journal article·2008-12

Notched stress–strain behavior of a conventional and a sequentially annealed highly crosslinked UHMWPE

Michael C. Sobieraj, Steven M. Kurtz, A. Wang, Michael M. Manley, Clare M. Rimnac

Biomaterials· Volume 29· Issue 35
Publisher page

Abstract

Contemporary total joint replacement designs contain stress-risers such as fillets, grooves, and undercuts; therefore, it is of interest to analyze the behavior of UHMWPEs in the presence of such design-related stress-risers. This study examined the engineering and true axial stress-strain behavior of smooth cylindrical and notched cylindrical test specimens, under applied axial tensile loading (2 displacement rates, 37 degrees C) for a conventional and a highly crosslinked second generation UHMWPE. Both materials were prepared from ram extruded GUR 1050. The conventional material (30 kGy) was gamma sterilized at 30 kGy in an inert N(2) environment. The sequentially annealed material (SA) was gamma irradiated at 30 kGy and annealed for 8h at 130 degrees C. The irradiation-annealing process was repeated two more times for an overall irradiation dose of 90 kGy. Differential scanning calorimetry (DSC) was utilized to investigate changes in crystallinity and lamellar thickness distributions upon loading. Fractographic analysis of scanning electron microscope (SEM) images of fracture surfaces was performed to investigate changes in fracture micromechanism with notching. Both the 30 kGy and SA materials, in the smooth condition, demonstrated substantial ductility and orientation hardening. With the introduction of a notch, both materials demonstrated an elevation in the yield stress (notch strengthening) and a reduction in the ultimate stress and ultimate strain at both displacement rates. Additionally, it was found that the uniaxial stress-state (smooth condition) allowed for greater changes in crystallinity and the lamellar thickness distributions, when compared to the untested materials, than the triaxial stress-state induced by the notched geometry.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Advanced ceramic materials synthesis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (30)

Show source-provided references
1. Kurtz S, Lau E, Zhao K, Mowat F, Ong K, Halpern M. Projections of Primary and Revision Hip and Knee Arthroplasties in the United States from 2005 to 2030. 73rd Annual Meeting of the American Academy of Orthopaedic Surgeons, 2006, Chicago, IL; 2006. p. Exhibit No. SE53.10.2106/00004623-200704000-00012 ↗
2. Musculoskeletal conditions in the United States
3. The Role of macrophages in osteolysis of total joint replacements10.1016/j.biomaterials.2004.04.035 ↗
4. 2004
5. Development of an extremely wear-resistant ultra-high molecular weight polyethylene for total hip replacement10.1002/jor.1100170203 ↗
6. Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra-high molecular weight polyethylene10.1016/j.biomaterials.2004.03.022 ↗
7. Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene
8. Interaction of oxidation and crosslinking in gamma-irradiated ultra-high molecular weight polyethylene10.1002/jbm.10203 ↗
9. Basis for a second generation highly crosslinked UHMWPE10.1097/01.blo.0000238856.61862.7d ↗
10. Wear, oxidation and mechanical properties of a sequentially irradiated and annealed UHMWPE in total joint replacement10.1088/0022-3727/39/15/S12 ↗
11. Notch strengthening and hardening behavior of conventional and highly crosslinked UHMWPE under applied tensile loading10.1016/j.biomaterials.2004.09.031 ↗
12. Crystal thickness distributions from melting homopolymers or random copolymers10.1002/(SICI)1099-0488(19991101)37:21<3131::AID-POLB22>3.0.CO;2-M ↗
13. Characterization of physical, chemical, and mechanical properties of UHMWPE
14. The effect of dose rate on the crystalline lamellar thickness distribution in gamma-radiation of UHMWPE10.1016/j.nimb.2005.04.037 ↗
15. Mechanical spectrometry of alpha-relaxations of high-density polyethylene10.1002/polb.1990.090280410 ↗
16. Multiple comparison procedures: the practical solution10.1080/00031305.1990.10475712 ↗
17. Failure micromechanisms during uniaxial tensile fracture of conventional and highly crosslinked ultra-high molecular weight polyethylenes used in total joint replacement10.1016/S0142-9612(03)00277-1 ↗
18. Thermomechanical behavior of virgin and highly crosslinked ultra-high molecular weight polyethylene used in total joint replacements10.1016/S0142-9612(02)00102-3 ↗
19. Orientation of plane strain-compressed ultra-high molecular weight polyethylene10.1016/S0032-3861(98)00163-3 ↗
20. Time resolved simultaneous small- and wide-angle X-ray scattering during polyethylene deformation-II. Cold drawing of linear polyethylene10.1016/S0032-3861(97)00226-7 ↗
21. Large deformation compression induced crystallinity degradation of conventional and highly crosslinked UHMWPEs10.1016/j.biomaterials.2005.04.008 ↗
22. Analysis of retrieved implants: crystallinity changes in ultra high molecular weight polyethylene10.1002/jbm.820160408 ↗
23. Phase transformation in explanted crystalline UHMWPE acetabular cups and debris after in vivo wear10.1016/j.molstruc.2005.09.029 ↗
24. Courtney · 2000
25. The effects of morphology on the yield behaviour of polyethylene copolymers10.1016/S0032-3861(00)00762-X ↗
26. A dislocation model for yield in polyethylene10.1080/14786439808206535 ↗
27. Microstructural changes induced in linear polyethylene by plastic deformation (rolling)10.1007/BF01124822 ↗
28. Wunderlich · 1980
29. Fractography evolution in accelerated aging of UHMWPE after gamma irradiation in air10.1016/S0142-9612(03)00464-2 ↗
30. Notch sensitivity and fractography of polyolefins10.1002/pen.760230303 ↗