Tight medial knot tying may increase retearing risk after transosseous equivalent repair of rotator cuff tendon
Source abstract
Abstract
BACKGROUND: Retearing mechanism after transosseous equivalent (TOE) rotator cuff repair has not been fully clarified yet. OBJECTIVE: The purposes of this study were to compare the stress distribution pattern in the tendon stump between knotted and knotless TOE repair and to investigate the role of suture tension applied during medial knot tying using a 3-dimensional finite element (3D-FE) method. METHODS: Both knotted and knotless TOE repairs were simulated on the 3D-FE human rotator cuff tear model. Elastic analysis was performed to compare the stress distribution pattern inside the tendon between the two models. The amount of compressive load applied to the medial-row sutures was then changed as 0, 20, 40, 60, 80, and 100 N in the knotted model. RESULTS: Knotted model demonstrated more distinct stress concentration inside the tendon around medial-row sutures than the knotless model. Mean von Mises equivalent stress in this area in the 0, 20, 40, 60, 80, and 100 N models was 0.26, 0.35, 0.50, 0.70, 1.11, and 1.14 MPa, respectively. CONCLUSIONS: In the knotted TOE repair, tight medial knot tying might cause a high stress concentration around medial knots, which may constitute one of the pathogenetic factors of postoperative retearing at this site.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Shoulder Injury and Treatment · Orthopedic Surgery and Rehabilitation · Shoulder and Clavicle Injuries
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Division of Orthopedics, Sendai City Hospital, Sendai, Japan
Strategic Innovation Research Hub Laboratory of Strength of Materials and Science, Teikyo University, Tokyo, Japan
References (40)
Show source-provided references
1. Rotator cuff: Biology and current arthroscopic techniques10.1007/s00167-012-1901-1 ↗
2. Arthroscopic rotator cuff repairs: An anatomic and biomechanical rationale for different suture-anchor repair configurations10.1016/j.arthro.2007.02.018 ↗
3. Which method of rotator cuff repair leads to the highest rate of structural healing? A systematic review10.1177/0363546509359679 ↗
4. “Transosseous-equivalent” rotator cuff repair technique10.1016/j.arthro.2006.07.017 ↗
5. Part II: Biomechanical assessment for a footprint-restoring transosseous-equivalent rotator cuff repair technique compared with a double-row repair technique10.1016/j.jse.2006.09.011 ↗
6. Part I: Footprint contact characteristics for a transosseous-equivalent rotator cuff repair technique compared with a double-row repair technique10.1016/j.jse.2006.09.010 ↗
7. Retear patterns after arthroscopic rotator cuff repair: Single-row versus suture bridge technique10.1177/0363546509350081 ↗
8. Arthroscopic rotator cuff repair using a suture bridge technique: Is the repair integrity actually maintained?10.1177/0363546510397171 ↗
9. Arthroscopic single-row modified Mason-Allen repair versus double-row suture bridge reconstruction for supraspinatus tendon tears: A matched-pair analysis10.1177/0363546512462123 ↗
10. Clinical and radiographic outcomes after arthroscopic repair of massive rotator cuff tears using a suture bridge technique: Assessment of repair integrity on magnetic resonance imaging10.1177/0363546511434546 ↗
11. Arthroscopic knotless-anchor rotator cuff repair: A clinical and radiological evaluation10.1007/s00167-014-3026-1 ↗
12. Arthroscopic rotator cuff repair using modified Mason–Allen medial row stitch: Knotless versus knot-tying suture bridge technique10.1177/0363546512459170 ↗
13. Arthroscopic rotator cuff repair using the opus knotless suture anchor fixation system10.1177/0363546508331203 ↗
14. Transosseous-equivalent rotator cuff repair: A systematic review on the biomechanical importance of tying the medial row10.1016/j.arthro.2012.11.008 ↗
15. A biomechanical comparison of 2 technical variations of double-row rotator cuff fixation: The importance of medial row knots10.1177/0363546507312640 ↗
16. Biomechanical characteristics of the horizontal mattress stitch: Implication for double-row and suture-bridge rotator cuff repair10.1007/s00776-013-0504-0 ↗
17. Functional and structural outcomes of single-row versus double-row versus combined double-row and suture-bridge repair for rotator cuff tears10.1177/0363546511415660 ↗
18. Effect of shoulder abduction angle on biomechanical properties of the repaired rotator cuff tendons with 3 types of double-row technique10.1177/0363546510388152 ↗
19. Prediction of strength and strain of the proximal femur by a CT-based finite element method10.1016/j.jbiomech.2006.08.003 ↗
20. Prediction of the strength and fracture location of the femoral neck by CT-based finite-element method: A preliminary study on patients with hip fracture10.1007/s00776-004-0824-1 ↗
21. Nonlinear stress analysis of the supraspinatus tendon using three-dimensional finite element analysis10.1007/s00167-012-2008-4 ↗
22. Finite element study on the anatomic transtibial technique for single-bundle anterior cruciate ligament reconstruction10.1007/s11517-015-1372-x ↗
23. Stress distribution in the supraspinatus tendon after tendon repair: Suture anchors versus transosseous suture fixation10.1177/0363546506296310 ↗
24. A 3D finite element model for geometrical and mechanical comparison of different supraspinatus repair techniques10.1016/j.jse.2015.09.002 ↗
25. Stress distribution within rotator cuff tendons with a crescent-shaped and an L-shaped tear10.1177/0363546513497565 ↗
26. Prediction of femoral fracture load using automated finite element modeling10.1016/S0021-9290(97)00123-1 ↗
27. A biomechanical study on the senile lntracapsular fracture of the femoral neck-speculation concerning the fracture lines and etiological factors
28. Preoperative assessment of the cancellous bone mineral density of the proximal humerus using CT data10.1007/s00256-011-1174-7 ↗
29. Tendons, ligaments, and capsule of the rotator cuff. Gross and microscopic anatomy10.2106/00004623-199274050-00010 ↗
30. Degeneration at the insertion weakens the tensile strength of the supraspinatus tendon: A comparative mechanical and histologic study of the bone-tendon complex10.1002/jor.1100150514 ↗
31. Non-mineralized fibrocartilage shows the lowest elastic modulus in the rabbit supraspinatus tendon insertion: Measurement with scanning acoustic microscopy10.1016/j.jse.2005.12.008 ↗
32. Material properties of common suture materials in orthopaedic surgery
33. Comparative evaluation on three-dimensional finite element models of the temporomandibular joint10.1016/j.clinbiomech.2007.12.011 ↗
34. Medial-row failure after arthroscopic double-row rotator cuff repair10.1016/j.arthro.2009.07.022 ↗
35. Characteristic retear patterns assessed by magnetic resonance imaging after arthroscopic double-row rotator cuff repair10.1016/j.arthro.2011.09.006 ↗
36. Arthroscopic single-row versus double-row suture anchor rotator cuff repair10.1177/0363546505279575 ↗
37. Medial rotator cuff failure after arthroscopic double-row rotator cuff repair10.1016/j.arthro.2008.03.009 ↗
38. Biomechanical comparison of 4 double-row suture-bridging rotator cuff repair techniques using different medial-row configurations10.1016/j.arthro.2010.02.013 ↗
39. Arthroscopic suture-bridge repair for small to medium size supraspinatus tear: Healing rate and retear pattern10.1016/j.arthro.2012.06.020 ↗
40. Biomechanical effects of a 2 suture-pass medial inter-implant mattress on transosseous-equivalent rotator cuff repair and considerations for a “technical efficiency ratio”10.1016/j.jse.2013.06.019 ↗