Analysis of Magnetic Resonance Signals from Diffusion Weighted Imaging using Compressed Sensitivity Encoding Technique
Source abstract
Abstract
The purpose of this study was to compare image quality between compressed sensitivity encoding (CS) and sensitivity encoding (SENSE) in single-shot turbo spin echo diffusion-weighted imaging (TSE-DWI). Signal-tonoise ratios (SNR), apparent diffusion coefficient (ADC) values, and geometric accuracies were measured with an American College of Radiology head phantom. CS showed a high mean SNR, less variation in ADC values, and reduced imaging acquisition time (21.7 %-39.3 %) compared with SENSE. In addition, there was no significant difference in geometric accuracy between SENSE and CS. In conclusion, in comparison with SENSE TSE-DWI, CS TSE-DWI resulted in a reduced imaging acquisition time and improved SNR, while maintaining image quality.