Long-term biocompatibility of a corrodible peripheral iron stent in the porcine descending aorta
Source abstract
Abstract
Currently there are no biodegradable stents available for treatment of vascular obstructions in patients with congenital heart defects. This study was performed to evaluate the safety of a corrodible stent produced from pure iron in a peripheral stent design (6-12mm diameter) in a slotted tube design similar to a commercially available 316-L stent which served as control. Both stents were implanted into the descending aorta of 29 minipigs with an overstretch injury without technical problems. Two animals died after the implantation not related to the iron stent. The remaining 27 minipigs were followed for 1-360 days. Histomorphometry and quantitative angiography showed no difference with regard to the amount of neointimal proliferation between 316-L and iron stents. Histopathological examination of heart, lung, spleen, liver, kidney and para-aortic lymphatic nodes demonstrated no signs of iron overload or iron-related organ toxicity. Adjacent to the iron stent struts, there was no evidence for local toxicity due to corrosion products. We conclude that iron is a suitable metal for the production of a large-size degradable stent with no local or systemic toxicity. A faster degradation rate, however, is desirable and further studies have to focus on the modification of the composition and design of the stent to expedite the degradation process.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Hydrogen embrittlement and corrosion behaviors in metals
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (42)
Show source-provided references
1. Transluminally placed coilspring endarterial tube grafts. Long-term patency in canine popliteal artery10.1097/00004424-196909000-00008 ↗
2. Intravascular stents to prevent occlusion and restenosis after transluminal angioplasty10.1056/NEJM198703193161201 ↗
3. Implantation of balloon-expandable intravascular grafts by catheterization in pulmonary arteries and systemic veins10.1161/01.CIR.77.1.188 ↗
4. Use of endovascular stents in congenital heart disease10.1161/01.CIR.83.6.1923 ↗
5. Intravascular and intracardiac stents used in congenital heart disease10.1097/00001573-200103000-00002 ↗
6. Post-intervention vessel remodeling10.1038/sj.gt.3301866 ↗
7. Redilation of endovascular stents in congenital heart disease: factors implicated in the development of restenosis and neointimal proliferation10.1016/S0735-1097(01)01406-1 ↗
8. A method for the study of undecalcified bones and teeth with attached soft tissues. The Sage-Schliff (sawing and grinding) technique10.1111/j.1600-0714.1982.tb00172.x ↗
9. Staining calcified tissues in plastic embedding medium
10. Interventional cardiac catheterization at Duke Medical Center10.1016/0002-9149(88)91529-9 ↗
11. Biodegradable intracoronary stents in adult dogs
12. Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries10.1161/01.CIR.94.7.1690 ↗
13. Collapse pressures of biodegradable stents10.1016/S0142-9612(02)00640-3 ↗
14. Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans10.1161/01.CIR.102.4.399 ↗
15. Macrophages related to dental implant failure10.1097/01.ID.0000041425.36813.A9 ↗
16. Corrosion behavior of a welded stainless-steel orthopedic implant10.1016/S0142-9612(00)00185-X ↗
17. Microvascular response of striated muscle to metal debris. A comparative in vivo study with titanium and stainless steel10.1302/0301-620X.85B1.12749 ↗
18. Observations at the articular surface of hip prostheses: an analytical electron microscopy study on wear and corrosion10.1002/1097-4636(20010315)54:4<591::AID-JBM160>3.0.CO;2-A ↗
19. Nickel and molybdenum contact allergies in patients with coronary in-stent restenosis10.1016/S0140-6736(00)03262-1 ↗
20. Why stainless steel corrodes10.1038/415770a ↗
21. Nickel release after implantation of the Amplatzer occluder10.1067/mhj.2003.7 ↗
22. Corroded nitinol wires in explanted aortic endografts: an important mechanism of failure?10.1177/152660280100800303 ↗
23. Dissolving tungsten coils lack systemic toxicity but lead to elevated serum tungsten levels and recanalization of the previously occluded vessel10.1017/S1047951102000513 ↗
24. Degradation of tungsten coils implanted into the subclavian artery of New Zealand white rabbits is not associated with local or systemic toxicity10.1016/S0142-9612(02)00352-6 ↗
25. A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal. results 6–18 months after implantation into New Zealand white rabbits10.1136/heart.86.5.563 ↗
26. Nickel and molybdenum contact allergies in patients with coronary in-stent restenosis10.1016/S0140-6736(00)03262-1 ↗
27. Corroded nitinol wires in explanted aortic endografts: an important mechanism of failure?10.1177/152660280100800303 ↗
28. Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology?10.1136/heart.89.6.651 ↗
29. Preliminary results after application of absorbable metal stents in patients with critical limb ischemia10.1583/04-1349R.1 ↗
30. Drug-eluting bioabsorbable magnesium stent10.1111/j.1540-8183.2004.04081.x ↗
31. Stent endothelialization. time course, impact of local catheter delivery, feasibility of recombinant protein administration, and response to cytokine expedition10.1161/01.CIR.95.2.438 ↗
32. The electrochemical aspects of the thrombogenicity of a material
33. Endothelial cell migration onto metal stent surfaces under static and flow conditions10.1615/JLongTermEffMedImplants.v10.i12.90 ↗
34. Acute and chronic tissue response to coronary stent implantation: pathologic findings in human specimen10.1016/S0735-1097(99)00486-6 ↗
35. Coronary stenting and inflammation10.1016/j.amjcard.2005.09.064 ↗
36. In vitro host response assessment of biomaterials for cardiovascular stent manufacture10.1023/B:JMSM.0000021123.51752.11 ↗
37. Effect of endovascular stent strut geometry on vascular injury, myointimal hyperplasia, and restenosis10.1067/mva.2002.122878 ↗
38. Morphological predictors of restenosis after coronary stenting in humans10.1161/01.CIR.0000019071.72887.BD ↗
39. Coronary stent symmetry and vascular injury determine experimental restenosis10.1136/heart.83.4.462 ↗
40. Breakable stent for interventions in infants and neonates. An animal study and histopathological findings10.1136/hrt.2005.062166 ↗
41. Molecular control of iron metabolism10.1016/j.beha.2004.10.004 ↗
42. Iron homeostasis and inherited iron overload disorders: an overview10.1016/j.hoc.2004.06.018 ↗