journal article·2026-02-16Open access

Bio-Compatibility Analysis of Newly Developed Plug and Cuff Electrodes for Future Neuronal Interface Applications

Eleni Zingkou, Georgios Pampalakis, Asimina Kolianou, Nafsika Rossopoulou, Aikaterini Skiada, Lydia Galouni, Patryk Śniarowski, Longina Madej-Kiełbik, Georgia Sotiropoulou, Karolina Gzyra-Jagieła, Theodora Katsila, Carmen Moldovan, Marian Ion, Octavian Narcis Ionescu, Eduard Franti, David Dragomir, Gerd Siekmeyer, Patrick Grotemeyer

Biomimetics· Volume 11· Issue 2

Abstract

The NerveRepack project is a European initiative that aims to develop biomimetic exoskeletons/exoprostheses for amputated or paralyzed leg patients that will receive and transmit signals to enable movements and sensations for the patient. To implement the project, it is fundamental to develop implantable neuronal electrodes that will allow bidirectional signaling between the sensors placed on the exoskeletons/exoprostheses and the nervous system. In this direction, two electrodes, plug and cuff, have been designed as integral parts of the final implantable device. The electrodes should comply with strict regulations to ensure their safe implantation in patients. The purpose of this study was to support the compliance of the implant platforms of certain key components with the ISO and ASTM standards that would be required for clinical applications. We have used an indirect method to assess the biocompatibility of the developed electrodes against neuronal cells, fibroblasts, and keratinocytes. Also, we assessed hemocompatibility, i.e., the potential of implantable electrodes to induce hemolysis or complement activation. Finally, the mutagenic/genotoxic potential was tested against the internationally recommended CHO cells. Both representative plug and cuff electrode components were found non-cytotoxic, non-mutagenic, and unable to induce hemolysis. Therefore, from the point of early evaluation of in vitro material and process biocompatibility, the selected implant platforms for the electrodes could be implanted in preclinical models to delineate their potential in vivo applications as neuronal interface with the biomimetic exoskeleton/exoprostheses.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Neuroscience and Neural Engineering · Muscle activation and electromyography studies · Prosthetics and Rehabilitation Robotics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Pharmacy, School of Health Sciences, University of Patras, 26504 Rion-Patras, Greece

Institute of Chemical Biology, National Hellenic Research Foundation, 48 Vasileos Constantinou Avenue, 11635 Athens, Greece

Łukasiewicz Research Network, Łódź Institute of Technology, Marii Skłodowskiej-Curie 19/27 St, 90-570 Łódź, Poland

Interdisciplinary Doctoral School, Łódź University of Technology, Żeromskiego 116 St, 90-924 Łódź, Poland

IMT Bucharest, 77190 Bucharest, Romania

Acquandas-Kaiserstraße 2, Building F, 1st floor room F-021, 24143 Kiel, Germany

Funding

European Union’s Horizon Europe

References (40)

Show source-provided references
1. Long-term stability strategies of deep brain flexible neural interface10.1038/s41528-025-00410-x ↗
2. Peripheral nerve bionic interface: A review of electrodes10.1007/s41315-019-00086-3 ↗
3. Feng, C., Frewin, C.L., Tanjil, M.R.E., Everly, R., Bieber, J., Kumar, A., Wang, M.C., and Saddow, S.E. (2021). A flexible a-sic-based neural interface utilizing pyrolyzed-photoresist film (C) active sites. Micromachines, 12.10.3390/mi12070821 ↗
4. Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: A review10.1186/s12984-020-00667-5 ↗
5. Multidimensional advances in neural interface technology for peripheral nerve repair: From material innovation to clinical translation10.1016/j.mtbio.2025.102092 ↗
6. Implanted high density cuff electrodes functionally activate human tibial and peroneal motor units without chronic detriment to peripheral nerve health10.1111/ner.13110 ↗
7. Recent advances in electrical neural interface engineering: Minimal invasiveness, longevity, and scalability10.1016/j.neuron.2020.10.011 ↗
8. Multifunctional nanomaterials for advancing neural interfaces: Recording, stimulation, and beyond10.1021/acs.accounts.4c00138 ↗
9. A review: Flexible devices for nerve stimulation10.20517/ss.2023.36 ↗
10. Long-term stability of neural signals from microwire arrays implanted in common marmoset motor cortex and striatum10.1088/2057-1976/aada67 ↗
11. Rokaya, D., Skallevold, H.E., Srimaneepong, V., Marya, A., Shah, P.K., Khurshid, Z., Zafar, M.S., and Sapkota, J. (2023). Shape Memory Polymeric Materials for Biomedical Applications: An Update. J. Compos. Sci., 7.10.3390/jcs7010024 ↗
12. Stretchable, dynamic covalent polymers for soft, long-lived bioresorbable electronic stimulators designed to facilitate neuromuscular regeneration10.1038/s41467-020-19660-6 ↗
13. Nanoporous gold as a neural interface coating: Effects of topography, surface chemistry, and feature size10.1021/acsami.5b00410 ↗
14. (2022). Biological Evaluation of Medical Devices. Part 2 “Animal Welfare Requirements” 3rd Edition (Standard No. ISO 10993).
15. (2014). Biological Evaluation of Medical Devices. Part 3 “Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity 3rd Edition (Standard No. ISO 10993).
16. (2009). Biological Evaluation of Medical Devices. Part 5 “Tests for In Vitro Cytotoxicity” 3rd Edition (Standard No. ISO 10993).
17. (2021). Biological Evaluation of Medical Devices. Part 12 “Sample Preparation and Reference Materials” 5th Edition (Standard No. ISO 10993).
18. (2018). Standard Guide for Performance of Chinese Hamster Ovary Cell/Hypoxanthine Guanine Phosphoribosyl Transferase Gene Mutation Assay (Standard No. ASTM E1262-88).
19. (2018). Standard Practice for Testing for Whole Complement Activation in Serum by Solid Materials (Standard No. ASTM F1984-99).
20. (2017). Standard Practice for Assessment of Hemolytic Properties of Materials (Standard No. ASTM F756-17).
21. (2025). Standard Practice for Extraction of Materials Used in Medical Devices (Standard No. ASTM F619-20).
22. The European Medical Device Regulation-What Biomedical Engineers Need to Know10.1109/JTEHM.2022.3194415 ↗
23. The Challenges of Setting Up a Clinical Study with the New European Union Medical Device Regulation10.20344/amp.19423 ↗
24. Hospitals as medical device manufacturers: Keeping to the Medical Device Regulation (MDR) in the EU10.1136/bmjinnov-2023-001150 ↗
25. A soft and stretchable multielectrode cuff for selective peripheral nerve stimulation10.1002/admt.202201322 ↗
26. Różańska, A., Walkowicz, M., Bulanda, M., Kasperski, T., Synowiec, E., Osuch, P., and Chmielarczyk, A. (2023). Evaluation of the efficacy of UV-C radiation in eliminating microorganisms of special epidemiological importance from touch surfaces under laboratory conditions and in the hospital environment. Healthcare, 11.10.3390/healthcare11233096 ↗
27. Multiple application of Alamar blue as an indicator of metabolic function and cellular health in cell viability bioassays10.3390/s120912347 ↗
28. Considerations for the use of SH-SY5Y neuroblastoma cells in neurobiology10.1007/978-1-62703-640-5_2 ↗
29. Properties of differentiated SH-SY5Y grown on carbon-based materials10.1039/D0RA03383A ↗
30. Behaviour of SH-SY5Y neuroblastoma cell line grown in different media and on different chemically modified substrates10.1016/j.biomaterials.2007.02.022 ↗
31. Material matters: Degradation products affect regenerating Schwann cells10.1016/j.bioadv.2024.213825 ↗
32. Biocompatibility and bioactivity of designer self-assembling nanofiber scaffold containing FGL motif for rat dorsal root ganglion neurons10.1002/jbm.a.32910 ↗
33. Adenovirus activates complement by distinctly different mechanisms in vitro and in vivo: Indirect complement activation by virions in vivo10.1128/JVI.00082-09 ↗
34. Mödinger, Y., Teixeira, G.Q., Neidlinger-Wilke, C., and Ignatius, A. (2018). Role of the complement system in the response to orthopedic biomaterials. Int. J. Mol. Sci., 19.10.3390/ijms19113367 ↗
35. Hemolysis, free hemoglobin toxicity, and scavenger protein therapeutics10.1182/blood.2022015596 ↗
36. Differential sensitivity of Chinese hamster V79 and Chinese hamster ovary (CHO) cells in the in vitro micronucleus screening assay10.1016/S1383-5718(01)00199-1 ↗
37. Zingkou, E., Kolianou, A., Angelis, G., Lykouras, M., Orkoula, M., Pampalakis, G., and Sotiropoulou, G. (2025). Cytocompatibility study of stainless steel 316l against differentiated SH-SY5Y cells. Biomimetics, 10.10.3390/biomimetics10030169 ↗
38. Ionescu, O.N., Franti, E., Carbunaru, V., Moldovan, C., Dinulescu, S., Ion, M., Dragomir, D.C., Mihailescu, C.M., Lascar, I., and Oproiu, A.M. (2024). System of implantable electrodes for neural signal acquisition and stimulation for wirelessly connected forearm prosthesis. Biosensors, 14.10.3390/bios14010031 ↗
39. (2025). Biological Evaluation of Medical Devices. Part 1 “Requirements and General Principles for the Evaluation of Biological Safety Within a Risk Management Process” 5th Edition (Standard No. ISO 10993).
40. (2017). Biological Evaluation of Medical Devices. Part 4 “Selection of Tests for Interactions with Blood” 3rd Edition (Standard No. ISO 10993).