journal article·2025-04-29Open access

Dynamic blood compatibility assessment of coronary stents using an innovative pneumatic blood pump to generate physiological flow rates in small sample chambers

Anne-Sophie Hönicke, Emilia Paz Gross, Simon Beydoun, Felix Pfisterer, Michael Kirschbaum

Abstract

Abstract According to ISO 10993-4, shear-dependent processes must be considered in the blood compatibility testing of cardiovascular implants, which demands controlled and physiological flow conditions in the test environment. For very small test objects such as coronary stents with a diameter below 5 mm, this is difficult to achieve, as previous test methods are either unable to represent the required flow velocities in small-lumen sample chambers, or can only do so under high pump shear stress, which in extreme cases can mask the effects of the tiny test objects. In this paper, we present a novel concept for dynamic in vitro models based on pneumatically generated blood flow ( AirDrive technology), which can achieve flow rates and velocities up to 450 ml min −1 or 930 mm s −1 , respectively. This allows for hemocompatibility testing of coronary stents under physiologically high flow rates without generating major shear stress from the pump mechanism. In an initial feasibility study, single coronary stents with different surface finishes (polished/unpolished) were subjected to flow velocities of up to 560 mm s −1 in a 3.2 mm diameter sample chamber. Blood samples were collected before and after perfusion, and haematological and coagulation markers were analysed. Unpolished stents elicited higher shear-induced blood responses than polished stents. This demonstrates that our experimental setup is highly sensitive and enables precise and robust investigation of blood compatibility under physiologically relevant flow conditions, even for the smallest objects under investigation. To show the superiority of this novel model in creating high flow rates while maintaining minimal blood damage, the AirDrive system was compared to the commonly used roller pump closed-loop system, with the former exhibiting significantly less blood damage. This further confirms that the AirDrive technology we present in this paper is of the highest value for developmental or regulatory testing of blood compatibility of small- and medium-sized test objects. Graphical Abstract
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical Circulatory Support Devices · Cardiac and Coronary Surgery Techniques · Cardiac Valve Diseases and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (31)

Show source-provided references
1. van Oeveren W. Obstacles in hemocompatibility testing. Scientifica. 2013;2013:392584. https://doi.org/10.1155/2013/39258410.1155/2013/392584 ↗
2. Weber M. Blood-contacting biomaterials: in vitro evaluation of the hemocompatibility. Front Bioeng Biotechnol. 2018;6:99. https://doi.org/10.3389/fbioe.2018.0009910.3389/fbioe.2018.00099 ↗
3. van Oeveren W, Tielliu IF, de Hart J. Comparison of modified chandler, roller pump, and ball valve circulation models for in vitro testing in high blood flow conditions: application in thrombogenicity testing of different materials for vascular applications. Int J Biomater. 2012;2012:673163. https://doi.org/10.1155/2012/67316310.1155/2012/673163 ↗
4. Streller U, Sperling C, Hübner J, Hanke R, Werner C. Design and evaluation of novel blood incubation systems for in vitro hemocompatibility assessment of planar solid surfaces. J Biomed Mater Res B Appl Biomater. 2003;66:379–90. https://doi.org/10.1002/jbm.b.1001610.1002/jbm.b.10016 ↗
5. International Organization for Standardization. Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood (ISO/FDIS 10993-4)ISO 10993-4:2017. Published online 2016.
6. Harboe M. A method for determination of hemoglobin in plasma by near-ultraviolet spectrophotometry. Scand J Clin Lab Invest. 1959;11:66–70. https://doi.org/10.3109/0036551590906041010.3109/00365515909060410 ↗
7. Knüppel F, Thomas I, Wurm F-H, Torner B. Suitability of different blood-analogous fluids in determining the pump characteristics of a ventricular assist device. Fluids. 2023. https://doi.org/10.3390/fluids8050151.10.3390/fluids8050151 ↗
8. Zamir M. Hemo-Dynamics. Springer International Publishing; 2016. https://doi.org/10.1007/978-3-319-24103-610.1007/978-3-319-24103-6 ↗
9. Standard Practice for Assessment of Hemolytic Properties of Materials (zugegriffen 15. Februar 2023). https://www.astm.org/f0756-17.html
10. Zhang BL, Bianco RW, Schoen FJ. Preclinical assessment of cardiac valve substitutes: current status and considerations for engineered tissue heart valves. Front Cardiovasc Med. 2019;6:72. https://doi.org/10.3389/fcvm.2019.0007210.3389/fcvm.2019.00072 ↗
11. Cooper SL, Guan J. Advances in polyurethane biomaterials. (Cambridge: Woodhead Publishing, 2016), pp. 691.
12. Thorsen T, Klausen H, Lie RT, Holmsen H. Bubble-induced aggregation of platelets: effects of gas species, proteins, and decompression. Undersea Hyperb Med. 1993;20:101–19.
13. Brandes R, Lang F, Schmidt RF. Hrsg., Physiologie des Menschen: mit Pathophysiologie: mit 850 Farbabbildungen, 32. Auflage. In: Lehrbuch. Berlin: Springer; 2019.10.1007/978-3-662-56468-4 ↗
14. Goodwill AG, Dick GM, Kiel AM, Tune JD. Regulation of coronary blood flow. Compr Physiol. 2017;7:321–82. https://doi.org/10.1002/cphy.c16001610.1002/cphy.c160016 ↗
15. Hall JE. Guyton and hall textbook of medical physiology, 14. Aufl. Philadelphia: Elsevier; 2020.
16. Ratner BD. The blood compatibility catastrophe. J Biomed Mater Res. 1993;27:283–7. https://doi.org/10.1002/jbm.82027030210.1002/jbm.820270302 ↗
17. Jaffer IH, Fredenburgh JC, Hirsh J, Weitz JI. Medical device-induced thrombosis: what causes it and how can we prevent it? J Thromb Haemost. 2015;13:S72–81. https://doi.org/10.1111/jth.1296110.1111/jth.12961 ↗
18. Biran R, Pond D. Heparin coatings for improving blood compatibility of medical devices. Adv Drug Deliv Rev. 2017;112:12–23. https://doi.org/10.1016/j.addr.2016.12.00210.1016/j.addr.2016.12.002 ↗
19. Gore S, Andersson J, Biran R, Underwood C, Riesenfeld J. Heparin surfaces: impact of immobilization chemistry on hemocompatibility and protein adsorption. J Biomed Mater Res B Appl Biomater. 2014;102:1817–24. https://doi.org/10.1002/jbm.b.3315410.1002/jbm.b.33154 ↗
20. Gao W, Lin T, Li T, Yu M, Hu X, Duan D, et al. Sodium alginate/heparin composites on PVC surfaces inhibit the thrombosis and platelet adhesion: applications in cardiac surgery. Int J Clin Exp Med. 2013;6:259–68.
21. Leslie DC. A bioinspired omniphobic surface coating on medical devices prevents thrombosis and biofouling. Nat Biotechnol. 2014. https://doi.org/10.1038/nbt.3020.10.1038/nbt.3020 ↗
22. Chan CHH. Discrete responses of erythrocytes, platelets, and von Willebrand factor to shear. J Biomech. 2022;130:110898. https://doi.org/10.1016/j.jbiomech.2021.11089810.1016/j.jbiomech.2021.110898 ↗
23. Köhne I. Haemolysis induced by mechanical circulatory support devices: unsolved problems. Perfusion. 2020;35:474–83. https://doi.org/10.1177/026765912093130710.1177/0267659120931307 ↗
24. Blok SLJ, van Oeveren W, Engels GE. The optimal incubation time for in vitro hemocompatibility testing: assessment using polymer reference materials under pulsatile flow with physiological wall shear stress conditions. J Biomed Mater Res B Appl Biomater. 2019;107:2335–42. https://doi.org/10.1002/jbm.b.3432610.1002/jbm.b.34326 ↗
25. Mueller M, Krolitzki B, Glasmacher B. Dynamic in vitro hemocompatibility testing – improving the signal to noise ratio. Biomed Eng Biomed Tech. 2012;57:549–52. https://doi.org/10.1515/bmt-2012-421110.1515/bmt-2012-4211 ↗
26. Walter T. Thrombogenicity of sirolimus-eluting stents and bare metal stents: evaluation in the early phase after stent implantation. Vivo. 2010;24:635–9.
27. Brubert J, Krajewski S, Wendel HP, Nair S, Stasiak J, Moggridge GD. Hemocompatibility of styrenic block copolymers for use in prosthetic heart valves. J Mater Sci Mater Med. 2016;27:32. https://doi.org/10.1007/s10856-015-5628-710.1007/s10856-015-5628-7 ↗
28. Olsen AL, Long M. Comparison of catheter thrombogenicity in a modified Chandler loop model using goat blood. J Biomed Mater Res Part A. 2018;106:S. 3143–51. https://doi.org/10.1002/jbm.a.3650910.1002/jbm.a.36509 ↗
29. Vafa Homann M, Johansson D, Wallen H, Sanchez J. Improved ex vivo blood compatibility of central venous catheter with noble metal alloy coating. J Biomed Mater Res B Appl Biomater. 2016;104:1359–65. https://doi.org/10.1002/jbm.b.3340310.1002/jbm.b.33403 ↗
30. Wacker M. An in vitro hemodynamic loop model to investigate the hemocytocompatibility and host cell activation of vascular medical devices. J Vis Exp. 2020. https://doi.org/10.3791/61570.10.3791/61570 ↗
31. Engels GE, Blok SLJ, van Oeveren W. In vitro blood flow model with physiological wall shear stress for hemocompatibility testing—An example of coronary stent testing. Biointerphases. 2016;11:031004. https://doi.org/10.1116/1.495897910.1116/1.4958979 ↗