2D ultrasonic elastography with lateral displacement estimation using statistics
Source abstract
Abstract
Ultrasound elastography is the method of obtaining relative stiffness information of biological tissue, which plays an important role in early diagnosis. Generally, a gradient-based strain imaging algorithm assumes that motion only occurs in an axial direction. However, because tissue has different relative stiffness, the scatter presents lateral motion under high freehand compression. Therefore, errors occur in estimating the cross-correlation phase in the calculation window. A 2D elastography algorithm with lateral displacement estimation using statistics was proposed to reduce errors. The new method was investigated through simulation, and the experiment confirmed that errors introduced by lateral tissue movement have been greatly reduced with no sacrifice of real-time ultrasonic imaging quality.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ultrasound Imaging and Elastography · Ultrasonics and Acoustic Wave Propagation · Elasticity and Material Modeling
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Computer Science and Technology, Changsha College, Changsha 410003, China
School of Computer Science, Sichuan University, Chengdu, Sichuan 610065, China
Detection & Guidance Solution, GE Healthcare, Chengdu, Sichuan 611731, China
References