The influence of surface oxides on the distribution and release of nickel from Nitinol wires
Source abstract
Abstract
The patterns of Ni release from Nitinol vary depending on the type of material (Ni-Ti alloys with low or no processing versus commercial wires or sheets). A thick TiO(2) layer generated on the wire surface during processing is often considered as a reliable barrier against Ni release. The present study of Nitinol wires with surface oxides resulting from production was conducted to identify the sources of Ni release and its distribution in the surface sublayers. The chemistry and topography of the surfaces of Nitinol wires drawn using different techniques were studied with XPS and SEM. The distribution of Ni into surface depth and the surface oxide thickness were evaluated using Auger spectroscopy, TEM with FIB and ELNES. Ni release was estimated using either ICPA or AAS. Potentiodynamic potential polarization of selected wires was performed in as-received state with no strain and in treated strained samples. Wire samples in the as-received state showed low breakdown potentials (200 mV); the improved corrosion resistance of these wires after treatment was not affected by strain. It is shown how processing techniques affect surface topography, chemistry and also Ni release. Nitinol wires with the thickest surface oxide TiO(2) (up to 720 nM) showed the highest Ni release, attributed to the presence of particles of essentially pure Ni whose number and size increased while approaching the interface between the surface and the bulk. The biological implications of high and lasting Ni release are also discussed.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Metal and Thin Film Mechanics · Shape Memory Alloy Transformations · Electrodeposition and Electroless Coatings
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (27)
Show source-provided references
1. Biocompatibility of nickel–titanium metal and its corrosion behaviour in human cell cultures10.1002/(SICI)1097-4636(19970615)35:4<451::AID-JBM5>3.0.CO;2-G ↗
2. Electrochemical and surface characterization of NiTi alloy10.1016/S0142-9612(97)00210-X ↗
3. New oxidation treatment of NiTi shape memory alloys to obtain Ni-free surfaces and to improve biocompatibility10.1002/jbm.b.30441 ↗
4. Effect of surface treatment of NiTi alloy on its corrosion behavior in Hank's solution10.1002/jbm.10114 ↗
5. Diamond-like carbon coatings on orthodontic archwires10.1016/j.diamond.2004.11.036 ↗
6. Effect of diamond-like carbon (DLC) on the properties of NiTi alloys10.1016/j.diamond.2006.03.004 ↗
7. Influence of nitinol wire surface treatment on oxide thickness and composition and its subsequent effect on corrosion resistance and nickel ion release10.1002/jbm.a.30720 ↗
8. Surface conditions of nitinol wires, tubing, and as-cast alloys: the effect of chemical etching, aging n boiling water, and heat treatment10.1002/jbm.b.10001 ↗
9. Critical overview of nitinol surfaces and their modifications for medical applications10.1016/j.actbio.2008.01.013 ↗
10. Effect of nitinol wire surface properties on albumin adsorption10.1016/j.actbio.2006.07.006 ↗
11. The corrosion behavior of nickel titanium shape memory alloy10.1016/0010-938X(90)90004-O ↗
12. Moulder · 1992
13. Surface oxidation of NiTi shape memory alloys10.1016/S0142-9612(02)00244-2 ↗
14. Tian H, Schryvers D, Shabalovskaya S, Van Humbeeck J. Microstructure of surface and sub-surface layers of a Ni–Ti shape memory microwire. Microsc Microanal, in press, doi:10.1017/S1431927609090059.10.1017/S1431927609090059 ↗
15. Carroll W, Kelly M, Brien B. Corrosion behavior of Nitinol wires in body fluids environment. In: Proceedings of international conference on shape memory and superelastic technologies. Antwerpen, Belgium: 1999. p. 240–9.
16. Effect of compressive straining on corrosion resistance of a shape memory Ni–Ti alloy in Ringer's solution10.1002/(SICI)1097-4636(199612)32:4<583::AID-JBM11>3.0.CO;2-F ↗
17. Brown S. On methods used for corrosion testing of NiTi. In: Proceedings of international conference shape memory and superelastic technologies. Asilomar, CA, Pacific Grove: 2000. p. 271–9.
18. Shabalovskaya S, Anderegg J, Rondelli G, Xiong J. The effect of surface particulates on the corrosion resistance of Nitinol wire. In: Proceedings of international conference on shape memory and superelastic technologies. Asilomar, CA, Pacific Grove: 2003. p. 399–408.
19. Heßing C. Surface modification of NiTi shape memory alloys. Electrolyte treatments, high temperature oxidation and coatings. Ph.D. thesis, Bochum, 2007.
20. Electro discharge machining of nickel–titanium shape memory alloys10.1016/j.msea.2003.09.115 ↗
21. Continued expansion of nitinol self-expandable stent angiographic analysis and 1-year clinical follow-up10.1016/S0002-8703(99)70120-1 ↗
22. Long-term vessel response of self-expending coronary stent: a serial volumetric intravascular ultrasound analysis from the ASSURE trial10.1016/S0735-1097(01)01162-7 ↗
23. Early and late results of self-expandable nitinol stets: interim report from multicenter European study10.1111/j.1540-8183.1997.tb00032.x ↗
24. Self-expandable and highly flexible nitinol stent: immediate and long-term results in dogs10.1016/0002-8703(94)90582-7 ↗
25. Neointimal hyperplasia in low-profile nitinol stents, palmaz stents and wall stents: a comparative experimental study10.1007/BF02577644 ↗
26. Corroded nitinol wires in explanted aortic endografts: an important mechanism or failure?10.1177/152660280100800303 ↗
27. What can we learn from explanted endovascular devices?10.1053/ejvs.2002.1677 ↗