Research on Change Factors for Accurate Autocontouring of Organs when RTPs

Jeong-Ho Kim, Man-Seok Han, Min-Cheol Jeon, Se-Jong Yoo

Journal of Magnetics· Volume 28· Issue 4
Publisher page

Abstract

When planning treatment using Radiation Treatment Planning system (RTPs), errors in the contouring process, which is the pre-planning stage, reduce the accuracy of the treatment. Therefore, we wanted to evaluate the error rate of RTPs, Window Width (WW), Window Level (WL), Hounsfield Unit (HU) values at the set point, and the error rate by organ, which are the causes of errors in the contouring process. The results showed that the error rate of WW, WL and HU of certain areas decreased over time. There were no significant changes in the other variables. Therefore, when performing auto-contouring for each organ, appropriate WW, WL and HU values should be set and additional manual contouring should be performed to ensure the accuracy of the treatment plan.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Radiomics and Machine Learning in Medical Imaging · Medical Imaging and Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References

Reference metadata was not supplied for this record.