10047-020-01240-6 ↗
33. Evaluation of Eulerian and Lagrangian models for hemolysis estimation10.1097/MAT.0b013e318254833b ↗
34. Complications common to ventricular assist device support are rare with 90 days of DeBakey VAD support in calves10.1097/00002480-200105000-00026 ↗
35. A fluid dynamic analysis using flow visualization of the Baylor/NASA implantable axial flow blood pump for design improvement10.1111/j.1525-1594.1995.tb02306.x ↗
36. Minimization of hemolysis in centrifugal blood pumps: influence of different geometries10.1177/039139889301600705 ↗
37. Blood coagulation10.1016/S0140-6736(00)02225-X ↗
38. Shear-induced activation of platelets10.1016/0021-9290(79)90150-7 ↗
39. Enhanced hemocompatibility and antibacterial activity on titania nanotubes with tanfloc/heparin polyelectrolyte multilayers10.1002/jbm.a.36876 ↗
40. From unfractionated heparin to pentasaccharide: Paradigm of rigorous science growing in the understanding of the in vivo thrombin generation10.1016/j.blre.2019.100613 ↗
41. New anticoagulant coatings and hemostasis assessment tools to avoid complications with pediatric left ventricular assist devices10.1016/j.jtcvs.2017.03.149 ↗
42. A honokiol-mediated robust coating for blood-contacting devices with anti-inflammatory, antibacterial and antithrombotic properties
43. White · 2011
44. Heparin coating on 3D printed poly (l-lactic acid) biodegradable cardiovascular stent via mild surface modification approach for coronary artery implantation10.1016/j.cej.2019.122116 ↗
45. Fluid dynamics characterization and thrombogenicity assessment of a levitating centrifugal pump with different impeller designs10.1016/j.medengphy.2020.07.008 ↗
46. Reynolds stresses and hemolysis in turbulent flow examined by threshold analysis10.3390/fluids1040042 ↗
47. Evaluation of the hemolysis and fluid dynamics of a ventricular assist device under the pulsatile flow condition10.1007/s42241-018-0154-y ↗
48. Estimation of shear stress-related blood damage in heart valve prostheses--in vitro comparison of 25 aortic valves10.1177/039139889001300507 ↗
49. Numerical analysis of blood damage potential of the HeartMate II and HeartWare HVAD rotary blood pumps10.1111/aor.12542 ↗
50. An estimation method of hemolysis within an axial flow blood pump by computational fluid dynamics analysis10.1046/j.1525-1594.2003.00034.x ↗
51. Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump10.1177/0391398818785558 ↗