Fusion Algorithm in Magnetic Resonance/Positron Emission Tomography using Attenuation Correction based on Ultrashort Echo-Time Pulse Sequences
Source abstract
Abstract
Positron emission tomography (PET) images in magnetic resonance (MR)/PET fusion imaging systems are corrected for parts attenuated by gamma-rays by various MR pulse sequences. We aimed to study the quality of MR/PET images using an attenuation correction method based on ultrashort echo-time (UTE) pulse sequences. The proposed image-quality improvement algorithm was modeled as a convergence of the median-modified Wiener filter (MMWF) for noise reduction and Prewitt operator for emphasizing the edge area. By applying the proposed algorithm to MR/PET images, superior contrast to noise ratio and coefficient of variation values were obtained compared to those in the original image. The edge rise distance data of the proposed algorithm exhibited a very small difference from that of the original MR/PET image (difference: 2.21 %). In conclusion, we confirmed the applicability and usefulness of MMWF and Prewitt operator in UTE-based MR/PET imaging.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Advanced MRI Techniques and Applications · Atomic and Subatomic Physics Research · Nuclear Physics and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.