journal article·2012-03

In vivo detection of glutathione disulfide and oxidative stress monitoring using a biosensor

Hui-Bog Noh, Pranjal Chandra, Jeon Ok Moon, Yoon-Bo Shim

Biomaterials· Volume 33· Issue 9
Publisher page

Abstract

A highly sensitive in vivo biosensor for glutathione disulfide (GSSG) is developed using covalently immobilized-glutathione reductase (GR) and -β-nicotinamide adenine dinucleotide phosphate (NADPH) on gold nanoparticles deposited on poly[2,2':5',2″-terthiophene-3'-(p-benzoic acid)] (polyTTBA). The fabricated biosensor was characterized with SEM, TEM, XPS, and QCM. Analytical parameters affecting the biosensor performance were optimized in terms of applied potential, NADPH:GR ratio, temperature, and pH. A linear calibration plot is obtained using chronoamperometry in the dynamic range between 0.1 μM and 2.5 mM of GSSG, with a detection limit of 12.5 ± 0.5 nM. The developed biosensor is applied to detect GSSG in a real plasma sample. A microbiosensor was applied to detect the in vivo GSSG concentration to monitor the oxidative stress caused by diquat and t-butyl hydroperoxide. The results obtained are reliable, implying a promising approach for a GSSG biosensor in clinical diagnostics and oxidative stress monitoring.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Sulfur Compounds in Biology · Electrochemical sensors and biosensors · Molecular Sensors and Ion Detection

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (29)

Show source-provided references
2. Interactions between biosynthesis, compartmentation and transport in the control of glutathione homeostasis and signalling10.1093/jexbot/53.372.1283 ↗
3. Altered lung antioxidant status in patients with mild asthma10.1016/S0140-6736(99)01812-7 ↗
4. Increased levels of oxidized glutathione in CD4+ lymphocytes associated with disturbed intracellular redox balance in human immunodeficiency virus type 1 infection10.1182/blood.V86.1.258.bloodjournal861258 ↗
5. Oxidative stress and endothelial function in chronic renal failure10.1681/ASN.V12122747 ↗
6. Assay for quantitative determination of glutathione and glutathione disulfide levels using enzymatic recycling method10.1038/nprot.2006.378 ↗
7. 3-Iodoacetylaminobenzanthrone as a fluorescent derivatizing reagent for thiols in high-performance liquid chromatography10.1016/j.aca.2004.03.002 ↗
8. Determination of glutathione disulfide levels in biological samples using thiol-disulfide exchanging agent, dithiothreitol10.1002/bmc.1083 ↗
9. Simultaneous determination of reduced and oxidized glutathione in freshly isolated rat hepatocytes and cardiomyocytes by HPLC with electrochemical detection10.1002/1099-0801(200011)14:7<468::AID-BMC10>3.0.CO;2-C ↗
10. Determination of glutathione and glutathione disulfide in hepatocytes by liquid chromatography with an electrode modified with functionalized carbon nanotubes10.1016/j.jchromb.2004.12.034 ↗
11. Catalytic cathodic stripping voltammetry of oxidized glutathione at a hanging mercury drop electrode in the presence of nickel ion10.1016/0039-9140(94)00232-H ↗
12. Simultaneous electrochemical determination of glutathione and glutathione disulfide at a nanoscale copper hydroxide composite carbon ionic liquid electrode10.1021/ac900501j ↗
13. Functionalized conducting polymer as an enzyme-immobilizing substrate: an amperometric glutamate microbiosensor for in vivo measurements10.1021/ac050558v ↗
14. Clinical applications of micro- and nanoscale biosensors
15. Disposable amperometric sensors for thiols with special reference to glutathione10.1002/elan.200804247 ↗
16. Direct DNA hybridization detection based on the oligonucleotide-functionalized conductive polymer10.1021/ac015572w ↗
17. Electrochemical detection of peroxynitrite using a biosensor based on a conducting polymer-manganese ion complex10.1021/ac102041u ↗
18. Triggering the redox reaction of cytochrome c on a biomimetic layer and elimination of interferences for NADH detection10.1016/j.biomaterials.2010.06.052 ↗
19. Electropolymerized self-assembled layer on gold nanoparticles: detection of inducible nitric oxide synthase in neuronal cell culture10.1021/ac2006558 ↗
20. Trace analysis of DNA: preconcentration, separation, and electrochemical detection in microchip electrophoresis using Au nanoparticles10.1021/ac0701177 ↗
21. Marked interanimal differences in susceptibility of Sprague-Dawley rats to diquat-induced oxidative stress in the liver: correlation with hepatic uptake of diquat10.1016/S0022-3565(25)10431-X ↗
22. Electrochemistry of conductive polymers. 22. Electrochemical and spectroelectrochemical studies of polyazulene growth and its characterization10.1149/1.1837954 ↗
23. Detection of daunomycin using phosphatidylserine and aptamer co-immobilized on Au nanoparticles deposited conducting polymer10.1016/j.bios.2011.04.060 ↗
24. Direct electrochemical oxidation of NADPH at a low potential on the carbon nanotube modified glassy carbon electrode10.1002/cjoc.20040220212 ↗
25. Biosensing approach for glutathione detection using glutathione reductase (GR) with multi-walled carbon nanotubes on gold electrode10.1016/S1452-3981(23)18304-0 ↗
26. Hepatic glutathione reductase. I. Purification and general kinetic properties10.1016/S0021-9258(19)83745-6 ↗
27. A microelectrode biosensor for real time monitoring of L-glutamate release10.1016/j.aca.2009.04.048 ↗
28. Plasma glutathione and glutathione disulfide in the rat: regulation and response to oxidative stress10.1016/S0022-3565(25)22134-6 ↗
29. Glutathione disulfide as index of oxidant stress in rat liver during hypoxia