journal article·2004-12

Comparison of metal concentrations in rat tibia tissues with various metallic implants

Yoshimitsu Okazaki, Emiko Gotoh, Takeshi Manabe, Kihei Kobayashi

Biomaterials· Volume 25· Issue 28
Publisher page

Abstract

To compare metal concentrations in tibia tissues with various metallic implants, SUS316L stainless steel, Co-Cr-Mo casting alloy, and Ti-6Al-4V and V-free Ti-15Zr-4Nb-4Ta alloys were implanted into the rat tibia for up to 48 weeks. After the implant was removed from the tibia by decalcification, the tibia tissues near the implant were lyophilized. Then the concentrations of metals in the tibia tissues by microwave acid digestion were determined by inductively coupled plasma-mass spectrometry. Fe concentrations were determined by graphite-furnace atomic absorption spectrometry. The Fe concentration in the tibia tissues with the SUS316L implant was relatively high, and it rapidly increased up to 12 weeks and then decreased thereafter. On the other hand, the Co concentration in the tibia tissues with the Co-Cr-Mo implant was lower, and it increased up to 24 weeks and slightly decreased at 48 weeks. The Ni concentration in the tibia tissues with the SUS316L implant increased up to 6 weeks and then gradually decreased thereafter. The Cr concentration tended to be higher than the Co concentration. This Cr concentration linearly increased up to 12 weeks and then decreased toward 48 weeks in the tibia tissues with the SUS316L or Co-Cr-Mo implant. Minute quantities of Ti, Al and V in the tibia tissues with the Ti-6Al-4V implant were found. The Ti concentration in the tibia tissues with the Ti-15Zr-4Nb-4Ta implant was lower than that in the tibia tissues with the Ti-6Al-4V implant. The Zr, Nb and Ta concentrations were also very low. The Ti-15Zr-4Nb-4Ta alloy with its low metal release in vivo is considered advantageous for long-term implants.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (23)

Show source-provided references
1. Release and excretion of metal in patients who have a total hip-replacement component made of titanium-base alloy10.2106/00004623-199173100-00005 ↗
2. Metal release in patients who have had a primary total hip arthroplasty10.2106/00004623-199810000-00006 ↗
3. Metallic wear in failed titanium-alloy total hip replacements10.2106/00004623-198870030-00005 ↗
4. Nickel-, chrom- and cobalt-concentration in human tissue and body fluids of hip prosthesis patients
5. Metal determination in organic fluids of patients with stainless steel hip arthroplasty10.3109/17453678308992891 ↗
6. Cobalt, chromium, and nickel concentrations in body fluids of patients with porous-coated knee or hip prostheses10.1002/jor.1100070302 ↗
7. Metal release and excretion from cementless titanium alloy total knee replacements10.1097/00003086-199901000-00021 ↗
8. Implications of metallic corrosion in total knee arthroplasty10.1097/00003086-199110000-00004 ↗
9. Okazaki Y, Nishimura E. Comparison of metal release from metallic biomaterials in Vitro. Biomaterials, in press.
10. Light and electron microscopic studies of bone-titanium interface in the tibiae of young and mature rats10.1002/(SICI)1097-4636(199604)30:4<523::AID-JBM11>3.0.CO;2-I ↗
11. A comparative study of bone tissue reaction to Ti implants in the tibiae between mature and young rats10.2186/jjps.37.858 ↗
12. Surface analysis of Ti–15Zr–4Nb–4Ta alloy after implantation in rat tibia10.1016/S0142-9612(00)00221-0 ↗
13. Effects of Mo and Pd on corrosion resistance of V-free titanium alloys for medical implants10.2320/matertrans1989.38.344 ↗
14. Corrosion resistance of implant alloys in pseudo physiological solution and role of alloying elements in passive films10.2320/matertrans1989.38.78 ↗
15. Titanium alloys as implant materials
16. The mechanisms of passive dissolution of titanium in a model physiological environment10.1002/jbm.820260305 ↗
17. Passive dissolution of titanium in biological environments
18. Metal ion release after total hip replacement10.1016/0142-9612(80)90016-2 ↗
19. Elution and biocompatibility of titanium in the rabbit tissue
20. Local accumulation of titanium released from a titanium implant in the absence of wear10.1002/(SICI)1097-4636(199606)31:2<227::AID-JBM9>3.0.CO;2-P ↗
21. Metal release from plates used in jaw fracture treatment10.1016/S0901-5027(89)80102-X ↗
22. Increased blood cobalt and chromium after total hip replacement
23. Elevated serum cobalt with metal-on-metal articulating surfaces10.1302/0301-620X.79B2.0790316 ↗