Mechanical loading characteristics of total hip prosthetics subjected to dynamic loading cycles

M. Kalayarasan, S. Shankar, M. Manikandan, K. Adithan

Publisher page

Abstract

The selection of best material pair in the hip prosthetics design for improved performance and life relies on the estimation of hip joint contact stresses and contact pressure distribution during various dynamic loading cycles: Climbing Upstairs, Climbing downstairs and Knee bending. The maximum Von Mises stress, contact pressure and deformation are considered factors in selecting the material pair in this current study. This is done by analysis of a three-dimensional finite element model of the acetabular component during the different dynamics cycles using ANSYS®. The different material combination of bearing couples considered for this analysis are metal in contact with plastic, metal on metal, metal on ceramic, ceramic on plastic, ceramic on metal and ceramic on ceramic. The numerical results were validated by comparing them with the FEA results of Hai-Bo Jiang et al. for the existing material combinations and a high correlation of 92% was observed. We found that the Alumina femoral head paired with ultra-high molecular weight polyethylene (UHMWPE) cup reduces the maximum Von Mises stress and maximum contact pressure developed at the interface amongst other material pairs.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

, , , India

References (42)

Show source-provided references
1. Campbell · 2001
2. Tate · 2009
3. Schünke · 2006
4. D.P. Byrne, K.J. Mulhall and Baker J.F., Anatomy & biomechanics of the hip, The Open Sports Medicine Journal 4(1) (2010).10.2174/1874387001004010051 ↗
5. Total hip arthroplasty–a cure?
6. Total hip arthroplasty
7. Morlock · 2011 · 1110.1007/978-3-642-19429-0_2 ↗
8. Posterior hip dislocation and ipsilateral isolated femoral neck fracture: A novel fixation method10.1093/jscr/2012.10.14 ↗
9. Villarraga · 2001 · 33
10. Biomechanics of the hip
11. Endurance verification of custom-made hip prostheses10.1016/S0142-1123(00)00056-6 ↗
12. A finite element analysis of hollow stemmed hip prostheses as a means of reducing stress shielding of the femur10.1016/S0021-9290(01)00072-0 ↗
13. Three-dimensional finite element analyses of functionally graded femoral prostheses with different geometrical configurations10.1016/j.matdes.2013.12.054 ↗
14. Static, dynamic and fatigue behavior of newly designed stem shapes for hip prosthesis using finite element analysis10.1016/j.matdes.2006.02.015 ↗
15. Probabilistic design of a newly designed cemented hip prosthesis using finite element method10.1016/j.matdes.2007.03.024 ↗
16. Hip implants VII: Finite element analysis and optimization of cross-sections10.1016/j.matdes.2007.09.002 ↗
17. Primary stability of an anatomical cementless hip stem: A statistical analysis10.1016/j.jbiomech.2005.03.024 ↗
18. Finite element modelling of primary hip stem stability: The effect of interference fit10.1016/j.jbiomech.2007.10.009 ↗
19. Finite element analysis of hip stem designs10.1016/j.matdes.2006.12.014 ↗
20. The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method10.1016/j.matdes.2006.08.012 ↗
21. A new method to analyse the mechanical behaviour of skeletal parts10.3109/17453677208998949 ↗
22. A review of stairway falls and stair negotiation: Lessons learned and future needs to reduce injury10.1016/j.gaitpost.2016.06.030 ↗
23. Loading of hip measured by hip contact forces at different speeds of walking and running10.1002/jbmr.2483 ↗
24. Material selection of acetabular component in human hip prosthesis using finite element concepts10.1504/IJECB.2013.056519 ↗
25. Finite element analysis of different contact bearing couples for human hip prosthesis10.1504/IJBET.2013.053712 ↗
26. Dynamic contact analysis of total hip prosthesis during normal active walking cycle10.1504/IJBET.2014.062742 ↗
27. Dynamic contact analysis of total hip prosthesis during stumbling cycle10.1142/S0219519414500419 ↗
28. Failure analysis on retrieved ultra high molecular weight polyethylene (UHMWPE) acetabular cups10.1016/j.engfailanal.2006.11.005 ↗
29. Biotribology of artificial hip joints10.5312/wjo.v6.i1.77 ↗
30. Life estimation of hip joint prosthesis10.1007/s40032-014-0159-4 ↗
31. Metal-on-metal hip arthroplasty: A review of adverse reactions and patient management10.3390/jfb6030486 ↗
32. The lubrication performance of the ceramic-on-ceramic hip implant under starved conditions10.1016/j.jmbbm.2015.06.001 ↗
33. Fluid load support and contact mechanics of hemiarthroplasty in the natural hip joint10.1016/j.medengphy.2010.09.009 ↗
34. The effect of acetabular cup size on the short-term stability of revision hip arthroplasty: A finite element investigation10.1243/0954411041560992 ↗
35. An axisymmetric contact model of ultra high molecular weight polyethylene cups against metallic femoral heads for artificial hip joint replacements10.1243/0954411991535158 ↗
36. Finite element algorithm reproducing hip squeak measured in experiment10.1016/j.jsv.2017.01.015 ↗
37. Optimization of femoral prosthesis based on comprehensive evaluation of structure and material properties10.1142/S0219519416400133 ↗
38. Comparison of finite element models of natural hip joint10.1016/j.jbiomech.2011.02.031 ↗
39. Static and dynamic mechanics analysis on artificial hip joints with different interface designs by the finite element method10.1016/S1672-6529(07)60024-9 ↗
40. Finite element simulation of the hip joint during stumbling: A comparison between static and dynamic loading10.1016/S0924-0136(03)00352-2 ↗
41. Hip contact forces and gait patterns from routine activities10.1016/S0021-9290(01)00040-9 ↗
42. Computational wear assessment of hard on hard hip implants subject to physically demanding tasks10.1007/s11517-017-1739-2 ↗