Structural evaluation, preliminary in vitro stability and electrochemical behavior of apatite coatings on Ti6Al4V substrates
Source abstract
Abstract
Abstract Medical-grade alloys, such as Ti-6Al-4V, have been used for fixation of fractured bone and for the total replacement of defective bone. Their bioactivity could be improved by applying a bone-like apatite layer onto their surfaces. This, in turn, enhances their integration with the surrounding tissues upon implantation. In addition, the presence of a bioactive bone-like coating minimizes the likelihood of corrosion. Various methods are known for the formation of apatite coating onto Ti-6Al-4V, among which sputtering has shown its promise as a simple direct method. In the current work, a sputtering technique was used to develop a 300 nm-thick bone-like apatite layer onto Ti-6Al-4V. Structural composition, integrity and morphology of the as-coated and thermally treated coatings were investigated. Coated substrates were further evaluated after soaking them in a simulated body fluid (SBF) for up to 14 days. Results showed the formation of an amorphous apatite layer onto the alloy, that was further shown to partially crystallize upon heat treatment. As a result of SBF treatment, the apatite layer was found to remodel through a dissolution-precipitation mechanism due to its amorphous and non-stoichiometric nature, forming a smooth layer with better homogeneity and decreased surface roughness. Electrochemical analysis of the coated alloys showed the enhanced corrosion protection of the alloy surfaces by coating them with apatite. In addition, pre-grinding of the alloy surfaces before the formation of the coating was also found to improve the corrosion inhibition of the alloy surfaces in aqueous media.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Titanium Alloys Microstructure and Properties · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (48)
Show source-provided references
1. Ratner · 1996
2. Composite technology in load-bearing orthopedic implants10.1016/S0142-9612(97)00217-2 ↗
3. American Standards for Testing and Measurements (ASTM) F138-13a, Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673).
4. American Standards for Testing and Measurements (ASTM) F75-12, ASTM Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075)
5. American Standards for Testing and Measurements (ASTM) F799-06, ASTM Standard Specification for Cobalt-28Chromium-6Molybdenum Alloy Forgings for Surgical Implants (UNS R31537, R31538, R31539)
6. American Standards for Testing and Measurements (ASTM) F67-13, ASTM Standard Specification for Unalloyed Titanium, for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700)
7. American Standards for Testing and Measurements (ASTM) F136-13, ASTM Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)
8. Tissue reactions to metallic wear and corrosion products in human patients10.1002/jbm.820080304 ↗
9. Comparison of metal concentrations in rat tibia tissues with various metallic implants10.1016/j.biomaterials.2004.01.064 ↗
10. The release of cobalt and nickel from a new total finger joint prosthesis made of vitallium10.1002/jbm.820170410 ↗
11. Effects of surface treatment of Ti-6Al-4V titanium alloy on biocompatibility in cultured human umbilical vein endothelial cells10.1016/j.actbio.2004.11.001 ↗
12. Total hip arthroplasty with hydroxyapatite-coated prostheses10.2106/00004623-199612000-00018 ↗
13. Orthopedic coatings10.3928/0147-7447-19940301-07 ↗
14. Microstructural characterizations of plasma sprayed hydroxyapatite coatings10.1016/S0257-8972(02)00855-1 ↗
15. Influence of spraying conditions on thermal and velocity properties of plasma sprayed hydroxyapatite10.1016/j.msec.2006.05.002 ↗
16. Direct electron microscopy studies of the bone-hydroxyapatite interface10.1002/jbm.820180702 ↗
17. Osteoconductive coatings for total joint arthroplasty10.1097/00003086-200202000-00007 ↗
18. Surface structural heterogeneity of high power plasma-sprayed hydroxyapatite coatings10.1016/j.jallcom.2016.06.163 ↗
19. Titanium dioxide reinforced hydroxyapatite coatings deposited by high-velocity oxy-fuel (HVOF) spray10.1016/S0142-9612(01)00082-5 ↗
20. Crystallite orientation and anisotropic strains in thermally sprayed hydroxyapatite coatings10.1016/0142-9612(95)99696-J ↗
21. Influence of thickness on the properties of hydroxyapatite coatings deposited by KrF laser ablation10.1016/S0142-9612(00)00408-7 ↗
22. Thin hydroxyapatite coating produced by the ion beam dynamic mixing method10.1016/0142-9612(94)90019-1 ↗
23. Calcium phosphate coating on nickel-titanium shape memory alloys. Coating procedure and adherence of leukocytes and platelets10.1016/S0142-9612(03)00241-2 ↗
24. Sol-gel hydroxyapatite coatings on stainless steel substrates10.1016/S0142-9612(01)00157-0 ↗
25. Electrophoretic deposition of porous hydroxyapatite scaffold10.1016/S0142-9612(03)00203-5 ↗
26. Biomimetic hydroxyapatite coating on metal implants10.1111/j.1151-2916.2002.tb00126.x ↗
27. Hot isostatic pressing-processed hydroxyapatite-coated titanium implants: light microscopic and scanning electron microscopy investigations
28. In vivo dissolution behavior of various RF magnetron sputtered Ca-P coatings on roughened titanium implants10.1016/S0142-9612(03)00067-X ↗
29. Properties and immersion behavior of magnetron-sputtered multi-layered hydroxyapatite/titanium composite coatings10.1016/S0142-9612(03)00315-6 ↗
30. A review on calcium phosphate coatings produced using a sputtering process–—an alternative to plasma spraying10.1016/j.biomaterials.2004.02.029 ↗
31. Strontium-substituted hydroxyapatite coatings deposited via a co-deposition sputter technique10.1016/j.msec.2014.10.046 ↗
32. How useful is SBF in predicting in vivo bone bioactivity?10.1016/j.biomaterials.2006.01.017 ↗
33. Human mesenchymal stem cell proliferation and osteogenic differentiation in fibrin gel in vitro10.1089/ten.2006.12.2385 ↗
34. Hydroxyapatite-coated Ti-6Al-4V Part 2: the effects of post deposition heat treatment at low temperatures10.1016/S0142-9612(01)00320-9 ↗
35. Oxidation of Ti-6Al-4V alloy10.1016/j.jallcom.2008.04.024 ↗
36. XPS analysis of chemical properties of calcium phosphate thin films and osteoblastic hos cell response
37. Characterization of bioactive films on Ti-6Al-4V alloy10.1016/j.electacta.2012.12.081 ↗
38. Control of biomimetic hydroxyapatite deposition on polymer substrates using different protein adsorption abilities10.1016/j.colsurfb.2015.04.010 ↗
39. Evaluation in vitro and in vivo of biomimetic hydroxyapatite-coated on titanium dental implants10.1016/j.msec.2004.08.044 ↗
40. Boukamp · Solid State Ion. · 1986 · 20 · 3110.1016/0167-2738(86)90031-7 ↗
41. Macdonald · 1987
42. Satpati · Mater. Chem. Phys. · 2008 · 109 · 35210.1016/j.matchemphys.2007.12.002 ↗
43. Mertens · Corrosion · 1997 · 53 · 38110.5006/1.3280481 ↗
44. R. Macdonald, D.R. Franceschetti, in: J.R. Macdonald (Ed.), Impedance Spectroscopy, Wiley New York, 1987
45. Bellanger · Corros. Sci. · 2006 · 48 · 137910.1016/j.corsci.2005.05.022 ↗
46. Sherif · Electrochim. Acta · 2006 · 51 · 466510.1016/j.electacta.2006.01.007 ↗
47. Satpati · Mater. Chem. Phys. · 2008 · 109 · 35210.1016/j.matchemphys.2007.12.002 ↗
48. Gallardo · Corros. Sci. · 2004 · 46 · 79510.1016/S0010-938X(03)00185-9 ↗