Standardized Testing and Quantitative Safety Assessment for Upper Limb Rehabilitation Robots: A Bionic Robotic Platform and Integrated Evaluation Framework
Source abstract
Abstract
To address the lack of standardized safety assessment tools for upper-limb rehabilitation robots, this study developed an integrated testing platform and a quantitative safety assessment framework, demonstrated with FlexoArm1 as a proof-of-concept. A 6-degree-of-freedom bionic arm equipped with multiple sensors was constructed, and a fuzzy PID control algorithm was employed to improve motion trajectory tracking accuracy. A fuzzy multi-criteria safety assessment model was established by combining the Analytic Hierarchy Process (AHP) with the entropy weight method. Experiments were conducted on the rehabilitation robot FlexoArm1. The platform reliably replaced human subjects in range-of-motion testing, interactive torque measurement (peak torque approximately 6.2 N·m in fully active mode), and spasticity simulation, with angular data showing close agreement with Inertial Measurement Unit (IMU) measurements. The assessment model assigned a comprehensive safety score of 70.23 to the tested device, successfully identifying weaknesses in fault detection capability and structural safety design. The proposed bionic-arm-based testing platform and the accompanying safety assessment methodology provide practical tools and a quantitative basis for standardizing safety evaluation and guiding design optimization of upper-limb rehabilitation robots.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Stroke Rehabilitation and Recovery · Prosthetics and Rehabilitation Robotics · Motor Control and Adaptation
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
Shanghai Engineering Research Center of Assistive Devices, Shanghai 200093, China
Funding
National Natural Science Foundation of China
References (39)
Show source-provided references
1. Global and regional burden of stroke during 1990–2010: Findings from the Global Burden of Disease Study 201010.1016/S0140-6736(13)61953-4 ↗
2. Advances and challenges in stroke rehabilitation10.1016/S1474-4422(19)30415-6 ↗
3. Part I: Frequency of depression after stroke: An updated systematic review and meta-analysis of observational studies10.1111/ijs.12357 ↗
4. Post-stroke depression: Mechanisms and pharmacological treatment10.1016/j.pharmthera.2017.11.005 ↗
5. Stroke rehabilitation: Strategies to enhance motor recovery10.1146/annurev.med.60.042707.104248 ↗
6. Qassim, H.M., and Wan Hasan, W.Z. (2020). A review on upper limb rehabilitation robots. Appl. Sci., 10.10.3390/app10196976 ↗
7. A scoping review of design requirements for a home-based upper limb rehabilitation robot for stroke10.1080/10749357.2021.1943797 ↗
8. System framework of robotics in upper limb rehabilitation on poststroke motor recovery10.1155/2018/6737056 ↗
9. Bessler, J., Prange-Lasonder, G.B., Schaake, L., Saenz, J.F., Bidard, C., Fassi, I., Valori, M., Lassen, A.B., and Buurke, J.H. (2021). Safety assessment of rehabilitation robots: A review identifying safety skills and current knowledge gaps. Front. Rob. AI, 8.10.3389/frobt.2021.602878 ↗
10. Mehrholz, J., Pohl, M., Platz, T., Kugler, J., and Elsner, B. (2018). Electromechanical and robot-assisted arm training for improving activities of daily living, arm function, and arm muscle strength after stroke. Cochrane Database Syst. Rev., CD006876.10.1002/14651858.CD006876.pub5 ↗
11. Robot assisted training for the upper limb after stroke (RATULS): A multicentre randomised controlled trial10.1016/S0140-6736(19)31055-4 ↗
12. Zhao, S., Liu, J., Gong, Z., Lei, Y., OuYang, X., Chan, C.C., and Ruan, S. (2020). Wearable physiological monitoring system based on electrocardiography and electromyography for upper limb rehabilitation training. Sensors, 20.10.3390/s20174861 ↗
13. Poststroke shoulder pain in subacute patients and its correlation with upper limb recovery after robotic or conventional treatment: A secondary analysis of a multicenter randomized controlled trial10.1177/1747493020937192 ↗
14. Standards for the safety of exoskeletons used by industrial workers performing manual handling activities: A contribution from the robo-mate project to their future development10.1016/j.promfg.2015.07.306 ↗
15. Measurement of contact behavior including slippage of cuff when using wearable physical assistant robot10.1109/TNSRE.2015.2464719 ↗
16. Toth, A., Pilissy, T., Bauer, M.O., Al-Absi, G., David, S., and Fazekas, G. (2022, January 25–29). Testing the limit range of motion safety function of upper limb rehabilitation robots with an anthropometrically adjustable and sensorized dummy limb. Proceedings of the 2022 International Conference on Rehabilitation Robotics (ICORR), Rotterdam, The Netherlands.10.1109/ICORR55369.2022.9896575 ↗
17. Design and analysis of bio-mimicking tactile sensor for upper limb prosthesis10.1007/s00542-023-05540-8 ↗
18. (2019). Medical Electrical Equipment—Part 2-78: Particular Requirements for Basic Safety and Essential Performance of Medical Robots for Rehabilitation, Assessment, Compensation, or Alleviation (Standard No. IEC 80601-2-78:2019).
19. (2019). Medical Devices—Application of Risk Management to Medical Devices (Standard No. ISO 14971:2019).
20. (2014). Robots and Robotic Devices—Safety Requirements for Personal Care Robots (Standard No. ISO 13482:2014).
21. (2016). Robots and Robotic Devices—Collaborative Robots (Standard No. ISO/TS 15066:2016).
22. (2011). Robots and Robotic Devices—Safety Requirements for Industrial Robots (Standard No. ISO 10218:2011).
23. Poststroke spasticity: Sequelae and burden on stroke survivors and caregivers10.1212/WNL.0b013e3182764c86 ↗
24. Neuromuscular abnormalities associated with spasticity of upper extremity muscles in hemiparetic stroke10.1152/jn.00049.2007 ↗
25. Adjustments to Zatsiorsky-Seluyanov’s segment inertia parameters10.1016/0021-9290(95)00178-6 ↗
26. Winter, D.A. (2009). Biomechanics and Motor Control of Human Movement, John Wiley & Sons. [4th ed.].10.1002/9780470549148 ↗
27. Passive resistive torques about long bone axes of major human joints
28. Passive visco-elastic properties of the structures spanning the human elbow joint10.1007/BF00430956 ↗
29. Tucan, P., Vaida, C., Plitea, N., Pisla, A., Carbone, G., and Pisla, D. (2019). Risk-based assessment engineering of a parallel robot used in post-stroke upper limb rehabilitation. Sustainability, 11.10.3390/su11102893 ↗
30. A comparison of analytic hierarchy process and conjoint analysis methods in assessing treatment alternatives for stroke rehabilitation10.2165/11587140-000000000-00000 ↗
31. Analytical hierarchy process: Revolution and evolution10.1007/s10479-021-04432-2 ↗
32. Yuan, W., Yang, M., Liu, D., Fu, X., Yu, L., and Wang, K. (2024). Evaluation of the applicability of hospital-affiliated green spaces to patient recovery using the entropy weight method and grey relational analysis. Front. Public Health, 12.10.3389/fpubh.2024.1362884 ↗
33. Computerized cuff pressure algometry as guidance for circumferential tissue compression for wearable soft robotic applications: A systematic review10.1089/soro.2017.0046 ↗
34. Normal range of motion of joints in Male subjects10.2106/00004623-197961050-00017 ↗
35. Range of motion measurements: Reference values and a database for comparison studies10.1111/j.1365-2516.2010.02399.x ↗
36. Tucan, P., Gherman, B., Major, K., Vaida, C., Major, Z., Plitea, N., Carbone, G., and Pisla, D. (2020). Fuzzy logic-based risk assessment of a parallel robot for elbow and wrist rehabilitation. Int. J. Environ. Res. Public Health, 17.10.3390/ijerph17020654 ↗
37. The reliability of three-dimensional kinematic gait measurements: A systematic review10.1016/j.gaitpost.2008.09.003 ↗
38. Biomechanical reactions of exoskeleton neurorehabilitation robots in spastic elbows and wrists10.1109/TNSRE.2017.2714203 ↗
39. The safety issues of medical robotics10.1016/S0951-8320(01)00037-0 ↗