journal article·2010-05

In vitro evaluation of the long-term stability of polyimide as a material for neural implants

Birthe Rubehn, Thomas Stieglitz

Biomaterials· Volume 31· Issue 13
Publisher page

Abstract

In order to be used as a material for neural prostheses, polyimide has to withstand the body environment over years. To obtain more information about the long-term stability of this material, we tested three commercially available polyimides (PI2611 - HD-Microsystems (type: BPDA-PPD), U-Varnish-S - UBE (type: BPDA-PPD), Durimide 7510 - Fujifilm (type: information not provided)). Specimens were stored in phosphate buffered saline (PBS) at 37 degrees C to simulate body temperature and at elevated temperatures of 60 degrees C and 85 degrees C to accelerate aging. During the course of 20 months, stress-strain curves were measured monthly by tensile testing. From these curves the Young's modulus, the stress and strain at break, the stress at 10% strain as well as the fracture energy were extracted and used to characterize and to statistically evaluate the mechanical material properties. Mass loss was determined by weighing polyimide foils (Upilex25S - UBE) and optical measurements were conducted to examine changes in chemical structure and crystallinity. At 37 degrees C and 60 degrees C no change in material behavior could be observed, except for an increase of the Young's modulus of the BPDA-PPD type stored at 60 degrees C. This demonstrates the long-term stability of all tested polyimides with respect to PBS. All extracted quantities mentioned above, as well as the mass, decreased in specimens stored in PBS at 85 degrees C. As a comparison, BPDA-PPD type specimens stored at 85 degrees C in deionized water showed no change in any property but behaved similarly to the reference material.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Neuroscience and Neural Engineering · Conducting polymers and applications · Muscle activation and electromyography studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. Thin-film multielectrode arrays for a cochlear prosthesis10.1109/T-ED.1982.20671 ↗
2. A flexible perforated microelectrode array for extended neural recordings10.1109/10.108125 ↗
3. Micromachined, polyimide-based devices for flexible neural interfaces10.1023/A:1009955222114 ↗
4. A MEMS-based flexible multichannel ECoG-electrode array10.1088/1741-2560/6/3/036003 ↗
5. Aqueous degradation of polyimides10.1002/app.1971.070151206 ↗
6. Thermal regeneration of the tensile properties of hydrolytically degraded polyimide film10.1002/app.1972.070161116 ↗
7. Environmental aging and deadhesion of polyimide dielectric films10.1115/1.1773853 ↗
8. Effect of water on the properties of a glass-polyimide laminate10.1007/BF00754486 ↗
9. Ghosh · 1996
10. A biohybrid system to interface peripheral nerves after traumatic lesions: design of a high channel sieve electrode10.1016/S0956-5663(02)00019-2 ↗
11. Assessment of biocompatibility of chronically implanted polyimide and plantinum intrafascicular electrodes10.1109/TBME.2006.886617 ↗
12. Immunohistochemical characterization of axonal sprouting and reactive tissue changes after long-term implantation of a polyimide sieve electrode to the transected adult rat sciatic nerve10.1016/S0142-9612(00)00420-8 ↗
13. Polyimides as biomaterials: preliminary biocompatibility testing10.1016/0142-9612(93)90183-3 ↗
14. Assessment of the biocompatibility of photosensitive polyimide for implantable medical device use10.1002/jbm.a.32125 ↗
15. Campbell FJ, Brewer AK, Orr RJ, Janicke TA, Bruning AM. Hydrolytic deterioration of polyimide insulation on naval aircraft wiring. 1988 Annual Report: Conference on electrical insulation and dielectric phenomena, Oct.16–20, 1988, Ottawa, Ont., Canada; 1988. p. 180–8.10.1109/CEIDP.1988.26329 ↗
16. Bessonov · 1987
17. Morphologic and functional evaluation of peripheral nerve fibres regenerated through polyimide sieve electrodes over long-term implantation10.1002/jbm.10099 ↗
18. Long term assessment of axonal regeneration through polyimide regenerative electrodes to interface the peripheral nerve10.1016/j.biomaterials.2004.06.025 ↗
19. Investigations on explanted micromachined nerve electrodes
20. ASTM. 882-02 Standard test method for tensile properties of thin plastic sheeting; 2005.
21. General aging theory and simplified protocol for accelerated aging of medical devices
22. Scheirs · 2000
23. Birolini · 2007
24. Minges · 1989
25. 2009
26. Strobl · 2007
27. Grellmann · 2007
28. Thermal curing in polyimide films and coatings
29. A small-angle x-ray scattering study of an aromatic polyimide10.1002/pol.1984.180220613 ↗
30. Water-sorption behaviors of poly (3,4′-oxydiphenylene pyromellitimide) films depending on the thickness variation10.1002/1099-0488(20010315)39:6<669::AID-POLB1041>3.0.CO;2-# ↗
31. Anisotropic properties of high-temperature polyimide thin films: dielectric and thermal-expansion behaviors10.1063/1.351629 ↗
32. Surface-induced ordering of an aromatic polyimide10.1103/PhysRevLett.66.1181 ↗
33. The biodegradation of poly(ether urethanes)10.1002/jbm.820210908 ↗
34. Plasticizer motion and diffusion
35. Some effects of water immersion on the mechanical behavior of a polyimide film10.1115/1.2905704 ↗
36. Degradation and stability of polyimides
37. Water diffusion studies in polyimide thin films10.1002/app.1899 ↗
38. Synthesis of polyimides
39. Structural effects on the transport of water in polyimides10.1021/ma00092a011 ↗
40. Solubility of water in polyimides: quartz crystal microbalance measurements10.1002/polb.1991.090290111 ↗
41. Masterton · 2008