journal article·2011-02

Corrosion study of single crystal Ni–Mn–Ga alloy and Tb0.27Dy0.73Fe1.95 alloy for the design of new medical microdevices

Pierre Pouponneau, Oumarou Savadogo, Teko Napporn, L’Hocine Yahia, Sylvain Martel

Publisher page

Abstract

Once placed in a magnetic field, smart magnetic materials (SMM) change their shape, which could be use for the development of smaller minimally invasive surgery devices activated by magnetic field. However, the potential degradation and release of cytotoxic ions by SMM corrosion has to be determined. This paper evaluates the corrosion resistance of two SMM: a single crystal Ni-Mn-Ga alloy and Tb(0.27)Dy(0.73)Fe(1.95) alloy. Ni-Mn-Ga alloy displayed a corrosion potential (E (corr)) of -0.58 V/SCE and a corrosion current density (i (corr)) of 0.43 μA/cm(2). During the corrosion assay, Ni-Mn-Ga sample surface was partially protected; local pits were formed on 20% of the surface and nickel ions were mainly found in the electrolyte. Tb(0.27)Dy(0.73)Fe(1.95) alloy exhibited poor corrosion properties such as E (corr) of -0.87 V/SCE and i (corr) of 5.90 μA/cm(2). During the corrosion test, this alloy was continuously degraded, its surface was impaired by pits and cracks extensively and a high amount of iron ions was measured in the electrolyte. These alloys exhibited low corrosion parameters and a selective degradation in the electrolyte. They could only be used for medical applications if they are coated with high strain biocompatible materials or embedded in composites to prevent direct contact with physiological fluids.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Shape Memory Alloy Transformations · High Entropy Alloys Studies · Titanium Alloys Microstructure and Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (37)

Show source-provided references
1. Lanfranco AR, Castellanos AE, Desai JP, Meyers WC. Robotic surgery: a current perspective. Ann Surg. 2004;239(1):14–21.10.1097/01.sla.0000103020.19595.7d ↗
2. Abbott JJ, Nagy Z, Beyeler F, Nelson BJ. Robotics in the small. IEEE Robot & Automation Mag. 2007;14:92–103.10.1109/MRA.2007.380641 ↗
3. Li W, Guo W, Li M, Zhu Y, editors. A novel locomotion principle for endoscopic robot. 2006 IEEE International Conference on Mechatronics and Automation, ICMA 2006, Jun 25–28 2006. Luoyang, China.10.1109/ICMA.2006.257445 ↗
4. Schurr MO. Robotics and telemanipulation technologies for endoscopic surgery. A review of the ARTEMIS project. Surg Endosc. 2000;14(4):375–81.10.1007/s004640020067 ↗
5. Pouponneau P, Leroux JC, Martel S. Magnetic nanoparticles encapsulated into biodegradable microparticles steered with an upgraded magnetic resonance imaging system for tumor chemoembolization. Biomaterials. 2009;30(31):6327–32.10.1016/j.biomaterials.2009.08.005 ↗
6. Pouponneau P, Savadogo O, Napporn T, Yahia L, Martel S. Corrosion study of iron-cobalt alloys for MRI-based propulsion embedded in untethered microdevices operating in the vascular network. J Biomed Mater Res B Appl Biomater. 2010;93(1):203–11.10.1002/jbm.b.31575 ↗
7. Martel S, Mathieu J-B, Felfoul O, Chanu A, Aboussouan E, Tamaz S, et al. Automatic navigation of an untethered device in the artery of a living animal using a conventional clinical magnetic resonance imaging system. Appl Phys Lett. 2007;90(11):114105.10.1063/1.2713229 ↗
8. Soderberg O, Ge Y, Sozinov A, Hannula S-P, Lindroos VK. Recent breakthrough development of the magnetic shape memory effect in Ni–Mn–Ga alloys. Smart Mater and Struct. 2005;14(5):223–35.10.1088/0964-1726/14/5/009 ↗
9. Liu XW, Soderberg O, Ge Y, Lanska N, Ullakko K, Lindroos VK, editors. On the corrosion of non-stoichiometric martensitic Ni–Mn–Ga alloys. J. Phys IV. 2003;112 (II):935–38.10.1051/jp4:20031034 ↗
10. Likhachev AA, Sozinov A, Ullakko K. Different modeling concepts of magnetic shape memory and their comparison with some experimental results obtained in Ni–Mn–Ga. Mater Sci Eng A. 2004;378(1–2 SPEC ISS):513–8.10.1016/j.msea.2003.10.353 ↗
11. Sozinov A, Likhachev AA, Lanska N, Ullakko K. Giant magnetic-field-induced strain in NiMnGa seven-layered martensitic phase. Appl Phys Lett. 2002;80(10):1746.10.1063/1.1458075 ↗
12. Martin E, Epure L, Manceur A, Savadogo O, Yahia LH, editors. Biocompatibility study of NI2–MN–GA magnetic shape memory alloy for biomedical application. 7th Cansmart Workshop: Smart materials and structures; 21-22 October 2004; Montreal, Canada.
13. Stepan LL, Levi DS, Gans E, Mohanchandra KP, Ujihara M, Carman GP. Biocorrosion investigation of two shape memory nickel based alloys: Ni–Mn–Ga and thin film NiTi. J Biomed Mater Res A. 2007;82(3):768–76.10.1002/jbm.a.31192 ↗
14. McMasters OD. Development, application, markets, and patent situation of magnetostrictive TERFENOL-D. J. Rare Earths. 1995;13:295–301.
15. Minagawa H, Kamada K, Nagai H, Nakata Y, Okutani T. Synthesis of Tb0.3Dy0.7Fe1.9 magnetostrictive alloy by unidirectional solidification in magnetic field and microgravity. J Magn Magn Mater. 2002;248(2):230–5.10.1016/S0304-8853(02)00339-6 ↗
16. Cheng LM, Farrell SP, Ham-Su R, Hyatt CV. The influence of composition and thermomechanical treatments on the magnetic shape memory effect of Ni–Mn–Ga single crystals. Smart Structures and Materials 2004. Active Materials: Behavior and Mechanics, 15–18 March 2004; USA: SPIE-Int. Soc. Opt. Eng.10.1117/12.538234 ↗
17. Farrell SP, Dunlap RA, Cheng LM, Ham-Su R, Gharghouri MA, Hyatt CV. Magnetic properties of single crystals of Ni–Mn–Ga magnetic shape memory alloys. Smart Structures and Materials 2004. Active Materials: Behavior and Mechanics, 15–18 March 2004; 2004; USA: SPIE-Int. Soc. Opt. Eng.10.1117/12.539770 ↗
18. American Society For Testing and Materials. Annual Book of ASTM standard. Philadelphia, USA 2000.
19. Tian X, Emteborg H, Adams FC. Analytical performance of axial inductively coupled plasma time of flight mass spectrometry (ICP-TOFMS). J Ana At Spectrom. 1999;14(12):1807–14.10.1039/a906762k ↗
20. Starosvetsky D, Gotman I. Corrosion behavior of titanium nitride coated Ni–Ti shape memory surgical alloy. Biomaterials. 2001;22(13):1853–9.10.1016/S0142-9612(00)00368-9 ↗
21. De Assis SL, Wolynec S. Costa. Corrosion characterization of titanium alloys by electrochemical techniques. Electrochim Acta. 2006;51:1815–9.10.1016/j.electacta.2005.02.121 ↗
22. Revie R. Uhlig’s corrosion handbook. New York: John Wiley & Sons Inc; 2000.
23. Cisse O, Savadogo O, Wu M, Yahia LH. Effect of surface treatment of NiTi alloy on its corrosion behavior in Hanks’ solution. J Biomed Mater Res. 2002;61(3):339–45.10.1002/jbm.10114 ↗
24. Kaesche H. Corrosion of metals: physicochemical principles and current problems. Heidelberg, NY : Springer-Verlag; 2003.10.1007/978-3-642-96038-3 ↗
25. Pouponneau P, Yahia LH, Merhi Y, Epure LM, Martel S. Biocompatibility of candidate materials for the realization of medical microdevices. 28th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS’06, Aug 30-Sep 3 2006. NY, USA.10.1109/IEMBS.2006.260061 ↗
26. Barceloux DG. Nickel. J Toxicol Clin Toxicol. 1999;37(2):239–58.10.1081/CLT-100102423 ↗
27. Assis SL, Rogero SO, Antunes RA, Padilha AF, Costa I. A comparative study of the in vitro corrosion behavior and cytotoxicity of a superferritic stainless steel, a Ti-13Nb-13Zr alloy, and an austenitic stainless steel in Hank’s solution. J Biomed Mater Res B Appl Biomater. 2005;73(1):109–16.10.1002/jbm.b.30205 ↗
28. Chang LW (ed). Toxicology of metals: CRC Press, Inc.; 1996.
29. Gurrappa I. Corrosion studies related to suitability of permanent magnets for biomedical applications. Mater Charact. 2002;48(1):63–70.10.1016/S1044-5803(02)00248-6 ↗
30. Reclaru L, Luthy H, Eschler P-Y, Blatter A, Susz C. Corrosion behaviour of cobalt–chromium dental alloys doped with precious metals. Biomaterials. 2005;26(21):4358–65.10.1016/j.biomaterials.2004.11.018 ↗
31. Sachdeva D, Balasubramaniam R. The effect of hydrogen on corrosion of Tb0.3Dy0.7Fe1.92. J Electrochem Soc. 2008;155:C75–80.10.1149/1.2816226 ↗
32. Noar JH, Wahab A, Evans RD, Wojcik AG. The durability of parylene coatings on neodymium–iron–boron magnets. Eur J Orthod. 1999;21(6):685–93.10.1093/ejo/21.6.685 ↗
33. Yoshida M, Langer R, Lendlein A, Lahann J. From advanced biomedical coatings to multi-functionalized biomaterials. Polym Rev. 2006;46(4):347–75.10.1080/15583720600945394 ↗
34. Strickler F, Richard R, McFadden S, Lindquist J, Schwarz MC, Faust R, et al. In vivo and in vitro characterization of poly(styrene-b-isobutylene-b-styrene) copolymer stent coatings for biostability, vascular compatibility and mechanical integrity. J Biomed Mater Res A. 2010;92:773–82.10.1002/jbm.a.32418 ↗
35. Hosoda H, Takeuchi S, Inamura T, Wakashima K. Material design and shape memory properties of smart composites composed of polymer and ferromagnetic shape memory alloy particles. Sci Technol Adv Mater. 2004;5:503–9.10.1016/j.stam.2004.02.009 ↗
36. Min Z, Siu Wing O, Chan HLW. DC- and AC-magnetic field-induced strain effects in ferromagnetic shape memory composites of Ni–Mn–Ga single crystal and polyurethane polymer. J Appl Phys. 2010;107:09–942.
37. Kwon OY, Kim HY, Cha SI, Hong SH. Magnetostriction and magnetomechanical coupling of grain-aligned Tb 0.33 Dy0.67Fey/epoxy-filled composites. J Appl Phys. 2005;97(11):113905.10.1063/1.1925335 ↗